Liquid Calibration Phantoms in Ultra-Low-Dose QCT for the Assessment of Bone Mineral Density
- PMID: 30902572
- DOI: 10.1016/j.jocd.2019.02.005
Liquid Calibration Phantoms in Ultra-Low-Dose QCT for the Assessment of Bone Mineral Density
Abstract
Introduction: Cortical bone is affected by metabolic diseases. Some studies have shown that lower cortical bone mineral density (BMD) is related to increases in fracture risk which could be diagnosed by quantitative computed tomography (QCT). Nowadays, hybrid iterative reconstruction-based (HIR) computed tomography (CT) could be helpful to quantify the peripheral bone tissue. A key focus of this paper is to evaluate liquid calibration phantoms for BMD quantification in the tibia and under hybrid iterative reconstruction-based-CT with the different hydrogen dipotassium phosphate (K2HPO4) concentrations phantoms.
Methodology: Four ranges of concentrations of K2HPO4 were made and tested with 2 exposure settings. Accuracy of the phantoms with ash gravimetry and intermediate K2HPO4 concentration as hypothetical patients were evaluated. The correlations and mean differences between measured equivalent QCT BMD and ash density as a gold standard were calculated. Relative percentage error (RPE) in CT numbers of each concentration over a 6-mo period was reported.
Results: The correlation values (R2 was close to 1.0), suggested that the precision of QCT-BMD measurements using standard and ultra-low dose settings were similar for all phantoms. The mean differences between QCT-BMD and the ash density for low concentrations (about 93 mg/cm3) were lower than high concentration phantoms with 135 and 234 mg/cm3 biases. In regard to accuracy test for hypothetical patient, RPE was up to 16.1% for the low concentration (LC) phantom for the case of high mineral content. However, the lowest RPE (0.4 to 1.8%) was obtained for the high concentration (HC) phantom, particularly for the high mineral content case. In addition, over 6 months, the K2HPO4 concentrations increased 25% for 50 mg/cm3 solution and 0.7 % for 1300 mg/cm3 solution in phantoms.
Conclusion: The excellent linear correlations between the QCT equivalent density and the ash density gold standard indicate that QCT can be used with submilisivert radiation dose. We conclude that using liquid calibration phantoms with a range of mineral content similar to that being measured will minimize bias. Finally, we suggest performing BMD measurements with ultra-low dose scan concurrent with iterative-based reconstruction to reduce radiation exposure.
Keywords: cortical bone; hybrid iterative reconstruction; quantitative computed tomography; tibia; ultra-low dose.
Copyright © 2019 The International Society for Clinical Densitometry. All rights reserved.
Similar articles
-
Quantitative micro-computed tomography: a non-invasive method to assess equivalent bone mineral density.Bone. 2008 Aug;43(2):302-311. doi: 10.1016/j.bone.2008.04.009. Epub 2008 Apr 30. Bone. 2008. PMID: 18539557
-
Improving Bone Mineral Density Assessment Using Spectral Detector CT.J Clin Densitom. 2019 Jul-Sep;22(3):374-381. doi: 10.1016/j.jocd.2018.10.004. Epub 2018 Oct 26. J Clin Densitom. 2019. PMID: 30497869
-
A Novel Dipotassium Hydrogen Phosphate Phantom for Calibrating Computed Tomographic Bone Density Measurements in Birds.J Avian Med Surg. 2022 May;36(1):28-38. doi: 10.1647/20-00018. J Avian Med Surg. 2022. PMID: 35526162
-
Quantitative Computed Tomography (QCT) derived Bone Mineral Density (BMD) in finite element studies: a review of the literature.J Exp Orthop. 2016 Dec;3(1):36. doi: 10.1186/s40634-016-0072-2. Epub 2016 Dec 9. J Exp Orthop. 2016. PMID: 27943224 Free PMC article. Review.
-
Application of quantitative computed tomography for assessment of trabecular bone mineral density, microarchitecture and mechanical property.Clin Imaging. 2016 Mar-Apr;40(2):330-8. doi: 10.1016/j.clinimag.2015.09.016. Epub 2015 Sep 21. Clin Imaging. 2016. PMID: 26602163 Review.
Cited by
-
Development and validation of a fully automated system using deep learning for opportunistic osteoporosis screening using low-dose computed tomography scans.Quant Imaging Med Surg. 2023 Aug 1;13(8):5294-5305. doi: 10.21037/qims-22-1438. Epub 2023 Jul 20. Quant Imaging Med Surg. 2023. PMID: 37581046 Free PMC article.
-
Comparative evaluation of accuracy, precision, and radiation dose between mindways and low-dose iCare QCT for lumbar spine BMD using the European spine phantom.Front Med Technol. 2025 May 12;7:1575553. doi: 10.3389/fmedt.2025.1575553. eCollection 2025. Front Med Technol. 2025. PMID: 40433429 Free PMC article.
-
Diagnosis of Osteoporosis by Quantifying Volumetric Bone Mineral Density of Lumbar Vertebrae Using Abdominal CT Images and Two-Compartment Model.Healthcare (Basel). 2023 Feb 13;11(4):556. doi: 10.3390/healthcare11040556. Healthcare (Basel). 2023. PMID: 36833090 Free PMC article.
-
MDCT-QCT, QUS, and DXA in healthy adults: An intermodality comparison.Med J Islam Repub Iran. 2019 Dec 28;33:156. doi: 10.34171/mjiri.33.156. eCollection 2019. Med J Islam Repub Iran. 2019. PMID: 32280662 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous