In vitro comparison of delivery performance of 4 preloaded intraocular lens injector systems for corneal and sclerocorneal incisions
- PMID: 30853318
- DOI: 10.1016/j.jcrs.2018.10.050
In vitro comparison of delivery performance of 4 preloaded intraocular lens injector systems for corneal and sclerocorneal incisions
Abstract
Purpose: To compare preloaded intraocular lens (IOL) injector systems in porcine eyes.
Setting: Alcon Japan, Tokyo, Japan.
Design: Experimental study.
Methods: Freshly-excised porcine eyes were assigned to 11 study groups (10 eyes per group) consisting of UltraSert (system U: corneal incisions 2.2 mm and 2.4 mm), UltraSert long tip (system ULT: corneal incision 2.6 mm; sclerocorneal incisions 2.4 mm and 2.6 mm), iTec (system iT: corneal incisions 2.2 mm and 2.4 mm; sclerocorneal incisions 2.2 mm and 2.4 mm), and Vivinex iSert (system iS: corneal incision 2.0 mm; sclerocorneal incision 1.8 mm).
Results: There were a few cases with IOL adherence to the plunger tip, delayed trailing haptic, intrawound manipulation, and damage to IOL, although the occurrence was not confined to particular devices. Cartridge tip splitting was observed with system iS. System U 2.4 mm and system ULT 2.6 mm caused the least amount of incision enlargement for corneal and sclerocorneal incisions, respectively. System ULT 2.6 mm resulted in the largest final size for corneal incisions and sclerocorneal incisions. System iS induced the largest amount of incision enlargement for both incisions but had significantly smaller final incision size than all other systems for corneal incisions and sclerocorneal incisions, except for system U 2.2 mm for corneal incisions.
Conclusions: Intraocular lens implantation through a tighter incision and with cartridge-insertion technique (>wound-assisted technique) caused greater wound enlargement. A smaller preimplantation incision with a preloaded IOL delivery system did not necessarily result in significantly smaller final incision because of wound enlargement.
Copyright © 2018 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Experimental Study on Delivery Performance of an Automated Preloaded Intraocular Lens Injector System for Corneal and Sclerocorneal Incisions.J Ophthalmol. 2021 Mar 31;2021:5548493. doi: 10.1155/2021/5548493. eCollection 2021. J Ophthalmol. 2021. PMID: 33859834 Free PMC article.
-
Comparative assessment of the corneal incision enlargement of 4 preloaded IOL delivery systems.J Cataract Refract Surg. 2020 Jul;46(7):1041-1046. doi: 10.1097/j.jcrs.0000000000000214. J Cataract Refract Surg. 2020. PMID: 32304484
-
In vitro delivery performance assessment of a new preloaded intraocular lens delivery system.J Cataract Refract Surg. 2016 Dec;42(12):1814-1820. doi: 10.1016/j.jcrs.2016.10.014. J Cataract Refract Surg. 2016. PMID: 28007114
-
Incision sizes before and after implantation of 6-mm optic foldable intraocular lenses using Monarch and Unfolder injector systems.Ophthalmology. 2005 Jan;112(1):58-66. doi: 10.1016/j.ophtha.2004.06.030. Ophthalmology. 2005. PMID: 15629821 Review.
-
Characterisation of Intraocular Lens Injectors.Klin Monbl Augenheilkd. 2024 Aug;241(8):905-916. doi: 10.1055/a-2349-2158. Epub 2024 Aug 15. Klin Monbl Augenheilkd. 2024. PMID: 39146575 Review. English.
Cited by
-
Estimations of Critical Clear Corneal Incisions Required for Lens Insertion in Cataract Surgery: A Mathematical Aspect.Front Physiol. 2022 Apr 8;13:834214. doi: 10.3389/fphys.2022.834214. eCollection 2022. Front Physiol. 2022. PMID: 35464073 Free PMC article.
-
Nozzle tip damage in three generations of intraocular lens injector models: an experimental laboratory study.BMC Ophthalmol. 2023 Jan 4;23(1):7. doi: 10.1186/s12886-022-02726-y. BMC Ophthalmol. 2023. PMID: 36600209 Free PMC article.
-
Video analysis of optic-haptic-interaction during hydrophobic acrylic intraocular lens implantation using preloaded injectors.BMC Ophthalmol. 2023 Dec 19;23(1):515. doi: 10.1186/s12886-023-03216-5. BMC Ophthalmol. 2023. PMID: 38115049 Free PMC article.
-
Experimental analysis of recommended corneal incision sizes in cataract surgery using 13 intraocular lens injector systems.Sci Rep. 2023 Feb 15;13(1):2659. doi: 10.1038/s41598-023-29497-w. Sci Rep. 2023. PMID: 36792674 Free PMC article.
-
In vivo comparison of implantation behavior and laboratory analysis of two preloaded intraocular lens injectors.Eur J Ophthalmol. 2024 May;34(3):766-773. doi: 10.1177/11206721231204382. Epub 2023 Sep 26. Eur J Ophthalmol. 2024. PMID: 37750504 Free PMC article. Clinical Trial.