Abstract
Purpose: The aim of the present experimental study was to evaluate the biomechanical behavior in different types of osteosynthesis (titanium screws, bioresorbable pins and miniplates) used in management of intracapsular condylar head fractures.
Method: Experimental models of the condylar head fractures were simulated on 15 dry human cadaveric mandibles. Osteotomized mandibles were randomly divided into three groups with different fixation systems used: 1) 15 mm long titanium screws, 2) 15 mm long bioresorbable pins Sonic Pins Rx, 3) T-shaped titanium miniplate and 7 mm long titanium screws. Mandibles were loaded in TIRAtest testing machine (Germany). The main types of deformations, including torsion, bending and shearing, were simulated to study the biomechanical characteristics of the fixation systems.
Results: Titanium bicortical screws demonstrated the highest stiffness in standard loading conditions. The fixation with bioresorbable pins showed lower stiffness in both frontal and sagittal loads. This is indicative of the fact that resorbable pins, which have numerous advantages for clinical usage, cannot provide adequately stable fixation in maximal masticatory loads. The mandibles fixed with T-shaped plate had the lowest stiffness.
Conclusion: Screw or pin fixation, regardless of the material used, was not resistant to rotational loads. On the contrary, the stiffness of T-shaped plates was quite significant. In real clinical conditions, if rotational displacements are not effectively compensated by irregularities in the fracture surface and precise repositioning of the bone fragments, combined use of miniplates and bicortical titanium screws or two screws can be beneficial.
Keywords: Biomechanics; Condylar head; Fracture; Osteosynthesis.
Figures





Similar articles
-
Biomechanical Loading Comparison between Titanium and Bioactive Resorbable Screw Systems for Fixation of Intracapsular Condylar Head Fractures.Materials (Basel). 2020 Jul 15;13(14):3153. doi: 10.3390/ma13143153. Materials (Basel). 2020. PMID: 32679803 Free PMC article.
-
[Stability of osteosyntheses for condylar head fractures in the clinic and biomechanical simulation].Mund Kiefer Gesichtschir. 2004 Mar;8(2):63-74. doi: 10.1007/s10006-004-0529-9. Epub 2004 Feb 6. Mund Kiefer Gesichtschir. 2004. PMID: 15045529 German.
-
Biomechanical Evaluation of Mandibular Condyle Fracture Osteosynthesis Using the Rhombic Three-Dimensional Condylar Fracture Plate.J Oral Maxillofac Surg. 2019 Sep;77(9):1868.e1-1868.e15. doi: 10.1016/j.joms.2019.04.020. Epub 2019 Apr 25. J Oral Maxillofac Surg. 2019. PMID: 31112678
-
Biomechanical comparison of osteosynthesis with poly-L-lactic acid and titanium screw in intracapsular condylar fracture fixation: An experimental study.Niger J Clin Pract. 2015 Sep-Oct;18(5):589-93. doi: 10.4103/1119-3077.158946. Niger J Clin Pract. 2015. PMID: 26096234
-
A comparison of combinations of titanium and resorbable plating systems for repair of isolated zygomatic fractures in the adult: a quantitative biomechanical study.Ann Plast Surg. 2005 Apr;54(4):402-8. doi: 10.1097/01.sap.0000151484.59846.62. Ann Plast Surg. 2005. PMID: 15785282 Review.
Cited by
-
Application of CAD/CAM technology for surgical treatment of condylar head fractures: A preliminary study.J Oral Biol Craniofac Res. 2020 Oct-Dec;10(4):608-614. doi: 10.1016/j.jobcr.2020.08.018. Epub 2020 Aug 27. J Oral Biol Craniofac Res. 2020. PMID: 32963953 Free PMC article.
-
Change in Pull-Out Force during Resorption of Magnesium Compression Screws for Osteosynthesis of Mandibular Condylar Fractures.Materials (Basel). 2021 Jan 6;14(2):237. doi: 10.3390/ma14020237. Materials (Basel). 2021. PMID: 33418924 Free PMC article.
-
Current Frequency of Mandibular Condylar Process Fractures.J Clin Med. 2023 Feb 9;12(4):1394. doi: 10.3390/jcm12041394. J Clin Med. 2023. PMID: 36835931 Free PMC article.
-
A Novel Screw Drive for Allogenic Headless Position Screws for Use in Osteosynthesis-A Finite-Element Analysis.Bioengineering (Basel). 2021 Oct 1;8(10):136. doi: 10.3390/bioengineering8100136. Bioengineering (Basel). 2021. PMID: 34677209 Free PMC article.
References
-
- Boffano P., Roccia F., Zavattero E. Maxillofacial Trauma (EURMAT) project: a multicentre and prospective study. J Cranio-Maxillo-Fac Surg. 2015;43(1):62–70. - PubMed
-
- Luo S., Li B., Long X., Deng M., Cai H., Cheng Y. Surgical treatment of sagittal fracture of mandibular condyle using long-screw osteosynthesis. J OralMaxillofacSurg. 2011;69:1988–1994. - PubMed
-
- Duan D.H., Zhang Y. A clinical investigation on disc displacement in sagittal fracture of the mandibular condyle and its association with TMJ ankylosis development. Int J Oral Maxillofac Surg. 2011;40(2):134–138. - PubMed
-
- Neff A., Muhlberger G., Karoglan M. Stability of osteosyntheses for condylar head fractures in the clinic and biomechanical simulation. Mund Kiefer Gesichtschir. 2004;8(2):63–74. - PubMed
LinkOut - more resources
Full Text Sources