The value of tibial mounted inertial measurement units to quantify running kinetics in elite football (soccer) players. A reliability and agreement study using a research orientated and a clinically orientated system

Affiliations

The value of tibial mounted inertial measurement units to quantify running kinetics in elite football (soccer) players. A reliability and agreement study using a research orientated and a clinically orientated system

Tom Hughes et al. J Electromyogr Kinesiol. 2019 Feb.

Abstract

In elite football, measurement of running kinetics with inertial measurement units (IMUs) may be useful as a component of periodic health examination (PHE). This study determined the reliability of, and agreement between a research orientated IMU and clinically orientated IMU system for initial peak acceleration (IPA) and IPA symmetry index (SI) measurement during running in elite footballers. On consecutive days, 16 participants performed treadmill running at 14kmph and 18kmph. Both IMUs measured IPA and IPA SI concurrently. All measurements had good or excellent within-session reliability (intraclass correlation coefficient (ICC2,1) range = 0.79-0.96, IPA standard error of measurement (SEM) range = 0.19-0.62 g, IPA SI SEM range = 2.50-8.05%). Only the research orientated IMU demonstrated acceptable minimal detectable changes (MDCs) for IPA at 14kmph (range = 7.46-9.80%) and IPA SI at both speeds (range = 6.92-9.21%). Considering both systems, between-session IPA reliability ranged from fair to good (ICC2,1 range = 0.63-0.87, SEM range = 0.51-1.10 g) and poor to fair for IPA SI (ICC2,1 range = 0.32-0.65, SEM range = 8.07-11.18%). All MDCs were >10%. For IPA and SI, the 95% levels of agreement indicated poor between system agreement. Therefore, the use of IMUs to evaluate treadmill running kinetics cannot be recommended in this population as a PHE test to identify prognostic factors for injuries or for rehabilitation purposes.

Keywords: Accelerometer; Gait; Lower extremity; Symmetry.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Photograph of inertial measurement unit placement on left tibia.
Fig. 2
Fig. 2
Bland-Altman plots to demonstrate agreement between systems for IPA measures at 14 and 18kmph on both testing days. Key: IPA = initial peak acceleration; diff = difference; g = gravitational force; kmph = kilometres per hour. Note: Thick black lines correspond to 95% limits of agreement, dashed black line corresponds to observed mean agreement. Where the y axis is 0, this indicates line of perfect agreement.
Fig. 3
Fig. 3
Bland-Altman plots to demonstrate agreement between systems for IPA SI on both testing days. Key: SI = symmetry index; diff = difference; % = percentage; kmph = kilometres per hour. Note: Thick black lines corresponds to 95% limits of agreement, dashed black line correspond to observed mean agreement. Where the y axis is 0, this indicates line of perfect agreement.

Similar articles

Cited by

References

    1. Al-Amri M., Nicholas K., Button K., Sparkes V., Sheeran L., Davies J.L. Inertial measurement units for clinical movement analysis: reliability and concurrent validity. Sensors. 2018;18 - PMC - PubMed
    1. Bangsbo J., Mohr M., Krustrup P. Physical and metabolic demands of training and match-play in the elite football player. J. Sports Sci. 2006;24:665–674. - PubMed
    1. Bergamini E., Ligorio G., Summa A., Vannozzi G., Cappozzo A., Sabatini A.M. Estimating orientation using magnetic and inertial sensors and different sensor fusion approaches: accuracy assessment in manual and locomotion tasks. Sensors. 2014;14:18625–18649. - PMC - PubMed
    1. Bland J.M., Altman D.G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;346:307–310. - PubMed
    1. Brophy R.H., Backus S.I., Pansy B.S., Lyman S., Williams R.J. Lower extremity muscle activation and alignment during the soccer instep and side-foot kicks. J. Orthop. Sports Phys. Ther. 2007;37:260–268. - PubMed