Anterior segment fluorescein cineangiography
Abstract
We have developed a technique for performing anterior segment fluorescein cineangiography. Illumination is obtained with a halogen lamp of a standard slide projector that is fitted with a blue excitation filer. Cinematography occurs with a movie camera fitted with an absorption-type barrier filter and mounted to a photo slit lamp through a cineadapter. The technique has been successfully employed with animals, and it is anticipated that the light levels used are tolerable and safe for application with humans.