Observational Study
. 2019 May;63(3):284-288.
doi: 10.1007/s10384-019-00664-9. Epub 2019 Mar 8.

Efficacy of N95 amplitude of pattern electroretinogram measured from baseline to N95 trough in the traumatic optic neuropathy

Affiliations
Observational Study

Efficacy of N95 amplitude of pattern electroretinogram measured from baseline to N95 trough in the traumatic optic neuropathy

Kun Hae Kim et al. Jpn J Ophthalmol. 2019 May.

Abstract

Purpose: To investigate the utility of selected pattern electroretinogram (PERG) parameters-including N95 amplitude and N95/P50 ratio, and a BL-N95 amplitude-in the analysis of visual function(s) and for predicting changes in retinal ganglion cell structures in traumatic optic neuropathy.

Study design: A retrospective, observational case series performed at a single center.

Methods: Forty-four eyes from 36 patients diagnosed with optic neuropathy were included. A BL-N95 amplitude was defined as the amplitude measured from baseline to the trough of N95. PERG and pattern visual evoked potential (pVEP) measures were acquired within 1 week after onset of optic neuropathies. To compare functional and anatomical changes, mean temporal peripapillary retinal nerve fiber layer (pRNFL) and average and minimum ganglion cell-inner plexiform layer (GC-IPL) thicknesses were measured using optical coherence tomography.

Results: Thirty-six patients (20 men, 16 women; mean age 37.5 ± 17.6 years) were evaluated. The BL-N95 amplitude was significantly smaller than the N95 amplitude (1.01 ± 0.56 μV and 2.45 ± 1.02 μV, respectively; p < 0.0001). Both the N95 (r = - 0.38, p = 0.010) and BL-N95 r = - 0.32, p = 0.029) amplitudes were significantly correlated with visual acuity. Although P100 latency was not correlated with all PERG parameters, the N95 (r = 0.32, p = 0.032) and BL-N95 (r = 0.41, p = 0.005) amplitudes demonstrated a positive correlation with P100 amplitude in pVEP. PERG parameters, including the N95 and BL-N95 amplitudes, and N95/P50 ratio, were not correlated with pRNFL thickness in optical coherence tomography. Only the BL-N95 amplitude demonstrated a significant correlation with GC-IPL.

Conclusion: The BL-N95 amplitude-measured from baseline to the trough of N95-was valuable in the analysis of visual function(s) and for predicting changes in retinal ganglion cell structures in traumatic optic neuropathy.

Keywords: Electrophysiology; Optic neuropathy; Pattern electroretinogram; Pattern visual evoked potential.

PubMed Disclaimer

Similar articles

References

    1. Invest Ophthalmol Vis Sci. 2000 Aug;41(9):2797-810 - PubMed
    1. Prog Retin Eye Res. 2001 Jul;20(4):531-61 - PubMed
    1. Ophthalmologica. 2002 Nov-Dec;216(6):415-9 - PubMed
    1. J Clin Neurosci. 2006 Jan;13(1):55-9 - PubMed
    1. Electroencephalogr Clin Neurophysiol. 1991 Jan;78(1):18-26 - PubMed

Publication types

MeSH terms

LinkOut - more resources