Comparative Study
. 2019 Jun;45(6):840-846.
doi: 10.1016/j.jcrs.2018.10.050. Epub 2019 Mar 8.

In vitro comparison of delivery performance of 4 preloaded intraocular lens injector systems for corneal and sclerocorneal incisions

Affiliations
Comparative Study

In vitro comparison of delivery performance of 4 preloaded intraocular lens injector systems for corneal and sclerocorneal incisions

Tetsuro Oshika et al. J Cataract Refract Surg. 2019 Jun.

Abstract

Purpose: To compare preloaded intraocular lens (IOL) injector systems in porcine eyes.

Setting: Alcon Japan, Tokyo, Japan.

Design: Experimental study.

Methods: Freshly-excised porcine eyes were assigned to 11 study groups (10 eyes per group) consisting of UltraSert (system U: corneal incisions 2.2 mm and 2.4 mm), UltraSert long tip (system ULT: corneal incision 2.6 mm; sclerocorneal incisions 2.4 mm and 2.6 mm), iTec (system iT: corneal incisions 2.2 mm and 2.4 mm; sclerocorneal incisions 2.2 mm and 2.4 mm), and Vivinex iSert (system iS: corneal incision 2.0 mm; sclerocorneal incision 1.8 mm).

Results: There were a few cases with IOL adherence to the plunger tip, delayed trailing haptic, intrawound manipulation, and damage to IOL, although the occurrence was not confined to particular devices. Cartridge tip splitting was observed with system iS. System U 2.4 mm and system ULT 2.6 mm caused the least amount of incision enlargement for corneal and sclerocorneal incisions, respectively. System ULT 2.6 mm resulted in the largest final size for corneal incisions and sclerocorneal incisions. System iS induced the largest amount of incision enlargement for both incisions but had significantly smaller final incision size than all other systems for corneal incisions and sclerocorneal incisions, except for system U 2.2 mm for corneal incisions.

Conclusions: Intraocular lens implantation through a tighter incision and with cartridge-insertion technique (>wound-assisted technique) caused greater wound enlargement. A smaller preimplantation incision with a preloaded IOL delivery system did not necessarily result in significantly smaller final incision because of wound enlargement.