Combined Conventional and Amplitude-Integrated EEG Monitoring in Neonates: A Prospective Study

Affiliations

Combined Conventional and Amplitude-Integrated EEG Monitoring in Neonates: A Prospective Study

Sarah Grace Buttle et al. J Child Neurol. 2019 May.

Abstract

Background/objective: Seizure monitoring via amplitude-integrated EEG is standard of care in many neonatal intensive care units; however, conventional EEG is the gold standard for seizure detection. We compared the diagnostic yield of amplitude-integrated EEG interpreted at the bedside, amplitude-integrated EEG interpreted by an expert, and conventional EEG.

Methods: Neonates requiring seizure monitoring received amplitude-integrated EEG and conventional EEG in parallel. Clinical events and amplitude-integrated EEG were interpreted at bedside. Subsequently, amplitude-integrated EEG and conventional EEG were independently analyzed by experienced neonatology and neurology readers. Sensitivity and specificity of bedside amplitude-integrated EEG as compared to expert amplitude-integrated EEG interpretation and conventional EEG were evaluated.

Results: Thirteen neonates were monitored for an average duration of 33 hours (range 15-94, SD 25). Fourteen seizure-like events were detected by clinical observation, and 12 others by bedside amplitude-integrated EEG analysis. One of the clinical, and none of the bedside amplitude-integrated EEG events were confirmed as seizures on conventional EEG. Post hoc expert amplitude-integrated EEG interpretation revealed eight suspected seizures, all different from the ones detected by the bedside amplitude-integrated EEG team, of which one was confirmed via conventional EEG. Eight seizures were recorded on conventional EEG. Expert amplitude-integrated EEG interpretation had a sensitivity of 13% with 46% specificity for individual seizure detection, and a sensitivity of 50% with 46% specificity for detecting patients with seizures.

Conclusion: Real-world bedside amplitude-integrated EEG monitoring failed to detect all seizures evidenced via conventional EEG, while misclassifying other events as seizures. Even post hoc expert amplitude-integrated EEG interpretation provided limited sensitivity and specificity. Considering the poor sensitivity and specificity of bedside amplitude-integrated EEG interpretation, combined monitoring may provide limited clinical benefit.

Keywords: diagnostic accuracy; electroencephalography (EEG); neonatal neurology; neurocritical care; newborn.