. 2019 Feb 25;9(1):2687.
doi: 10.1038/s41598-018-37478-7.

Validation of clinic-based cryptococcal antigen lateral flow assay screening in HIV-infected adults in South Africa

Affiliations

Validation of clinic-based cryptococcal antigen lateral flow assay screening in HIV-infected adults in South Africa

Paul K Drain et al. Sci Rep. .

Abstract

Since rapid cryptococcal antigen lateral flow assays (CrAg LFA) may expedite treatment of HIV-associated cryptococcal infections, we sought to validate clinic-based CrAg LFA testing. Among newly-diagnosed HIV-infected adults in South Africa, a trained nurse performed clinic-based testing of urine, fingerprick capillary and venous whole blood with rapid CrAg LFA (Immy Diagnostics, Norman, USA). We performed matched laboratory-based serum cryptococcal antigen testing with an enzyme immunoassay (EIA). We assessed diagnostic accuracy using EIA as the gold-standard, and performed additional validation testing on serum and among hospitalized adults with cryptococcal meningitis. Among 5,618 participants enrolled, 1,296 were HIV-infected and screened for cryptococcal antigenemia. Overall CrAg prevalence by serum EIA was 3.6% (95% CI 2.0-6.0%) for adults with CD4 < 200 cells/mm3, and 5.7% (95% CI 2.8-10.2%) for adults with CD4 < 100 cells/mm3. Using expanded screening guidelines (CD4 < 200 cells/mm3), CrAg LFA testing of venous whole blood, fingerprick capillary blood, and urine had diagnostic sensitivities of 46% (95% CI 19-75%), 38% (95% CI 14-68%), and 54% (95% CI 25-81%), and specificities of 97%, 97%, and 86%, respectively. When tested on serum samples, CrAg LFA had sensitivity of 93% (95% CI 66-100%) and specificity of 100% (95% CI 88-100%). All venous and fingerprick whole blood CrAg LFA tests were positive among 30 hospitalized adults with cryptococcal meningitis. Two independent readers had strong agreement for all LFA results (p < 0.0001). When performed at the point-of-care by trained nurses, CrAg LFA testing was feasible, had the highest accuracy on serum specimens, and may accelerate treatment of HIV-associated cryptococcal infections.

PubMed Disclaimer

Conflict of interest statement

The authors declare no competing interests.

Figures

Figure 1
Figure 1
Consort diagram of study enrollment and testing results.
Figure 2
Figure 2
Venn Diagram for agreement between all positive laboratory-based serum CrAg EIA results (“Lab crypto”) and point-of-care (POC) CrAg by venous and fingerprick whole blood.

References

    1. Rajasingham R, et al. Global burden of disease of HIV-associated cryptococcal meningitis: an updated analysis. Lancet Infect Dis. 2017;17:873–881. doi: 10.1016/S1473-3099(17)30243-8. - DOI - PMC - PubMed
    1. Jarvis JN, et al. Adult meningitis in a setting of high HIV and TB prevalence: findings from 4961 suspected cases. BMC Infect Dis. 2010;10:67. doi: 10.1186/1471-2334-10-67. - DOI - PMC - PubMed
    1. French N, et al. Cryptococcal infection in a cohort of HIV-1-infected Ugandan adults. AIDS. 2002;16:1031–1038. doi: 10.1097/00002030-200205030-00009. - DOI - PubMed
    1. Liechty CA, et al. Asymptomatic serum cryptococcal antigenemia and early mortality during antiretroviral therapy in rural Uganda. Trop Med Int Health. 2007;12:929–935. doi: 10.1111/j.1365-3156.2007.01874.x. - DOI - PubMed
    1. Jarvis JN, et al. Screening for cryptococcal antigenemia in patients accessing an antiretroviral treatment clinic in South Africa. Clin Infect Dis. 2009;48:856–862. doi: 10.1086/597262. - DOI - PMC - PubMed