Comparison of anterior segment measurements obtained using a swept-source optical coherence tomography biometer and a Scheimpflug-Placido tomographer
- PMID: 30851806
- DOI: 10.1016/j.jcrs.2018.10.033
Comparison of anterior segment measurements obtained using a swept-source optical coherence tomography biometer and a Scheimpflug-Placido tomographer
Abstract
Purpose: To assess the repeatability and agreement of anterior segment measurements between an optical biometer (IOLMaster 700) using swept-source optical coherence tomography and a Scheimpflug-Placido tomographer (Sirius) in patients scheduled for corneal refractive surgery.
Setting: Eye Hospital of Wenzhou Medical University, Zhejiang, China.
Design: Prospective case series.
Methods: The parameters, including the central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth, keratometry (K) values, astigmatism, J0 (Jackson cross-cylinder with axes at 180 degree and 90 degrees) and J45 vectors (Jackson cross-cylinder with axes at 45 degrees and 135 degrees), and horizontal corneal diameter, were measured 3 times per instrument. The intraobserver repeatability was assessed using test-retest repeatability, the coefficient of variation (CoV), and the intraclass correlation coefficient (ICC). Paired t tests were used to compare the difference between the 2 methods. Agreement was evaluated using Bland-Altman analyses.
Results: The study enrolled 100 right eyes. All measurements of the 2 devices showed high repeatability. The CoV of all measured parameters was less than 0.93 and the ICC value was more than 0.85. The paired t test did not show a significant difference in CCT, astigmatism, J0 or J45 vector. The mean ACD and aqueous depth were statistically lower with the optical biometer; however, the 95% limits of agreement (LoA) were low (from -0.20 to -0.03 mm). The K and corneal diameter values obtained using the optical biometer were higher by 0.1 diopter and 0.1 mm, respectively, but with narrow 95% LoA.
Conclusions: The optical biometer and the Scheimpflug-Placido tomographer showed high repeatability for all anterior segment measurements. The measured parameters showed good agreement and could be used interchangeably.
Copyright © 2018. Published by Elsevier Inc.
Similar articles
-
Comparison of Anterior Segment Measurements With a New Swept-source Optical Coherence Tomography Biometer and a Scheimpflug-Placido Topographer.J Refract Surg. 2025 May;41(5):e413-e420. doi: 10.3928/1081597X-20250317-03. Epub 2025 May 1. J Refract Surg. 2025. PMID: 40340682
-
Repeatability and Agreement of a Swept-Source Optical Coherence Tomography-Based Biometer IOLMaster 700 Versus a Scheimpflug Imaging-Based Biometer AL-Scan in Cataract Patients.Eye Contact Lens. 2020 Jan;46(1):35-45. doi: 10.1097/ICL.0000000000000603. Eye Contact Lens. 2020. PMID: 30985487
-
Agreement Between Swept-Source Optical Biometry and Scheimpflug-based Topography Measurements of Anterior Segment Parameters.Am J Ophthalmol. 2016 Sep;169:73-78. doi: 10.1016/j.ajo.2016.06.020. Epub 2016 Jun 16. Am J Ophthalmol. 2016. PMID: 27320057
-
Current Developments in Corneal Topography and Tomography.Diagnostics (Basel). 2021 Aug 13;11(8):1466. doi: 10.3390/diagnostics11081466. Diagnostics (Basel). 2021. PMID: 34441401 Free PMC article. Review.
-
Clinical role of swept source optical coherence tomography in anterior segment diseases: a review.Semin Ophthalmol. 2021 Nov 17;36(8):684-691. doi: 10.1080/08820538.2021.1897854. Epub 2021 Mar 10. Semin Ophthalmol. 2021. PMID: 33689554 Review.
Cited by
-
The Homburg-Adelaide toric IOL nomogram: How to predict corneal power vectors from preoperative IOLMaster 700 keratometry and total corneal power in toric IOL implantation.Acta Ophthalmol. 2025 Feb;103(1):e19-e30. doi: 10.1111/aos.16742. Epub 2024 Jul 16. Acta Ophthalmol. 2025. PMID: 39011876 Free PMC article.
-
Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract.Int J Ophthalmol. 2024 Aug 18;17(8):1437-1446. doi: 10.18240/ijo.2024.08.08. eCollection 2024. Int J Ophthalmol. 2024. PMID: 39156774 Free PMC article.
-
The effect of implantable collamer Lens V4c on ocular biometric measurements and intraocular lens power calculation based on Pentacam-AXL and IOLMaster 500.BMC Ophthalmol. 2022 Nov 5;22(1):421. doi: 10.1186/s12886-022-02644-z. BMC Ophthalmol. 2022. PMID: 36335303 Free PMC article.
-
Repeatability and agreement of a Scheimpflug-Placido tomography and a Partial Coherence Interferometer in keratoconus and control eyes.Int Ophthalmol. 2025 Mar 22;45(1):121. doi: 10.1007/s10792-025-03474-w. Int Ophthalmol. 2025. PMID: 40119977
-
Measurements of Anterior and Posterior Corneal Curvatures with OCT and Scheimpflug Biometers in Patients with Low Total Corneal Astigmatism.J Clin Med. 2022 Nov 24;11(23):6921. doi: 10.3390/jcm11236921. J Clin Med. 2022. PMID: 36498495 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical