Results of 200 Consecutive Ceramic-on-Ceramic Cementless Hip Arthroplasties in Patients Up To 50 Years of Age: A 5-24 Years of Follow-Up Study
- PMID: 30777621
- DOI: 10.1016/j.arth.2019.01.057
Results of 200 Consecutive Ceramic-on-Ceramic Cementless Hip Arthroplasties in Patients Up To 50 Years of Age: A 5-24 Years of Follow-Up Study
Abstract
Background: Ceramic-on-ceramic (CoC) bearings due to their mechanical properties can be used in primary total hip arthroplasty (THA) especially in young patients requiring a long-lasting implant. The goal of this study is to analyze the results of a series of 200 consecutive CoC THAs in patients aged 50 years or less at the time of surgery.
Methods: A retrospective study was conducted on the first 200 consecutive CoC arthroplasties performed using the direct lateral approach on 105 females and 81 males (14 bilateral cases) with an average age of 44.2 (16-50) years. The diagnosis was primary or post-traumatic osteoarthritis in 94 cases, avascular necrosis of the femoral head in 47, displaced intracapsular femoral neck fracture in 29, osteoarthritis secondary to developmental dysplasia of the hip/Legg-Calvè-Perthes disease/slipped capital femoral epiphysis in 20, and rheumatic diseases in 10 cases. The preoperative Harris Hip Score was 32.5 on average (range 15-55). All the implants were cementless. In 177 THAs the coupling was alumina-on-alumina, and in 23 cases the coupling was AMC-on-AMC (alumina matrix composite).
Results: Twenty-five patients with 28 THAs were lost at the final follow-up, 2 stems were revised due to subsidence, 1 cup was revised due to malposition, 1 femoral head was changed because of impingement, and 1 THA was revised for periprosthetic infection. Three patients sustained a Vancouver B1 periprosthetic femoral fracture. At the final follow-up (mean 14.9 years; minimum 5 years to maximum 24 years), 172 THAs were eligible for clinical and radiographic evaluation: none was revised for wear and/or breakage of the ceramic components. Harris Hip Score rose up to a mean value of 90.1 (52-100).
Conclusion: The present report demonstrates that CoC coupling offers excellent long-term results in THA performed in young patients with very low wear and no adverse effects caused by the material.
Keywords: ceramic-on-ceramic; total hip arthroplasty; total hip replacement; tribology; wear.
Copyright © 2019 Elsevier Inc. All rights reserved.
Similar articles
-
A minimum 20-year outcome of 100 consecutive alumina-on-alumina arthroplasties performed by a single surgeon.Hip Int. 2018 Nov;28(2_suppl):10-14. doi: 10.1177/1120700018813218. Hip Int. 2018. PMID: 30755120
-
Quarter Century Outcomes of Alumina Ceramic-on-Ceramic Total Hip Arthroplasty.J Arthroplasty. 2025 Apr;40(4):992-998. doi: 10.1016/j.arth.2024.10.009. Epub 2024 Oct 16. J Arthroplasty. 2025. PMID: 39419413
-
What Is the Frequency of Fracture of Ceramic Components in THA? Results from the Norwegian Arthroplasty Register from 1997 to 2017.Clin Orthop Relat Res. 2020 Jun;478(6):1254-1261. doi: 10.1097/CORR.0000000000001272. Clin Orthop Relat Res. 2020. PMID: 32348085 Free PMC article.
-
Low Reinfection Rates But a High Rate of Complications in THA for Infection Sequelae in Childhood: A Systematic Review.Clin Orthop Relat Res. 2021 May 1;479(5):1094-1108. doi: 10.1097/CORR.0000000000001607. Clin Orthop Relat Res. 2021. PMID: 33617159 Free PMC article.
-
Outcomes of different bearings in total hip arthroplasty - implant survival, revision causes, and patient-reported outcome.Dan Med J. 2017 Mar;64(3):B5350. Dan Med J. 2017. PMID: 28260601 Review.
Cited by
-
The Wear Rate and Survivorship in Total Hip Arthroplasty Using a Third-generation Ceramic Head on a Conventional Polyethylene Liner: A Minimum of 15-year Follow-up.Hip Pelvis. 2022 Jun;34(2):115-121. doi: 10.5371/hp.2022.34.2.115. Epub 2022 Jun 7. Hip Pelvis. 2022. PMID: 35800129 Free PMC article.
-
Global diversity in bearings in primary THA.EFORT Open Rev. 2020 Oct 26;5(10):763-775. doi: 10.1302/2058-5241.5.200002. eCollection 2020 Oct. EFORT Open Rev. 2020. PMID: 33204520 Free PMC article. Review.
-
Finite Element Analysis of Acetabulum Prosthesis' Lining Damage Zone with Different Implanting Angle.Comput Math Methods Med. 2023 Jun 2;2023:9350474. doi: 10.1155/2023/9350474. eCollection 2023. Comput Math Methods Med. 2023. PMID: 37304323 Free PMC article.
-
The race for the classification of proximal periprosthetic femoral fractures : Vancouver vs Unified Classification System (UCS) - a systematic review.BMC Musculoskelet Disord. 2022 Mar 23;23(1):280. doi: 10.1186/s12891-022-05240-w. BMC Musculoskelet Disord. 2022. PMID: 35321671 Free PMC article.
-
Uncemented versus cemented total hip arthroplasty for displaced femoral neck fractures in elderly patients with osteoporosis: A retrospective analysis.J Int Med Res. 2020 Aug;48(8):300060520944663. doi: 10.1177/0300060520944663. J Int Med Res. 2020. PMID: 32772652 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical