Biocompatibility of new low-cost (α + β)-type Ti-Mo-Fe alloys for long-term implantation
- PMID: 30889729
- DOI: 10.1016/j.msec.2019.01.133
Biocompatibility of new low-cost (α + β)-type Ti-Mo-Fe alloys for long-term implantation
Abstract
In this work, two new α + β titanium alloys with low contents of ubiquitous and low-cost alloying elements (i.e., Mo and Fe) were designed on the basis of the electronic parameters and molybdenum equivalent approaches. The designed Ti - 2Mo - 0.5Fe at. % (TMF6) and Ti - 3Mo - 0.5Fe at. % (TMF8) alloys were produced using arc melting process for studying their mechanical, electrochemical and cytotoxicity compatibilities and comparing these compatibilities to those of Ti-6Al-4V ELI alloy. The cost of the used raw materials for producing the TMF6 and TMF8 alloys are almost 1/6 of those for producing the Ti-6Al-4V ELI alloy. The hardness of the two alloys are higher than that of the Ti-6Al-4V ELI alloy, while their Young's moduli (in the range of 85-82 GPa) are lower than that of the Ti-6Al-4V ELI alloy (110 GPa). Increasing the Mo equivalent from 6 (in TMF6 alloy) to 8 (in TMF8 alloy) led to an increase in the plastic strain percent from 4% to 17%, respectively, and a decrease in the ultimate tensile strength from 949 MPa to 800 MPa, respectively. The microstructure of TMF6 alloy consists of α'/α″ phases, while TMF8 alloy substantially consists of α″ phase. The corrosion current densities and the film resistances of the new alloys are in the range of 0.70-1.07 nA/cm2 and on the order of 105 Ω·cm2, respectively. These values are more compatible with biomedical applications than those measured for the Ti-6Al-4V ELI alloy. Furthermore, the cell viabilities of the TMF6 and TMF8 alloys indicate their improved compatibility compared to that of the Ti-6Al-4V ELI alloy. The CCK-8 (Cell Counting Kit-8) assay was conducted to investigate the cytotoxicity, proliferation, and shape index of the cells of the candidate alloys. Overall, the measured compatibility of the new V-free low-cost alloys, particularly TMF8, makes them promising candidates for replacing the Ti-6Al-4V ELI alloy in biomedical applications.
Keywords: CCK-8 assay; Cytocompatibility; Electrochemical properties; Low-cost titanium alloy; Mechanical properties.
Copyright © 2019 Elsevier B.V. All rights reserved.
Similar articles
-
Performance evaluation of a low-cost Ti-Mo-Fe (TMF8) as a replacement for Ti-6Al-4V for internal fixation implants used in mandibular angular fractures: a finite element analysis study.J Mater Sci Mater Med. 2024 Nov 22;35(1):69. doi: 10.1007/s10856-024-06842-7. J Mater Sci Mater Med. 2024. PMID: 39576372 Free PMC article.
-
Screening on binary Zr-1X (X = Ti, Nb, Mo, Cu, Au, Pd, Ag, Ru, Hf and Bi) alloys with good in vitro cytocompatibility and magnetic resonance imaging compatibility.Acta Biomater. 2013 Dec;9(12):9578-87. doi: 10.1016/j.actbio.2013.07.035. Epub 2013 Aug 6. Acta Biomater. 2013. PMID: 23928334
-
Investigations into Ti-(Nb,Ta)-Fe alloys for biomedical applications.Acta Biomater. 2016 Mar 1;32:336-347. doi: 10.1016/j.actbio.2015.12.010. Epub 2015 Dec 12. Acta Biomater. 2016. PMID: 26689463
-
Comparative analysis of corrosion resistance between beta titanium and Ti-6Al-4V alloys: A systematic review.J Trace Elem Med Biol. 2020 Dec;62:126618. doi: 10.1016/j.jtemb.2020.126618. Epub 2020 Jul 9. J Trace Elem Med Biol. 2020. PMID: 32663743
-
Recent Advances and Prospects in β-type Titanium Alloys for Dental Implants Applications.ACS Biomater Sci Eng. 2024 Oct 14;10(10):6029-6060. doi: 10.1021/acsbiomaterials.4c00963. Epub 2024 Aug 30. ACS Biomater Sci Eng. 2024. PMID: 39215386 Free PMC article. Review.
Cited by
-
Application of additive manufacturing in customized titanium mandibular implants for patients with oral tumors.Oncol Lett. 2020 Oct;20(4):51. doi: 10.3892/ol.2020.11912. Epub 2020 Jul 27. Oncol Lett. 2020. PMID: 32788938 Free PMC article.
-
A Novel Design Method of Gradient Porous Structure for Stabilized and Lightweight Mandibular Prosthesis.Bioengineering (Basel). 2022 Aug 30;9(9):424. doi: 10.3390/bioengineering9090424. Bioengineering (Basel). 2022. PMID: 36134969 Free PMC article.
-
The Effect of Rod Pattern, Outrigger, and Multiple Screw-Rod Constructs for Surgical Stabilization of the 3-Column Destabilized Cervical Spine - A Biomechanical Analysis and Introduction of a Novel Technique.Neurospine. 2020 Sep;17(3):610-629. doi: 10.14245/ns.2040436.218. Epub 2020 Sep 30. Neurospine. 2020. PMID: 33022166 Free PMC article.
-
Performance evaluation of a low-cost Ti-Mo-Fe (TMF8) as a replacement for Ti-6Al-4V for internal fixation implants used in mandibular angular fractures: a finite element analysis study.J Mater Sci Mater Med. 2024 Nov 22;35(1):69. doi: 10.1007/s10856-024-06842-7. J Mater Sci Mater Med. 2024. PMID: 39576372 Free PMC article.
-
Predicting Workability of a Low-Cost Powder Metallurgical Ti-5Al-2Fe-3Mo Alloy Using Constitutive Modeling and Processing Map.Materials (Basel). 2021 Feb 9;14(4):836. doi: 10.3390/ma14040836. Materials (Basel). 2021. PMID: 33572427 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials