In vitro Evaluation of the Accuracy of the Root Zx in the Presence of Naocl 2.5% and Chlorhexidine 0.2
- PMID: 30607220
- PMCID: PMC6311399
- DOI: 10.4317/jced.54193
In vitro Evaluation of the Accuracy of the Root Zx in the Presence of Naocl 2.5% and Chlorhexidine 0.2
Abstract
Background: Establishing the working length at the apical constriction is considered ideal for root canal therapy. Because of the limitations of radiography and complicacy of the apex of the root, electronic measurement of canal length has become valuable for endodontic treatment. This study was designed to evaluation of the accuracy of the Root zx electronic apex locator in the presence of NaOCl 2.5% and chlorhexidine 0.2%.
Material and methods: Thirty extracted human premolars with complete root formation were enrolled. The actual length (AL) was assessed visually (under stereo microscope) and teeth mounted in the saline model. The electronic length (EL) measurements were recorded in the presence of NaOCl 2.5% and chlorhexidine 0.2% and the differences between the EL and AL were compared.
Results: By accepting the error of 0.5 and 1 mm, the accuracy of Root zx was 76.7% and 96.7% in the presence of chlorhexidine 0.2% and 90% and 100% in the presence of NaOCl 2.5%, respectively. No statistical differences was found between the measured groups (P=0.223).
Conclusions: Our results confirmed that Root zx can accurately determine the apical constriction in presence of both NaOCl 2.5% and chlorhexidine 0.2%. Key words:Chlorhexidine, Sodium Hypochlorite, Root ZX.
Conflict of interest statement
Conflict of interest statement: The authors have declared that no conflict of interest exist.
References
-
- Plotino G, Grande NM, Brigante L, Lesti B, Somma F. Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex. Int Endod J. 2006;39:408–14. - PubMed
-
- Vieyra JP, Acosta J, Mondaca JM. Comparison of working length determination with radiographs and two electronic apex locators. Int Endod J. 2010;43:16–20. - PubMed
-
- Alothmani OS. The accuracy of Root ZX electronic apex locator. Saudi Endod J. 2012;2:115–30.
LinkOut - more resources
Full Text Sources
Miscellaneous