Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Multicenter Study
. 2019 Oct 1;56(4):793-799.
doi: 10.1093/ejcts/ezz055.

Minimally invasive aortic valve replacement with sutureless and rapid deployment valves: a report from an international registry (Sutureless and Rapid Deployment International Registry)†

Affiliations
Free article
Multicenter Study

Minimally invasive aortic valve replacement with sutureless and rapid deployment valves: a report from an international registry (Sutureless and Rapid Deployment International Registry)†

Paolo Berretta et al. Eur J Cardiothorac Surg. .
Free article

Abstract

Objectives: The impact of sutureless and rapid deployment (SURD) valves on the clinical outcomes of patients undergoing minimally invasive aortic valve replacement (MI-AVR) has still to be defined. The aim of this study was to assess clinical characteristics and in-hospital results of patients receiving SURD-AVR through less invasive approaches in the large population of the Sutureless and Rapid Deployment International Registry (SURD-IR).

Methods: Of the 1935 patients who received primary isolated SURD-AVR between 2009 and 2018, a total of 1418 (73.3%) underwent MI interventions and were included in this analysis. SURD-AVR was performed using upper ministernotomy in 56.4% (n = 800) of cases and anterior right thoracotomy in 43.6% (n = 618). Perceval S was implanted in 1011 (71.3%) patients and Edwards Intuity or Intuity Elite in 407 (28.7%) patients.

Results: Overall in-hospital mortality and stroke rates were 1.7% and 2%, respectively. A definitive pacemaker implantation was reported in 9% of cases and significantly decreased over the observational period, from 20.6% to 5.6% (P = 0.002). The Perceval valve was associated with shorter operative times and was more frequently implanted in patients receiving anterior right thoracotomy incision. The Intuity valve was preferred in younger patients and revealed superior postoperative haemodynamic results.

Conclusions: SURD-AVR was largely performed through less invasive approaches and can be considered as a primary indication in MI surgery. In the SURD-IR cohort, MI SURD-AVR using both Perceval and Intuity valves appeared a safe and reproducible procedure associated with promising early results.

Keywords: Aortic valve replacement; Minimally invasive aortic valve replacement; Rapid deployment valve; Sutureless and Rapid-Deployment Aortic Valve Replacement International Registry; Sutureless valve.

PubMed Disclaimer

Comment in

Publication types