A 3-year-long investigation of the authenticity of the return results of hepatitis B virus qualitative testing in external quality assessment in East China
- PMID: 30761566
- DOI: 10.1002/jmv.25431
A 3-year-long investigation of the authenticity of the return results of hepatitis B virus qualitative testing in external quality assessment in East China
Abstract
Background: In external quality assessments (EQAs), it is important and necessary for participants to return authentic test results. However, to obtain better evaluation results, some laboratories have lacked confidence that their own and have preferred to rely upon the test results in surrounding laboratories. These violations covered up errors during testing and affected the authenticity of the returned data. The occurrence of violations in EQA launched by different EQA organizers is still unknown because these violations were carried out in private. In this paper, we adjusted the hepatitis B virus (HBV) qualitative EQA schemes to uncover some of the violations by using specially designed sample combinations in four EQA surveys to show the authenticity and accuracy of HBV testing in some prefectures of East China.
Methods: Four HBV qualitative EQA surveys were selected from 2016 to 2018, which were named 201602, 201701, 201702, and 201801, and a total of 474 laboratories were included. In the first two EQA surveys, the same EQA sample combinations were distributed. In the last two EQA surveys, three kinds of different sample combinations were designed and distributed with specific plans. The result of adding different sample combinations was that each laboratory only had a 33.3% chance of receiving the same sample combinations as those received by the surrounding laboratories. In addition, the laboratories were not aware that those samples with the same serial numbers might have different concentrations and target values. The sample concentrations, mode of delivery, requirements, results reports, and evaluation criteria were all the same as those of the first two EQA surveys.
Results: The number of laboratories with failed tests increased from fourfold to sevenfold between the first two and the last two EQA surveys. The number of failed tests in 201702 was eight times that in 201701 during the same year. Six (6 of 469, 1.28%) and three (3 of 472, 0.64%) poor performer laboratories (PPLabs) appeared in 201602 and 201701, respectively, while the number of those laboratories increased to 29 (29 of 474, 6.12%) in 201702 and 201801. Failed tests from PPLabs accounted for 70.5% (201702) and 68.7% (201801) of the total failed tests. The increase in the number of PPLabs was not universal but was concentrated in some prefectures.
Conclusions: The current EQA results for HBV qualitative testing were not as good as originally anticipated. Violations during the EQA surveys caused by exchanging test results with other laboratories and by modifying results before returning still occurred in some prefectures of East China. The laboratories that were in violation can be partly exposed as PPLabs by providing different kinds of sample combinations. Through such an EQA adjustment, the laboratory would have to re-establish confidence in its own testing results from HBV EQA samples. Then, evaluations of HBV qualitative testing samples in EQA would be more authentic and accurate in East China.
Keywords: authenticity; external quality assessment; qualitative testing; results exchange; target value; violations.
© 2019 Wiley Periodicals, Inc.
Similar articles
-
A 10-year human hepatitis B virus nucleic test external quality assessment in China: continual improvement.Clin Chim Acta. 2013 Oct 21;425:139-47. doi: 10.1016/j.cca.2013.07.026. Epub 2013 Aug 2. Clin Chim Acta. 2013. PMID: 23916788
-
[Evaluation of 2015-2016 MOTAKK HBV DNA and HCV RNA external quality assessment national program results].Mikrobiyol Bul. 2018 Oct;52(4):348-366. doi: 10.5578/mb.67081. Mikrobiyol Bul. 2018. PMID: 30522421 Turkish.
-
Interpretative comments and reference ranges in EQA programs as a tool for improving laboratory appropriateness and effectiveness.Clin Chim Acta. 2003 Jul 15;333(2):209-19. doi: 10.1016/s0009-8981(03)00188-8. Clin Chim Acta. 2003. PMID: 12849907 Review.
-
External quality assessment (EQA) and alternative assessment procedures (AAPs) in molecular diagnostics: findings of an international survey.Clin Chem Lab Med. 2020 May 8;59(2):301-306. doi: 10.1515/cclm-2020-0101. Clin Chem Lab Med. 2020. PMID: 32383687
-
Proficiency testing/external quality assessment: current challenges and future directions.Clin Chem. 2011 Dec;57(12):1670-80. doi: 10.1373/clinchem.2011.168641. Epub 2011 Sep 30. Clin Chem. 2011. PMID: 21965556 Review.
Cited by
-
Analytical comparison between two hematological analyzer systems: Mindray BC-5180 vs Sysmex XN-1000.J Clin Lab Anal. 2019 Oct;33(8):e22955. doi: 10.1002/jcla.22955. Epub 2019 Jun 20. J Clin Lab Anal. 2019. PMID: 31218736 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical