. 2018 Mar 20:5:13-18.
doi: 10.2147/RSRR.S142562. eCollection 2018.

A hand-guided robotic drill for cochleostomy on human cadavers

Affiliations

A hand-guided robotic drill for cochleostomy on human cadavers

Xinli Du et al. Robot Surg. .

Abstract

Background: An arm supported robotic drill has been recently demonstrated for preparing cochleostomies in a pilot research clinical trial. In this paper, a hand-guided robotic drill is presented and tested on human cadaver trials.

Methods: The innovative smart tactile approach can automatically detect drilling mediums and decided when to stop drilling to prevent penetrating the endosteum. The smart sensing scheme has been implemented in a concept of a hand guided robotic drill.

Results: Experiments were carried out on two adult cadaveric human bodies for verifying the drilling process and successfully finished cochleostomy on three cochlea. The advantage over a system supported by a mechanical arm includes the flexibility in adjusting the trajectory to initiate cutting without slipping. Using the same concept as a conventional drilling device, the user will also be benefit from the lower setup time and cost, and lower training overhead.

Conclusion: The hand-guided robotic drill was recently developed for testing on human cadavers. The robotic drill successfully prepared cochleostomies in all three cases.

Keywords: cochlear implantation; drilling cochleostomies; hand guided robot; hearing preservation; smart sensing; surgical robot.

PubMed Disclaimer

Conflict of interest statement

Disclosure The authors report no conflicts of interest in this work.

Figures

Figure 1
Figure 1
Diagram illustrating the anatomy of the ear and location of a cochleostomy. Note: Reproduced with the permission of John Wiley and Sons. Du X, Assadi MZ, Jowitt F, et al. Robustness analysis of a smart surgical drill for cochleostomy. Int J Med Robot. 2013;9(1):119–126.
Figure 2
Figure 2
The experimental hand-guided surgical robot drill system. Note: Open Access Creative Commons, Brett P, Du X, Zoka-Assadi M, Coulson C, Reid A, Proops D. Feasibility study of a hand guided robotic drill for cochleostomy. Biomed Res Int. 2014;2014:7.
Figure 3
Figure 3
The hand-guided robotic drill unit. Note: Open Access Creative Commons, Brett P, Du X, Zoka-Assadi M, Coulson C, Reid A, Proops D. Feasibility study of a hand guided robotic drill for cochleostomy. Biomed Res Int. 2014;2014:7.
Figure 4
Figure 4
Drilling cochleostomy using hand-guided robotic drill on cadaver.
Figure 5
Figure 5
The finished cochleostomy using both hand-guided robotic drill and conventional surgical drill with endosteal membrane.
Figure 6
Figure 6
A typical force and torque transient during the drilling.

References

    1. Drake JM, Joy M, Goldenberg A, Kreindler D. Computer and robotic assisted resection of brain tumours. Proceedings of the Fifth International Conference on Advanced Robotics; Pisa: ICAR; 1991. pp. 888–892.
    1. Taylor RH, Paul HA, Mittelstadt BD, et al. An image based robotic system for hip replacement surgery. J Robot Soc Jpn. 1990;8(5):111–116.
    1. Guthart GS, Salisbury JK. The intuitive telesurgery system: overview and application. IEEE International Conference on Robotics and Automation (ICRA ‘00); San Francisco, California. 2000. pp. 618–621.
    1. Jakopec M, Rodriguez y Baena F, Harris SJ, et al. The hands-on orthopaedic robot ‘acrobot’: early clinical trials of total knee replacement surgery. IEEE Trans Robot Autom. 2003;19(5):902–911.
    1. Lonner JH, John TK, Conditt MA. Robotic arm-assisted UKA improves tibial component alignment: a pilot study. Clin Orthop Relat Res. 2010;468(1):141–146. - PMC - PubMed

LinkOut - more resources