First-in-human retrieval of chronically implanted Micra transcatheter pacing system
- PMID: 30758053
- DOI: 10.1111/pace.13622
First-in-human retrieval of chronically implanted Micra transcatheter pacing system
Abstract
We report the first-in-human retrieval of a chronically implanted Micra transcatheter pacing system (TPS; Medtronic, Minneapolis, MN, USA). A 41-year-old woman suffering from third-degree atrioventricular block was successfully implanted with a TPS in the low septum of the right ventricle. After 29 months, due to an increased electrical threshold and 100% pacing percentage, the device showed low battery voltage. Despite the long implant duration, the attempt of device retrieval using a snare loop inserted in the delivery system of a TPS was successful. In the same procedure, a new leadless pacemaker was implanted in the high right ventricular septum with optimal pacing threshold.
Keywords: end-of-life management; leadless pacemaker; leadless pacemaker retrieval; micra transcatheter pacing system.