Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2019 Jan 24;19(3):470.
doi: 10.3390/s19030470.

Vital Sign Monitoring and Cardiac Triggering at 1.5 Tesla: A Practical Solution by an MR-Ballistocardiography Fiber-Optic Sensor

Affiliations

Vital Sign Monitoring and Cardiac Triggering at 1.5 Tesla: A Practical Solution by an MR-Ballistocardiography Fiber-Optic Sensor

Jan Nedoma et al. Sensors (Basel). .

Abstract

This article presents a solution for continuous monitoring of both respiratory rate (RR) and heart rate (HR) inside Magnetic Resonance Imaging (MRI) environments by a novel ballistocardiography (BCG) fiber-optic sensor. We designed and created a sensor based on the Fiber Bragg Grating (FBG) probe encapsulated inside fiberglass (fiberglass is a composite material made up of glass fiber, fabric, and cured synthetic resin). Due to this, the encapsulation sensor is characterized by very small dimensions (30 × 10 × 0.8 mm) and low weight (2 g). We present original results of real MRI measurements (conventionally most used 1.5 T MR scanner) involving ten volunteers (six men and four women) by performing conventional electrocardiography (ECG) to measure the HR and using a Pneumatic Respiratory Transducer (PRT) for RR monitoring. The acquired sensor data were compared against real measurements using the objective Bland⁻Altman method, and the functionality of the sensor was validated (95.36% of the sensed values were within the ±1.96 SD range for the RR determination and 95.13% of the values were within the ±1.96 SD range for the HR determination) by this means. The accuracy of this sensor was further characterized by a relative error below 5% (4.64% for RR and 4.87% for HR measurements). The tests carried out in an MRI environment demonstrated that the presence of the FBG sensor in the MRI scanner does not affect the quality of this imaging modality. The results also confirmed the possibility of using the sensor for cardiac triggering at 1.5 T (for synchronization and gating of cardiovascular magnetic resonance) and for cardiac triggering when a Diffusion Weighted Imaging (DWI) is used.

Keywords: MRI-compatible; ballistocardiography (BCG); cardiac triggering; fiber bragg grating (FBG); fiberglass; heart rate (HR); respiratory rate (RR).

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflicts of interest.

Figures

Figure 1
Figure 1
ECG signal in various types of magnetic fields (0 T, 1.5 T and 3 T).
Figure 2
Figure 2
Blog diagram of the FBG structure.
Figure 3
Figure 3
A prototype of created and designed measuring sensor.
Figure 4
Figure 4
A reflection spectrum of the Bragg grating before and after encapsulation.
Figure 5
Figure 5
The dependence of Bragg wavelength on the temperature during the temperature loading in the temperature box.
Figure 6
Figure 6
Sample recordings of electrocardiogram (ECG) and Ballistocardiography (BCG) signals.
Figure 7
Figure 7
Signal processing from the FBG sensor to determine respiratory and heart rate [44].
Figure 8
Figure 8
Positioning of the FBG sensor on the human body and the experimental setup.
Figure 9
Figure 9
(a) a schematic location of the FBG sensor on the human body; (b) a photo taken during a real measurement (test subject M1).
Figure 10
Figure 10
A sample recording of breathing activity using the FBG and the reference device: (a) M1 volunteer, and (b) F1 volunteer.
Figure 11
Figure 11
The full-time course of breathing activity over the conducted MRI examinations: (a) volunteer M1, and (b) volunteer F1.
Figure 12
Figure 12
Reproducibility of RR determination capabilities of our FBG sensor (based on comparison with Pneumatic Respiratory Transducer-based RR calculations) using the Bland–Altman method for the data acquired from: (a) volunteer M1, and (b) volunteer F1.
Figure 13
Figure 13
Ten-second long sample recordings of the heart activity; (a) M1 volunteer, and (b) F1 volunteer.
Figure 14
Figure 14
The full-time course of cardiac activity over the conducted MRI examinations: (a) volunteer M1, and (b) volunteer F1.
Figure 15
Figure 15
Reproducibility of HR determination capabilities of our FBG sensor (based on comparison with ECG-based HR calculations) using the Bland–Altman analysis method and data acquired from: (a) volunteer M1, and (b) volunteer F1.
Figure 16
Figure 16
A sample MRI image (axial view) showing where the FBG sensor was positioned: A GE T2-weighted image (400 × 400 pixels).
Figure 17
Figure 17
(a) An example of ECG signal interference in DWI MRI sequences, (b) An example of a signal obtained from the FBG sensor before and after filtering.
Figure 18
Figure 18
Example of analyzed R–J interval.

Similar articles

Cited by

References

    1. Weckesser M., Posse S., Olthoff U., Kemna L., Dager S., Müller-Gärtner H.W. Functional imaging of the visual cortex with bold-contrast MRI: Hyperventilation decreases signal response. Mag. Res. Med. 1999;41:213–216. doi: 10.1002/(SICI)1522-2594(199901)41:1<213::AID-MRM31>3.0.CO;2-S. - DOI - PubMed
    1. Giardino N.D., Friedman S.D., Dager S.R. Anxiety, respiration, and cerebral blood flow: Implications for functional brain imaging. Compr. Psychiatry. 2007;48:103–112. doi: 10.1016/j.comppsych.2006.11.001. - DOI - PMC - PubMed
    1. Tamaki S., Yamada T., Okuyama Y., Morita T., Sanada S., Tsukamoto Y. Cardiac iodine-123 metaiodobenzylguanidine imaging predicts sudden cardiac death independently of left ventricular ejection fraction in patients with chronic heart failure and left ventricular systolic dysfunction: results from a comparative study with signal-averaged electrocardiogram, heart rate variability, and QT dispersion. J. Am. Coll. Cardiol. 2009;53:426–435. - PubMed
    1. Zabel M., Acar B., Klingenheben T., Franz M.R., Hohnloser S.H., Malik M. Analysis of 12-lead T-wave morphology for risk stratification after myocardial infarction. Circulation. 2000;102:1252–1257. doi: 10.1161/01.CIR.102.11.1252. - DOI - PubMed
    1. Ogura R., Hiasa Y., Takahashi T., Yamaguchi K., Fujiwara K., Ohara Y., Hosokawa S. Specific findings of the standard 12-lead ECG in patients with takotsubo’ cardiomyopathy. Circ. J. 2003;67:687–690. doi: 10.1253/circj.67.687. - DOI - PubMed

LinkOut - more resources