Performance of the BD-FACS Presto for CD4 count and hemoglobin measurement in a district hospital and rural laboratory in Ghana
- PMID: 30794637
- PMCID: PMC6386386
- DOI: 10.1371/journal.pone.0212684
Performance of the BD-FACS Presto for CD4 count and hemoglobin measurement in a district hospital and rural laboratory in Ghana
Abstract
Introduction: In Ghana, initiation of Antiretroviral Therapy (ART) is recommended for all patients with an HIV diagnosis, regardless of CD4+ T-cell count. However, measurement of CD4 count remains an important metric for identifying patients with advanced HIV disease, and assessing a person's overall immune status, which informs the decision to offer opportunistic infection screening and prophylaxis. Access to CD4+ T cell count in rural health facilities remains a major challenge in Ghana and other resource-limited settings. This study aimed to validate the accuracy of the BD FACSPresto near-patient device for measurement of CD4 count and hemoglobin concentration against the FACSCount (CD4) and Sysmex (hemoglobin) diagnostic machines when operated in both a district hospital and rural laboratory, serving a network of health posts in Ashanti Region, Ghana.
Methodology: In the first phase of the study, patients were recruited from a district hospital, and both venous and capillary blood samples were tested using the FACSCount and Sysmex as reference tests and compared to results of the FACSPresto performed in the clinic laboratory at the district hospital. In the second phase, patients were recruited from both the hospital and from rural health clinics, and samples were tested using the FACSPresto at a rural laboratory. Sensitivity and specificity among samples categorized into different clinically relevant CD4 count ranges were calculated, along with correlation between the Presto and the reference measurements, and mean and relative bias with limits of agreement.
Results: The FACSPresto was successfully operated in both clinical settings. A total of 59 samples in the first phase and 48 samples in the second phase were included. Positive bias was observed when comparing CD4 count measured by BD FACSPresto to FACSCount in the district hospital (bias = 44, LOA -72,160) and in the rural laboratory setting (bias = 74, LOA -96, 244). In addition, capillary blood samples were shown to give higher measures when compared to venous blood samples from the same participant. All results were statistically significant (p<0.05) apart from hemoglobin measurement in venous blood in the rural laboratory. Correlation coefficients were high for CD4 count measures and lower for hemoglobin measures.
Conclusion: Overall, the Presto gave higher estimates of CD4 count compared to FACSCount, and hemoglobin measurements were higher than from Sysmex. Samples of capillary blood in turn gave higher results for both measurements compared to venous blood, consistent with previous analyses. These findings should be considered when selecting CD4 count machines for use at the point of care, especially in remote areas where capillary blood sampling may be preferable, but are likely balanced by device's ease of use, portability, and ability to expand access to services. These results are some of the first to demonstrate the accuracy of the FACSPresto in West Africa and show that this device can be successfully operated in a very rural lab setting and may therefore assist to provide CD4 count and hemoglobin concentration measurement to populations in need.
Conflict of interest statement
The authors have declared that no competing interests exist.
Figures


Similar articles
-
Performance evaluation of BD FACSPresto™ point of care CD4 analyzer to enumerate CD4 counts for monitoring HIV infected individuals in Nigeria.PLoS One. 2017 May 25;12(5):e0178037. doi: 10.1371/journal.pone.0178037. eCollection 2017. PLoS One. 2017. PMID: 28542359 Free PMC article.
-
Performance of the BD FACSPresto near to patient analyzer in comparison with representative conventional CD4 instruments in Cameroon.AIDS Res Ther. 2020 Aug 17;17(1):53. doi: 10.1186/s12981-020-00309-9. AIDS Res Ther. 2020. PMID: 32799909 Free PMC article.
-
The performance of BD FACSPresto™ for CD4 T-cell count, CD4% and hemoglobin concentration test in Ethiopia.PLoS One. 2017 Apr 27;12(4):e0176323. doi: 10.1371/journal.pone.0176323. eCollection 2017. PLoS One. 2017. PMID: 28448581 Free PMC article.
-
Absolute and percent CD4+ T-cell enumeration by flow cytometry using capillary blood.J Immunol Methods. 2011 Sep 30;372(1-2):1-6. doi: 10.1016/j.jim.2011.07.008. Epub 2011 Jul 20. J Immunol Methods. 2011. PMID: 21787779 Review.
-
Evaluation of a flow cytometry method for CD4 T cell enumeration based on volumetric primary CD4 gating using thermoresistant reagents.J Immunol Methods. 2011 Sep 30;372(1-2):7-13. doi: 10.1016/j.jim.2011.07.012. Epub 2011 Jul 29. J Immunol Methods. 2011. PMID: 21835181 Review.
Cited by
-
Validation of the BD FACSPresto system for the measurement of CD4 T-lymphocytes and hemoglobin concentration in HIV-negative and HIV-positive subjects.Sci Rep. 2020 Nov 11;10(1):19605. doi: 10.1038/s41598-020-76549-6. Sci Rep. 2020. PMID: 33177659 Free PMC article.
-
Geographical Access to Point-of-care diagnostic tests for diabetes, anaemia, Hepatitis B, and human immunodeficiency virus in the Bono Region, Ghana.BMC Health Serv Res. 2024 Oct 29;24(1):1303. doi: 10.1186/s12913-024-11830-2. BMC Health Serv Res. 2024. PMID: 39472915 Free PMC article.
-
Peripheral blood lymphocyte immunophenotyping (TBNK) - a comparison of BD FACSCanto II and BD FACSLyric flow cytometry analysers.Cent Eur J Immunol. 2024;49(1):45-51. doi: 10.5114/ceji.2024.135939. Epub 2024 Mar 5. Cent Eur J Immunol. 2024. PMID: 38812607 Free PMC article.
-
Accurate and reproducible enumeration of CD4 T cell counts and Hemoglobin levels using a point of care system: Comparison with conventional laboratory based testing systems in a clinical reference laboratory in Cameroon.PLoS One. 2024 Mar 20;19(3):e0297790. doi: 10.1371/journal.pone.0297790. eCollection 2024. PLoS One. 2024. PMID: 38507344 Free PMC article.
-
Comparison of Standard and Point-of-Care CD4+ T Lymphocyte Measurement Methods in HIV-1 Infected Turkish Patients.Medicina (Kaunas). 2024 Dec 21;60(12):2094. doi: 10.3390/medicina60122094. Medicina (Kaunas). 2024. PMID: 39768973 Free PMC article.
References
-
- World Health Organization. Guidelines for managing advanced HIV disease and rapid initiation of antiretroviral therapy, July 2017. 2017. - PubMed
-
- Daniel Francois Venter W, Chersich MF, Majam M, Akpomiemie G, Arulappan N, Moorhouse M, et al. CD4 cell count variability with repeat testing in South Africa: Should reporting include both absolute counts and ranges of plausible values? International journal of STD & AIDS. 2018:0956462418771768. - PubMed
-
- Hoffman S, Exner TM, Lince-Deroche N, Leu C-S, Phillip JL, Kelvin EA, et al. Immediate Blood Draw for CD4+ Cell Count Is Associated with Linkage to Care in Durban, South Africa: Findings from Pathways to Engagement in HIV Care. PLOS ONE. 2016;11(10):e0162085 10.1371/journal.pone.0162085 - DOI - PMC - PubMed
-
- Desai MA, Okal DO, Rose CE, Ndivo R, Oyaro B, Otieno FO, et al. Effect of point-of-care CD4 cell count results on linkage to care and antiretroviral initiation during a home-based HIV testing campaign: a non-blinded, cluster-randomised trial. The Lancet HIV. 2017;4(9):e393–e401. 10.1016/S2352-3018(17)30091-7. - DOI - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous