Optimal beam quality for chest flat panel detector system: realistic phantom study
- PMID: 30737566
- DOI: 10.1007/s00330-019-5998-1
Optimal beam quality for chest flat panel detector system: realistic phantom study
Abstract
Objective: To investigate optimal beam quality for chest flat panel detector (FPD) system by semi-quantitatively assessment using a realistic lung phantom.
Materials and methods: Chest FPD radiographs were obtained on a realistic lung phantom with simulated lung opacities using various X-ray tube voltage levels (90-140 kV) with/without copper filter. Entrance skin dose was set to maintain identical for all images (0.1 mGy). Three chest radiologists unaware of the exposure settings independently evaluated the image quality of each simulated opacity and normal structure using a 5-point scale (+ 2: clearly superior to the standard; + 1: slightly superior to the standard; 0: equal to the standard; - 1: slightly inferior to the standard; - 2: clearly inferior to the standard). The traditional FPD image obtained at a tube voltage of 120 kV was used as the standard. The scores of image quality were statistically compared using the Wilcoxon rank test with Bonferroni correction.
Results: FPD images using 90-kV shot with copper filter were superior to the traditional 120-kV shot without filter with respect to the visibility of vertebra, pulmonary vessels, and nodules overlapping diaphragm and heart (p < 0.05). There was no significant difference with respect to the visibility of all other simulated lung opacities (lung nodules except for overlying diaphragm/heart and honeycomb opacity) between each tube voltage level with/without copper filter and the traditional 120-kV shot without filter.
Conclusion: Image quality of FPD images using 90 kV with copper filtration is superior to that using standard tube voltage when dose is identical.
Key points: • FPD image quality using 90 kV with filter is superior to that using traditional beam. • Ninety-kilovolt shot with copper filter may be suitable for chest FPD image. • Clinical study dealing with chest FPD beam optimization would be warranted.
Keywords: Chest radiography; Digital radiography; Dose; Image quality; Physics.
Similar articles
-
[Optimal Beam Quality in Chest Radiography Using CsI-flat Panel Detector for Detection of Pulmonary Nodules].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2021;77(4):335-343. doi: 10.6009/jjrt.2021_JSRT_77.4.335. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2021. PMID: 33883367 Japanese.
-
Chest radiography with a flat-panel detector: image quality with dose reduction after copper filtration.Radiology. 2005 Nov;237(2):691-700. doi: 10.1148/radiol.2372041738. Epub 2005 Sep 28. Radiology. 2005. PMID: 16192324
-
Optimal beam quality for chest computed radiography.Invest Radiol. 1996 Mar;31(3):126-31. doi: 10.1097/00004424-199603000-00002. Invest Radiol. 1996. PMID: 8675419
-
Radiation dose and image quality in diagnostic radiology. Optimization of the dose-image quality relationship with clinical experience from scoliosis radiography, coronary intervention and a flat-panel digital detector.Acta Radiol Suppl. 2002 Mar;43(427):1-43. Acta Radiol Suppl. 2002. PMID: 12108231 Review.
-
Dose reduction in skeletal and chest radiography using a large-area flat-panel detector based on amorphous silicon and thallium-doped cesium iodide: technical background, basic image quality parameters, and review of the literature.Eur Radiol. 2004 May;14(5):827-34. doi: 10.1007/s00330-004-2243-2. Epub 2004 Feb 17. Eur Radiol. 2004. PMID: 14968260 Review.
Cited by
-
Influence of the use of various imaging units and projections on the radiation dose received by children during chest digital radiography.PLoS One. 2021 Aug 5;16(8):e0255749. doi: 10.1371/journal.pone.0255749. eCollection 2021. PLoS One. 2021. Retraction in: PLoS One. 2025 May 9;20(5):e0324184. doi: 10.1371/journal.pone.0324184. PMID: 34352022 Free PMC article. Retracted.
-
Comparison of organ dose from chest radiography with varying beam quality and constant exposure index.Phys Eng Sci Med. 2023 Sep;46(3):1143-1151. doi: 10.1007/s13246-023-01281-0. Epub 2023 May 28. Phys Eng Sci Med. 2023. PMID: 37245193
-
An evaluation of the effect of tube potential on clinical image quality using direct digital detectors for pelvis and lumbar spine radiographs.J Med Radiat Sci. 2020 Dec;67(4):260-268. doi: 10.1002/jmrs.403. Epub 2020 Jun 3. J Med Radiat Sci. 2020. PMID: 32495517 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical