In vitro Evaluation of the Accuracy of the Root Zx in the Presence of Naocl 2.5% and Chlorhexidine 0.2
- PMID: 30607220
- PMCID: PMC6311399
- DOI: 10.4317/jced.54193
In vitro Evaluation of the Accuracy of the Root Zx in the Presence of Naocl 2.5% and Chlorhexidine 0.2
Abstract
Background: Establishing the working length at the apical constriction is considered ideal for root canal therapy. Because of the limitations of radiography and complicacy of the apex of the root, electronic measurement of canal length has become valuable for endodontic treatment. This study was designed to evaluation of the accuracy of the Root zx electronic apex locator in the presence of NaOCl 2.5% and chlorhexidine 0.2%.
Material and methods: Thirty extracted human premolars with complete root formation were enrolled. The actual length (AL) was assessed visually (under stereo microscope) and teeth mounted in the saline model. The electronic length (EL) measurements were recorded in the presence of NaOCl 2.5% and chlorhexidine 0.2% and the differences between the EL and AL were compared.
Results: By accepting the error of 0.5 and 1 mm, the accuracy of Root zx was 76.7% and 96.7% in the presence of chlorhexidine 0.2% and 90% and 100% in the presence of NaOCl 2.5%, respectively. No statistical differences was found between the measured groups (P=0.223).
Conclusions: Our results confirmed that Root zx can accurately determine the apical constriction in presence of both NaOCl 2.5% and chlorhexidine 0.2%. Key words:Chlorhexidine, Sodium Hypochlorite, Root ZX.
Conflict of interest statement
Conflict of interest statement: The authors have declared that no conflict of interest exist.
Similar articles
-
A comparative assessment of the accuracy of electronic apex locator (Root ZX) in the presence of commonly used irrigating solutions.J Clin Exp Dent. 2014 Feb 1;6(1):e41-6. doi: 10.4317/jced.51230. eCollection 2014 Feb. J Clin Exp Dent. 2014. PMID: 24596634 Free PMC article.
-
An Invitro Evaluation of the Accuracy of Two Electronic Apex Locators to Determine Working Length in the Presence of Various Irrigants.Ethiop J Health Sci. 2016 Sep;26(5):457-462. doi: 10.4314/ejhs.v26i5.7. Ethiop J Health Sci. 2016. PMID: 28446851 Free PMC article.
-
Comparison of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study.J Conserv Dent. 2012 Apr;15(2):178-82. doi: 10.4103/0972-0707.94585. J Conserv Dent. 2012. PMID: 22557820 Free PMC article.
-
Evaluation of accuracy of an electronic apex locator in presence of sodium hypochlorite in primary teeth with and without resorption.J Clin Pediatr Dent. 2023 Nov;47(6):150-154. doi: 10.22514/jocpd.2023.089. Epub 2023 Nov 3. J Clin Pediatr Dent. 2023. PMID: 37997246
-
The effect of different root canal irrigations on the accuracy of apex locators: A systematic review.J Clin Exp Dent. 2024 Dec 1;16(12):e1538-e1546. doi: 10.4317/jced.62049. eCollection 2024 Dec. J Clin Exp Dent. 2024. PMID: 39822780 Free PMC article. Review.
Cited by
-
Factors influencing the accuracy of electronic apex locators: A scoping review.Dent Res J (Isfahan). 2025 Jun 27;22:26. doi: 10.4103/drj.drj_634_23. eCollection 2025. Dent Res J (Isfahan). 2025. PMID: 40655901 Free PMC article. Review.
-
Accuracy of Three Types of Apex Locators versus Digital Periapical Radiography for Working Length Determination in Maxillary Premolars: An In Vitro Study.Clin Pract. 2022 Dec 12;12(6):1043-1053. doi: 10.3390/clinpract12060107. Clin Pract. 2022. PMID: 36547115 Free PMC article.
-
The Effect of Long Service Life on the Accuracy of Dentaport ZX Electronic Apex Locator.Iran Endod J. 2022;17(4):195-199. doi: 10.22037/iej.v17i4.34623. Iran Endod J. 2022. PMID: 36703688 Free PMC article.
-
Ex vivo evaluation of the accuracy of four electronic foraminal locators during endodontic retreatment.J Clin Exp Dent. 2024 Mar 1;16(3):e257-e262. doi: 10.4317/jced.61133. eCollection 2024 Mar. J Clin Exp Dent. 2024. PMID: 38600923 Free PMC article.
References
-
- Plotino G, Grande NM, Brigante L, Lesti B, Somma F. Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex. Int Endod J. 2006;39:408–14. - PubMed
-
- Vieyra JP, Acosta J, Mondaca JM. Comparison of working length determination with radiographs and two electronic apex locators. Int Endod J. 2010;43:16–20. - PubMed
-
- Alothmani OS. The accuracy of Root ZX electronic apex locator. Saudi Endod J. 2012;2:115–30.
LinkOut - more resources
Full Text Sources
Miscellaneous