Implant cut-out following cephalomedullary nailing of intertrochanteric femur fractures: Are helical blades to blame?

Affiliations

Implant cut-out following cephalomedullary nailing of intertrochanteric femur fractures: Are helical blades to blame?

Ishaq Ibrahim et al. Injury. 2019 Apr.

Abstract

Introduction: Implant cut-out remains a common cause of cephalomedullary nail (CMN) failure and patient morbidity following surgical treatment of intertrochanteric femur fractures. Recent studies have suggested an increased rate of CMN cut-out with helical blades as opposed to lag screws. We compared rates of overall cut-out between helical blades and lag screws and used bivariate and multivariate analysis to determine the role of proximal fixation method among other variables on risk for cut-out. Subgroup analysis was performed on the basis of failure mechanism; superior migration (Fig. 2) versus medial perforation (Fig. 3).

Methods: Three-hundred and thirteen patient charts were retrospectively reviewed over an 8-year period; 245 patients were treated with helical blades and 68 with lag screws. Radiographs were reviewed for fracture pattern, Tip-Apex Distance (TAD), Parker's Ratio (PR) and reduction quality. Rate of implant cut-out was compared between groups and multiple logistic regression was used to analyze the ability of several independent variables to predict implant cut-out.

Results: Twenty cut-outs occurred; 15 with helical blades and 5 with lag screws. No difference in the rate of cut-out was observed between the two groups (p = 0.45). Poor fracture reduction was found to be a significant predictor of implant failure via bivariate and multiple logistic regression analysis (p = <0.01, OR 23.573). Helical blade fixation, fracture instability, TAD ≥ 25, and PR ≥ 0.45 were not predictive of implant cut-out during multivariate analysis. Similarly, patient smoking status and surgeon trauma fellowship training did not significantly increase the odds of implant cut-out. Failure by medial perforation occurred in 12 instances, all involving helical blades. Failure by superior migration occurred at a significantly higher rate with lag screws than helical blades (p = 0.02).

Conclusion: CMN cutout is likely multifactorial. A direct association between helical blade fixation and implant cut-out was not observed in our study. Amongst modifiable risk factors for implant failure, poorer fracture reduction was predictive of failure by cut-out. Subgroup analysis highlights differing modes of failure between lag screws and helical blades which warrants further investigation. Ideal TAD during helical blade fixation remains unknown.

Keywords: Cephalomeduallary nail; Cut-out; Helical blade; Hip fractures; Medial perforation; Tip-apex distance.