. 2019 Nov;158(5):1317-1328.e1.
doi: 10.1016/j.jtcvs.2019.01.073. Epub 2019 Jan 31.

Bioprosthetic valve fracture: Technical insights from a multicenter study

Collaborators, Affiliations
Free article

Bioprosthetic valve fracture: Technical insights from a multicenter study

Keith B Allen et al. J Thorac Cardiovasc Surg. 2019 Nov.
Free article

Abstract

Objective: Valve-in-valve transcatheter aortic valve replacement (VIV TAVR) can result in high residual gradients that are associated with increased mortality. Bioprosthetic valve fracture (BVF) has been shown to improve residual gradients following VIV TAVR; however, factors influencing the results of BVF have not been studied.

Methods: BVF was performed in 75 patients at 21 centers. Hierarchical multiple linear regression was performed to identify variables that were associated with lower final transvalvular gradient.

Results: Surgical valves with a median true internal diameter of 18.5 mm (interquartile range, 17.0-20.5 mm) were treated with VIV TAVR in conjunction with BVF using balloon-expandable (n = 43) or self-expanding (n = 32) transcatheter heart valves with a median size of 23 mm (interquartile range, 23-23 mm). There were no aortic root disruptions, coronary occlusions, or new pacemakers; in-hospital or 30-day mortality was 2.6% (2 out of 75). Final mean transvalvular gradient was 9.2 ± 6.3 mm Hg, but was significantly lower when BVF was performed after VIV TAVR compared with BVF first (8.1 ± 4.8 mm Hg vs 16.9 ± 10.1 mm Hg; P < .001). After adjusting for timing of BVF (ie, before or after VIV TAVR), transcatheter heart valve size/type, surgical valve mode of failure, true internal diameter, and baseline gradient and BVF balloon size, performing BVF after VIV TAVR (P < .001) and using a larger BVF balloon (P = .038) were the only independent predictors of lower final mean gradient.

Conclusions: BVF can be performed safely and results in reduced residual transvalvular gradients. Performing BVF after VIV TAVR and using larger balloon appears to achieve the best hemodynamic results.

Keywords: bioprosthetic valve fracture (BVF); valve in valve transcatheter aortic valve replacement (VIV TAVR).

PubMed Disclaimer

Comment in

  • Commentary: Can a fracture be a good thing, and when?
    Hicks GL. Hicks GL. J Thorac Cardiovasc Surg. 2019 Nov;158(5):1329-1330. doi: 10.1016/j.jtcvs.2019.01.089. Epub 2019 Feb 1. J Thorac Cardiovasc Surg. 2019. PMID: 30853226 No abstract available.
  • Discussion.
    [No authors listed] [No authors listed] J Thorac Cardiovasc Surg. 2019 Nov;158(5):1325-1328. doi: 10.1016/j.jtcvs.2019.01.131. Epub 2019 Mar 8. J Thorac Cardiovasc Surg. 2019. PMID: 30857815 No abstract available.
  • Commentary: Move fast and break things.
    Smith CR. Smith CR. J Thorac Cardiovasc Surg. 2019 Nov;158(5):1331. doi: 10.1016/j.jtcvs.2019.02.034. Epub 2019 Feb 21. J Thorac Cardiovasc Surg. 2019. PMID: 30871836 No abstract available.