First-in-human retrieval of chronically implanted Micra transcatheter pacing system
- PMID: 30758053
- DOI: 10.1111/pace.13622
First-in-human retrieval of chronically implanted Micra transcatheter pacing system
Abstract
We report the first-in-human retrieval of a chronically implanted Micra transcatheter pacing system (TPS; Medtronic, Minneapolis, MN, USA). A 41-year-old woman suffering from third-degree atrioventricular block was successfully implanted with a TPS in the low septum of the right ventricle. After 29 months, due to an increased electrical threshold and 100% pacing percentage, the device showed low battery voltage. Despite the long implant duration, the attempt of device retrieval using a snare loop inserted in the delivery system of a TPS was successful. In the same procedure, a new leadless pacemaker was implanted in the high right ventricular septum with optimal pacing threshold.
Keywords: end-of-life management; leadless pacemaker; leadless pacemaker retrieval; micra transcatheter pacing system.
© 2019 Wiley Periodicals, Inc.
Similar articles
-
Transcatheter non-acute retrieval of the tine-based leadless ventricular pacemaker.Europace. 2024 Oct 3;26(10):euae256. doi: 10.1093/europace/euae256. Europace. 2024. PMID: 39374848 Free PMC article.
-
Techniques for successful early retrieval of the Micra transcatheter pacing system: A worldwide experience.Heart Rhythm. 2018 Jun;15(6):841-846. doi: 10.1016/j.hrthm.2018.02.008. Epub 2018 Feb 8. Heart Rhythm. 2018. PMID: 29427820
-
Accelerometer-based atrioventricular synchronous pacing with a ventricular leadless pacemaker: Results from the Micra atrioventricular feasibility studies.Heart Rhythm. 2018 Sep;15(9):1363-1371. doi: 10.1016/j.hrthm.2018.05.004. Epub 2018 May 11. Heart Rhythm. 2018. PMID: 29758405
-
Design and evaluation of the Micra Transcatheter Pacing System for bradyarrhythmia management.Future Cardiol. 2019 Jan;15(1):9-15. doi: 10.2217/fca-2018-0077. Epub 2018 Dec 5. Future Cardiol. 2019. PMID: 30516083 Review.
-
Micra Extraction Out To 4.5 Years.Card Electrophysiol Clin. 2024 Jun;16(2):149-155. doi: 10.1016/j.ccep.2023.10.015. Epub 2023 Nov 11. Card Electrophysiol Clin. 2024. PMID: 38749634 Review.
Cited by
-
Extraction of a leadless pacemaker 28 months after implantation owing to right ventricular outflow tract obstruction.HeartRhythm Case Rep. 2023 Nov 13;10(2):124-127. doi: 10.1016/j.hrcr.2023.11.002. eCollection 2024 Feb. HeartRhythm Case Rep. 2023. PMID: 38404981 Free PMC article. No abstract available.
-
Cardiac Implantable Electronic Devices Breakthrough: Are We Ready to Face the Future?J Clin Med. 2022 Oct 26;11(21):6321. doi: 10.3390/jcm11216321. J Clin Med. 2022. PMID: 36362549 Free PMC article.
-
Micra leadless pacemaker retrieval for broken tines: a case report.Clin Res Cardiol. 2022 Nov;111(11):1295-1298. doi: 10.1007/s00392-022-02025-y. Epub 2022 Apr 13. Clin Res Cardiol. 2022. PMID: 35419628 No abstract available.
-
Transcatheter non-acute retrieval of the tine-based leadless ventricular pacemaker.Europace. 2024 Oct 3;26(10):euae256. doi: 10.1093/europace/euae256. Europace. 2024. PMID: 39374848 Free PMC article.
-
Intraoperative sensing increase predicts long-term pacing threshold in leadless pacemakers.J Interv Card Electrophysiol. 2022 Apr;63(3):679-686. doi: 10.1007/s10840-021-01111-x. Epub 2022 Jan 4. J Interv Card Electrophysiol. 2022. PMID: 34981291
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical