Characterization of the plastic scintillation detector Exradin W2 for small field dosimetry
- PMID: 30897221
- DOI: 10.1002/mp.13501
Characterization of the plastic scintillation detector Exradin W2 for small field dosimetry
Abstract
Purpose: Small field dosimetry has been an active area of research for over a decade. It is now known that large dosimetric errors can be introduced if proper detectors or correction factors are not used. The International Atomic Energy Agency (IAEA) through the technical report series No. 483 provides guidelines for small field dosimetry procedures as well as correction factors for most detectors available in the market. The plastic scintillator detector (PSD) Exradin W1 has been found to have a correction factor close to unity; however, it is not designed for beam scanning. To overcome this limitation, the new PSD Exradin W2 has been developed to be used as a scanning as well as a relative dosimeter. Characterization of this detector in small field dosimetry is presented in this study.
Methods: A 6 MV beam from a Varian-Edge linac was used to collect data for the characterization of a W2 detector. Cerenkov light ratio (CLR) is corrected through a separate new electrometer system that comes with the W2 detector. The parameters investigated include the dose and dose rate linearity, beam profiles, percent depth dose (PDD), field output factors, and temperature response. The results were compared with Gafchromic film (EBT-3 film) for beam profiles. The field output factor and temperature response were compared to the Exradin W1 detector.
Results: The dose linearity measured with 600 MU/min dose rate showed minimal variations (<0.5%) even for small MU, and similar results were seen for dose rate linearity. The comparison of field output factors between the W2 and W1 showed small differences for various depth and field sizes. The temperature response showed small variation when the temperature was varied from to . The slope was and for the W2 and the W1 detector, respectively. The differences in profiles are 0.5% in umbra and penumbra region for field size when compared to the EBT-3 film profile.
Conclusions: The W2 scintillator detector showed similar dosimetric and temperature properties to the W1 scintillator detector. The main advantage of the W2 detector among other plastic scintillators is the beam scanning capabilities that, combined with its correction factor of 1.0, make it an ideal detector for commissioning of SRS and SBRT techniques.
Keywords: Exradin W2; plastic scintillator detector; scintillator characterization; small field dose profiles; small field dosimetry; small field output factors.
© 2019 American Association of Physicists in Medicine.
Similar articles
-
Characterization of the Plastic Scintillator Detector System Exradin W2 in a High Dose Rate Flattening-Filter-Free Photon Beam.Sensors (Basel). 2022 Sep 8;22(18):6785. doi: 10.3390/s22186785. Sensors (Basel). 2022. PMID: 36146135 Free PMC article.
-
Dosimetric characterization and behaviour in small X-ray fields of a microchamber and a plastic scintillator detector.Br J Radiol. 2017 Jan;90(1069):20160596. doi: 10.1259/bjr.20160596. Epub 2016 Nov 9. Br J Radiol. 2017. PMID: 27826990 Free PMC article.
-
Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields.J Appl Clin Med Phys. 2022 Aug;23(8):e13728. doi: 10.1002/acm2.13728. Epub 2022 Jul 21. J Appl Clin Med Phys. 2022. PMID: 35861648 Free PMC article.
-
Dosimetry quality control based on percent depth dose rate variation for checking beam quality in radiotherapy.Rep Pract Oncol Radiother. 2020 Jul-Aug;25(4):484-488. doi: 10.1016/j.rpor.2020.03.016. Epub 2020 May 1. Rep Pract Oncol Radiother. 2020. PMID: 32477013 Free PMC article. Review.
-
High-Z Sensitized Plastic Scintillators: A Review.Adv Mater. 2018 Jul;30(27):e1706956. doi: 10.1002/adma.201706956. Epub 2018 May 7. Adv Mater. 2018. PMID: 29736994 Review.
Cited by
-
Characterization of the Plastic Scintillator Detector System Exradin W2 in a High Dose Rate Flattening-Filter-Free Photon Beam.Sensors (Basel). 2022 Sep 8;22(18):6785. doi: 10.3390/s22186785. Sensors (Basel). 2022. PMID: 36146135 Free PMC article.
-
High spatial resolution inorganic scintillator detector for high-energy X-ray beam at small field irradiation.Med Phys. 2020 Mar;47(3):1364-1371. doi: 10.1002/mp.14002. Epub 2020 Jan 23. Med Phys. 2020. PMID: 31883388 Free PMC article.
-
Validation of a New Scintillating Fiber Dosimeter for Radiation Dose Quality Control in Computed Tomography.Sensors (Basel). 2023 Feb 27;23(5):2614. doi: 10.3390/s23052614. Sensors (Basel). 2023. PMID: 36904817 Free PMC article.
-
Comparison and Validation of Multiple Detectors against Monte Carlo Simulation for the Use of Small-Field Dosimetry.J Med Phys. 2022 Jul-Sep;47(3):235-242. doi: 10.4103/jmp.jmp_35_22. Epub 2022 Nov 8. J Med Phys. 2022. PMID: 36684698 Free PMC article.
-
A comprehensive investigation of the performance of a commercial scintillator system for applications in electron FLASH radiotherapy.Med Phys. 2024 Jun;51(6):4504-4512. doi: 10.1002/mp.17030. Epub 2024 Mar 20. Med Phys. 2024. PMID: 38507253 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials