Accuracy of the ClearSight™ system in patients undergoing abdominal aortic aneurysm surgery
- PMID: 30904953
- DOI: 10.1007/s00540-019-02632-6
Accuracy of the ClearSight™ system in patients undergoing abdominal aortic aneurysm surgery
Abstract
Purpose: The ClearSight™ device monitors continuous pressure and cardiac output via pulse contour analysis. ClearSight™, however, may not be reliable in patients with reduced peripheral perfusion caused by high peripheral resistance. This study aimed to elucidate the accuracy and trending ability of ClearSight™ in patients undergoing abdominal aortic aneurysm (AAA) surgery by comparing the ClearSightTM-derived cardiac index (CICS) with that measured using three-dimensional echocardiography (CI3D).
Methods: The study included 20 patients who underwent elective AAA surgery. CICS and CI3D were measured simultaneously at eight time points during the surgery. Trending ability was investigated after aortic clamping and unclamping. We used CI3D as the reference method.
Results: Bland-Altman analysis showed a wide limit of agreement between CICS and CI3D (percentage error 41.3%). Subgroup analysis showed a lower percentage error (33.2%) in patients with CI ≥ 2.5 L/min/m2. The cubic splines related to the CI3D and CI discrepancy were negatively sloped, indicating that CI3D had significant influence on the CI discrepancy (p < 0.001). Four-quadrant plot analysis showed that the tracking ability of ClearSight™ after aortic clamping and declamping were clinically unacceptable (81.3% and 78.6%, respectively). Also, the polar plot analysis showed that the concordance rate of ClearSight™ after aortic clamping and declamping were clinically unacceptable (58.3% and 66.7%, respectively).
Conclusions: ClearSight™ was not sufficiently accurate in patients undergoing AAA surgery. The tracking ability of ClearSight™ after aortic clamping was below the acceptable limit.
Keywords: Blood pressure monitor; Cardiac output; ClearSight; Echocardiography; Ultrasound.
Similar articles
-
Accuracy and trending ability of the fourth-generation FloTrac/Vigileo System™ in patients undergoing abdominal aortic aneurysm surgery.J Anesth. 2018 Jun;32(3):387-393. doi: 10.1007/s00540-018-2491-y. Epub 2018 Apr 3. J Anesth. 2018. PMID: 29616345
-
Accuracy and Trending Ability of Cardiac Index Measured by the CNAP System in Patients Undergoing Abdominal Aortic Aneurysm Surgery.J Cardiothorac Vasc Anesth. 2021 May;35(5):1439-1446. doi: 10.1053/j.jvca.2020.08.012. Epub 2020 Aug 11. J Cardiothorac Vasc Anesth. 2021. PMID: 32888805
-
Clearsight™ use for haemodynamic monitoring during the third trimester of pregnancy - a validation study.Int J Obstet Anesth. 2018 Nov;36:85-95. doi: 10.1016/j.ijoa.2018.04.009. Epub 2018 May 8. Int J Obstet Anesth. 2018. PMID: 30392653
-
A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output.Anesth Analg. 2010 Nov;111(5):1180-92. doi: 10.1213/ANE.0b013e3181f08a5b. Epub 2010 Aug 24. Anesth Analg. 2010. PMID: 20736431 Review.
-
Open thoracic or thoracoabdominal aortic aneurysm repair after previous abdominal aortic aneurysm surgery.J Vasc Surg. 2008 Sep;48(3):761-8. doi: 10.1016/j.jvs.2008.02.006. Epub 2008 May 16. J Vasc Surg. 2008. PMID: 18486428 Review.
Cited by
-
Update on the assessment of fluid responsiveness.J Anesth. 2020 Apr;34(2):163-166. doi: 10.1007/s00540-019-02731-4. Epub 2020 Jan 7. J Anesth. 2020. PMID: 31912310 No abstract available.
-
Comparison of invasive and non-invasive measurements of cardiac index and systemic vascular resistance in living-donor liver transplantation: a prospective, observational study.BMC Anesthesiol. 2023 Nov 4;23(1):359. doi: 10.1186/s12871-023-02302-x. BMC Anesthesiol. 2023. PMID: 37924013 Free PMC article.
-
Non-invasive continuous blood pressure monitoring using the ClearSight system for pregnant women at high risks of post-partum hemorrhage: comparison with invasive blood pressure monitoring during cesarean section.Obstet Gynecol Sci. 2022 Jul;65(4):325-334. doi: 10.5468/ogs.22063. Epub 2022 Jun 27. Obstet Gynecol Sci. 2022. PMID: 35754365 Free PMC article.
-
Continuous Non-Invasive Hemodynamic Monitoring in Cirrhotic Patients-Friend or Foe?Medicina (Kaunas). 2025 Mar 19;61(3):536. doi: 10.3390/medicina61030536. Medicina (Kaunas). 2025. PMID: 40142347 Free PMC article.
-
Continuous Non-Invasive Arterial Pressure Monitoring (ClearSight System) and Ankle Blood Pressure Measurements as Alternatives to Conventional Arm Blood Pressure.J Clin Med. 2020 Nov 10;9(11):3615. doi: 10.3390/jcm9113615. J Clin Med. 2020. PMID: 33182672 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical