Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2019 Nov;66(11):3014-3025.
doi: 10.1109/TBME.2019.2899762. Epub 2019 Feb 15.

Portable System for Time-Domain Diffuse Correlation Spectroscopy

Portable System for Time-Domain Diffuse Correlation Spectroscopy

Davide Tamborini et al. IEEE Trans Biomed Eng. 2019 Nov.

Erratum in

Abstract

We introduce a portable system for clinical studies based on time-domain diffuse correlation spectroscopy (DCS). After evaluating different lasers and detectors, the final system is based on a pulsed laser with about 550 ps pulsewidth, a coherence length of 38 mm, and two types of single-photon avalanche diodes (SPAD). The higher efficiency of the red-enhanced SPAD maximizes detection of the collected light, increasing the signal-to-noise ratio, while the better timing response of the CMOS SPAD optimizes the selection of late photons and increases spatial resolution. We discuss component selection and performance, and we present a full characterization of the system, measurement stability, a phantom-based validation study, and preliminary in vivo results collected from the forearms and the foreheads of four healthy subjects. With this system, we are able to resolve blood flow changes 1 cm below the skin surface with improved depth sensitivity and spatial resolution with respect to continuous wave DCS.

PubMed Disclaimer

Figures

Fig. 1.
Fig. 1.
Example of pathlength distributions (dashed areas) of the photons at a sampling time of 1.5 ns. The optical properties used for the sample response function (SRF) are μa = 0.05 and μs’ = 10 cm−1. (a) Two Gaussian IRFs with 100 ps and 500 ps FWHM, centered on ts, and the corresponding pathlength distributions. (b) 500 ps Gaussian IRF with the addition of diffusion tails representative of two different detectors and the corresponding pathlength distributions. The IRFs are flipped so that the total delay (sum of the ToF and emission time relative to IRF peak) equals ts to for the photons in the distribution. Note that the longer tail causes increased contribution of photons with short ToF to the pathlength distribution at ts.
Fig. 2.
Fig. 2.
Block diagram of the system setup, where three MPD-RE detectors are used to measure the IRF and the light collected at 1 and 2 cm away from the source. A MPD-FG module collects the light at a 1 cm separation from the source. For the laser source, we chose the Vis-IR 500 because of the longer coherence length. A correlator board acquires the ToF measured by TDC-cards and pairs this information with the absolute arrival time of the photon, then transfers the data to a PC via USB 3.0. (b) A photo of the system, showing an enlarged view of the optical probe (the probe has also a 15 mm separation, not used in this work).
Fig. 3.
Fig. 3.
Autocorrelation functions (g2) of the CW equivalent photons (i.e. using the whole TPSF) when using a 12.5 MHz laser repetition rate (fL), 150 MHz correlator sampling frequency (fS), and 1 cm source-detector separation. For these measurements, we used a silicone oil liquid phantom with a very slow decay (in the 100 μs to 1 ms range) to achieve a constant pattern in the μs range and better visualize the oscillations. The g2 curve computed without any correction shows oscillations which are due to the fact fL is not equal to nor an integer multiple of fS. By applying either a jitter or a resampling correction the oscillations are completely removed.
Fig. 4.
Fig. 4.
Pulse width comparison (a) and g2 comparison (b) of the four lasers tested. The g2s are evaluated on a silicone oil liquid phantom at 1 cm source-detector separation, while selecting the photons in a gate of 60 ps centered at the TPSF peak. For narrower laser pulses, the g2 are much lower because of the shorter coherence length.
Fig. 5.
Fig. 5.
Beta measured at different gates centered on the TPSF peak. We changed the gate width from 10 to 500 ps (roughly the FWHM of the VisIR-500 laser). The beta drop changes slope around 180 ps, that corresponds to the coherence length of the laser (~40 mm).
Fig. 6.
Fig. 6.
IRF comparison using the two detectors. The MPD-FG IRF (green) has a fast extinction time of 585 ps to drop two orders of magnitude, while the MPD-RE IRF (blue) takes about 2.28 ns. A measurement of the VISIR-70 and VISIR-500 lasers (blue vs red curve) both acquired with the MPD-RE shows the effect of the source width on the long diffusion tail. For an increased pulse width, the diffusion tail contribution starts higher (at −13 dB instead of −18 dB).
Fig. 7.
Fig. 7.
Normalized count rate (black, left y-axis), IRF peak position (blue, right y-axis), and IRF FWHM (green, right y-axis) for a long acquisition time. The system stabilizes after about 25 minutes of warm-up time.
Fig. 8.
Fig. 8.
Results of the two-layer experiment. On the TD-DCS side, the g2 of the early (EP, 240 to 120 ps before the TPSF peak) and late (LP, 360 to 720 ps after the TPSF peak) photons show a large decay time difference between the Intralipid (IL) and the silicone oil (SO) phantoms, while the normalized TPSFs show minor differences due to the small difference in optical properties. For all detectors, the g2 of the 20 mm top-layer is close to the silicone oil one. When selecting the late photons, the g2 of the 0.5 and 1 cm top-layer phantoms show a decay closer to the Intralipid. On the CW-DCS side, the difference in the g2 decay between intralipid and silicone oil is evident at all source-detector separation. For the two-layer phantoms, the g2 show a decay closer to the one of the bottom layer whenthe top layer is thin and the source-detector separation is larger.
Fig. 9.
Fig. 9.
Comparison of the normalized BFi at different thickness between TD- and CW-DCS. TD-DCS with LP at 2 cm separation shows the maximum sensitivity to depth, while LP results at 1 cm are comparable to CW-DCS at 2 cm. The sensitivity is good up to 15 mm layer thickness, and becomes poor at 20 mm.
Fig. 10.
Fig. 10.
Comparison of the normalized BFi (average and standard deviation obtained by splitting the measurement in 60 parts of 10 s each) at different thickness between the MPD-FG (a) and the MPD-RE (b) detectors at 1 cm and the MPD-RE at 2 cm (c) separation when pushing the selection of photons to late gates, up to 1 ns after the TPSF peak. The sensitivity of the MPD-FG increases while the MPD-RE at 1 cm starts losing sensitivity at thin layer due to increasing contribution of the diffusion tail. The MPD-RE at 2 cm is less penalized by the detection tail thanks to the slower decay of the TPSF at larger source-detector separation, but the sensitivity at very-late and ultra-late gates (VLP and ULP) is not improved as in the MPD-FG detector.
Fig. 11.
Fig. 11.
Results of the flow experiment when the source-detector separation is 1 cm (a) and 2 and 3 cm (b). The highest contrast is achieved when the tube is centered between source and detector: x = 5 mm (a), x = 10 mm (b) and x = 15 mm (b, orange curve). At 1 cm source-detector separation, the flow is not detected by the early photons (EP), while the maximum flow contrast is obtainable when selecting the late photons (LP) and using the short tail detector (MPD-FG, brown line). Similar results are obtained at 2 cm source-detection separation. CW-DCS at 3 cm (orange) provides lower contrast and spatial resolution than LP TD-DCS.
Fig. 12.
Fig. 12.
Results of the cuff occlusion experiment. Cuff pressure shown in panel (a), blood flow index when selecting early photons (b) and late photons (c) at 1 cm source-detector separation. In both cases, and with both detectors, we see BFi drop to almost zero during ischemia. At late gates, the higher efficiency of the MPD-RE (blue curve) results in less noise than the MPD-FG (brown curve). The bold lines are means and faint lines the standard deviations of eight trials in total.
Fig. 13.
Fig. 13.
Relative BFi changes at 1 cm source-detector separation using the MPD-RE (blue) and the MPD-FG (brown) before and during the application of light manual pressure on the optical probe placed on the forehead. By using the early photons, the BFi drops by 31% after applying the pressure, as expected. By using the late photons, the MPD-FG shows a small drop of 7%, while the MPD-RE shows a drop of 22% due to the tail contribution.

References

    1. Dagal A and Lam AM, “Cerebral blood flow and the injured brain: how should we monitor and manipulate it?,” Current Opinion in Anaesthesiology, vol. 24, no. 2, pp. 131–137 (2011). - PubMed
    1. Boas DA and Franceschini MA, “Haemoglobin oxygen saturation as a biomarker: the problem and a solution,” Philos. Trans. Royal Soc. A, vol. 369, no. 1955, pp. 4407–4424 (2011). - PMC - PubMed
    1. Yoxall CW and Weindling AM, “Measurement of cerebral oxygen consumption in the human neonate using near infrared spectroscopy: cerebral oxygen consumption increases with advancing gestational age,” Pediatric Research, vol. 44, pp. 283–290 (1998). - PubMed
    1. Grubb RL, Raichle ME, Eichling JO et al., “The effects of changes in PaCO2 on cerebral blood volume, blood flow, and vascular mean transit time,” Stroke, vol. 5, no. 5, pp. 630–639 (1974). - PubMed
    1. Jones M, Berwick J and Mayhew J, “Changes in blood flow, oxygenation, and volume following extended stimulation of rodent barrel cortex,” Neuroimage, vol. 15, no. 3, pp. 474–487 (2002). - PubMed

Publication types

MeSH terms