Treatment of Coronary Drug-Eluting Stent Restenosis by a Sirolimus- or Paclitaxel-Coated Balloon
- PMID: 30898253
- DOI: 10.1016/j.jcin.2018.11.040
Treatment of Coronary Drug-Eluting Stent Restenosis by a Sirolimus- or Paclitaxel-Coated Balloon
Abstract
Objectives: The aim of this randomized controlled trial was to investigate a novel sirolimus-coated balloon (SCB) compared with the best investigated paclitaxel-coated balloon (PCB).
Background: Treatment of coronary in-stent restenosis (ISR) remains challenging. PCBs are an established treatment option outside the United States with a Class I, Level of Evidence: A recommendation in the European guidelines. However, their efficacy is better in bare-metal stent (BMS) ISR compared with drug-eluting stent (DES) ISR.
Methods: Fifty patients with DES ISR were enrolled in a randomized, multicenter trial to compare a novel SCB (SeQuent SCB, 4 μg/mm2) with a clinically proven PCB (SeQuent Please Neo, 3 μg/mm2) in coronary DES ISR. The primary endpoint was angiographic late lumen loss at 6 months. Secondary endpoints included procedural success, major adverse cardiovascular events, and individual clinical endpoints such as stent thrombosis, cardiac death, target lesion myocardial infarction, clinically driven target lesion revascularization, and binary restenosis.
Results: Quantitative coronary angiography revealed no differences in baseline parameters. After 6 months, in-segment late lumen loss was 0.21 ± 0.54 mm in the PCB group versus 0.17 ± 0.55 mm in the SCB group (p = NS; per-protocol analysis). Clinical events up to 12 months also did not differ between the groups.
Conclusions: This first-in-man comparison of a novel SCB with a crystalline coating shows similar angiographic outcomes in the treatment of coronary DES ISR compared with a clinically proven PCB. (Treatment of Coronary In-Stent Restenosis by a Sirolimus [Rapamycin] Coated Balloon or a Paclitaxel Coated Balloon [FIM LIMUS DCB]; NCT02996318).
Keywords: coronary in-stent restenosis; paclitaxel-coated balloon; sirolimus-coated balloon.
Copyright © 2019 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
-
Sirolimus-Coated Balloons: A Ferrari for Our Coronary Patients or Another Elegant Town Car?JACC Cardiovasc Interv. 2019 Mar 25;12(6):567-568. doi: 10.1016/j.jcin.2018.12.034. JACC Cardiovasc Interv. 2019. PMID: 30898254 No abstract available.
Publication types
MeSH terms
Substances
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous