Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes
- PMID: 30652032
- PMCID: PMC6311789
- DOI: 10.1155/2018/3258396
Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes
Abstract
Background: Secretions accumulate in endotracheal tubes' (ETT) lumens upon their placement in patients. The secretions impact airway resistance and pressure. Secretions potentiate prolonged mechanical ventilation and ventilator-associated pneumonia. Our primary objective in this study was to evaluate an ETT-clearing device (ETT-CD) in its ability to remove secretions from ex vivo ETT lumens.
Methods: Forty ETTs, obtained from intensive care patients at extubation, were individually placed into a ventilator field performance testing simulator at 37°C. The pressure drop through the ETTs was measured at a flow rate of 60 L/min before and after cleaning with the ETT-CD and compared with unused, similarly sized controls tubes. The ETT-CD was inserted into an ETT until the tip reached Murphy's eye (hole in the side) of the ETT. The wiper, set back from the tip, was expanded by ETT-CD handle activation. As the ETT-CD was removed, the distal wiper extracted secretions from the ETT lumen.
Results: Forty ETTs were tested with nonparametric Wilcoxon signed-rank tests. Before being cleared with the ETT-CD, the median pressure drop in the extubated 7.5 mm ETTs was 17.8 cm H2O; after ETT-CD use, it was 12.3. The cleared ETTs were significantly improved over the ETTs before being cleared (p < 0.001); however, there remained a significant difference between the cleared ETTs and the control tubes (p=0.005), indicating the clearing was not to the level of an unused ETT. Similar results were determined for the 8.0 mm ETTs.
Conclusions: For the 7.5 mm and the 8.0 mm EETs, the ETT-CD improved effective patency of the ETTs over the uncleared ETTs, independent of occlusion location, tube size, or length of tube. However, there remained a significant difference between the cleared tubes and controls.
Similar articles
-
Impact of endotracheal tube biofilm and respiratory secretions on airway resistance and mechanics of breathing in a neonatal lung model.J Appl Physiol (1985). 2018 Oct 1;125(4):1227-1231. doi: 10.1152/japplphysiol.00083.2018. Epub 2018 Jul 26. J Appl Physiol (1985). 2018. PMID: 30048200
-
Effects of prolonged mechanical ventilation with a closed suction system on endotracheal tube resistance and its reversibility by a closed suction cleaning system.Anaesth Intensive Care. 2013 Nov;41(6):728-35. doi: 10.1177/0310057X1304100607. Anaesth Intensive Care. 2013. PMID: 24180713
-
Endotracheal tubes and fluid aspiration: an in vitro evaluation of new cuff technologies.BMC Anesthesiol. 2017 Mar 4;17(1):36. doi: 10.1186/s12871-017-0328-0. BMC Anesthesiol. 2017. PMID: 28257624 Free PMC article.
-
Modifying endotracheal tubes to prevent ventilator-associated pneumonia.Curr Opin Infect Dis. 2011 Apr;24(2):157-62. doi: 10.1097/QCO.0b013e328343b733. Curr Opin Infect Dis. 2011. PMID: 21252657 Review.
-
Lubricity, wear prevention, and anti-biofouling properties of macromolecular coatings for endotracheal tubes.Biomater Sci. 2024 Feb 27;12(5):1228-1238. doi: 10.1039/d3bm01985c. Biomater Sci. 2024. PMID: 38230671 Review.
Cited by
-
Adherence of critical care nurses to endotracheal suctioning guidelines: a cross-sectional study.BMC Nurs. 2022 Nov 14;21(1):312. doi: 10.1186/s12912-022-01092-w. BMC Nurs. 2022. PMID: 36376904 Free PMC article.
-
Evaluation of a Closed Suction System With Integrated Tube-Scraping Technology.Respir Care. 2023 Aug;68(8):1023-1030. doi: 10.4187/respcare.10830. Epub 2023 Jan 20. Respir Care. 2023. PMID: 36669781 Free PMC article. Clinical Trial.
References
-
- Filiberto D. M., Croce M. A. Surgical Critical Care Therapy 2018. Cham, Switzerland: Springer; 2018. Ventilator-associated pneumonia; pp. 407–413. - DOI
LinkOut - more resources
Full Text Sources
Research Materials