Aluminium release by coated and uncoated fluid-warming devices
- PMID: 30793280
- DOI: 10.1111/anae.14601
Aluminium release by coated and uncoated fluid-warming devices
Abstract
The use of fluid-warming systems is recommended for infusion rates > 500 ml.h-1 to avoid peri-operative hypothermia. Some fluid-warming devices use disposable aluminium-heated plates for heat transfer, but there is no protective coating to separate the fluid from the heated aluminium surface. It is unknown if this could promote release of aluminium into infusion fluids. We investigated a coated (Fluido compact) and an uncoated (enFlow) fluid-warming device using normal saline or balanced electrolyte solution as infusion fluids, pumped through the heated disposables at flow rates of 2, 4 and 8 ml.min-1 for 60 min each. Aluminium concentrations in the fluid samples were analysed using graphite furnace atomic absorption spectrometry. With saline the coated and uncoated devices yielded aluminium concentrations below the level of quantification (< 128 μg.l-1 ). Similarly, balanced electrolyte solution in the coated device yielded aluminium concentrations < 128 μg.l-1 . However, balanced electrolyte solution in the uncoated device yielded aluminium concentrations of up to 6794 (3465-8002 [1868-7421]) μg.l-1 . Repeating this last study at a flow rate of 2 ml.min-1 resulted in quite high aluminium concentrations when the uncoated device was not heated (~1000 μg.l-1 ) and higher concentrations after the device was heated. We conclude that using uncoated aluminium plates in fluid-warming systems can lead to a risk of administering potentially harmful concentrations of aluminium when balanced crystalloid solutions are used. The mechanism is unclear, but heat is in part involved. Coating for aluminium within medical devices in direct contact with infusion fluids should be recommended.
Keywords: fluid warming; hypothermia: prevention; paediatrics: warming techniques.
© 2019 Association of Anaesthetists.
Comment in
-
Medical device regulation: the need for clinical vigilance and oversight.Anaesthesia. 2019 Jun;74(6):693-695. doi: 10.1111/anae.14603. Epub 2019 Feb 21. Anaesthesia. 2019. PMID: 30793284 No abstract available.
-
Aluminium release during fluid warming.Anaesthesia. 2019 Jun;74(6):819. doi: 10.1111/anae.14673. Anaesthesia. 2019. PMID: 31063203 No abstract available.
-
An experimental comparison of two methods for warming intravenous crystalloid solutions.Anaesthesia. 2019 Jul;74(7):946. doi: 10.1111/anae.14708. Anaesthesia. 2019. PMID: 31168804 No abstract available.
-
Quantified aluminium levels released into blood and fluids using the Level 1 Fast Flow Fluid Warmer.Anaesthesia. 2020 Feb;75(2):271-272. doi: 10.1111/anae.14881. Anaesthesia. 2020. PMID: 31916604 No abstract available.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous