Randomized Controlled Trial
. 2019 Jul;54(7):1052-1059.
doi: 10.1002/ppul.24295. Epub 2019 Mar 25.

Performance of a novel reusable pediatric pulse oximeter probe

Affiliations
Randomized Controlled Trial

Performance of a novel reusable pediatric pulse oximeter probe

Carina King et al. Pediatr Pulmonol. 2019 Jul.

Abstract

Objective: To assess the performance of reusable pulse oximeter probe and microprocessor box combinations, of varying price-points, in the context of a low-income pediatric setting.

Methods: A prospective, randomized cross-over study comparing time to biologically plausible oxygen saturation (SpO2 ) between: (1) Lifebox LB-01 probe with Masimo Rad-87 box (L + M) and (2) a weight-appropriate reusable Masimo probe with Masimo Rad-87 box (M + M). A post hoc secondary analysis comparison with historical usability testing data with the Lifebox LB-01 probe and Lifebox V1.5 box (L + L) was also conducted. Participants, children aged 0 to 35 months, were recruited from pediatric wards and outpatient clinics in the central region of Malawi. The primary outcome was time taken to achieve a biologically plausible SpO 2 measurement, compared using t tests for equivalence.

Results: We recruited 572 children. Plausible SpO2 measurements were obtained in less than 1 minute, 71%, 70%, and 63% for the M + M, L + M, and L + L combinations, respectively. A similar pattern was seen for less than 2 minutes, however, this effect disappeared at less than 5 minutes with 96%, 96%, and 95% plausible measurements. Using a ±10 second threshold for equivalence, we found L + M and M + M to be equivalent, but were under-powered to assess equivalence for L + L.

Conclusions: The novel reusable pediatric Lifebox probe can achieve a quality SpO2 measurement within a pragmatic time range of weight-appropriate Masimo equivalent probes. Further research, which considers the cost of the devices, is needed to assess the added value of sophisticated motion tolerance software.

Keywords: LMIC; hypoxemia; pediatric; pulse oximeter.

PubMed Disclaimer

Conflict of interest statement

The authors declare that there are no conflict of interests.

Figures

Figure 1
Figure 1
Photograph of the LB‐01 probe in use [Color figure can be viewed at wileyonlinelibrary.com]
Figure 2
Figure 2
Mean difference in time to biologically plausible oxygen saturation measurement. L + L, Lifebox LB‐01 probe and Lifebox V1.5 box; L + M, Lifebox LB‐01 probe with Masimo Rad‐87 box; [Color figure can be viewed at wileyonlinelibrary.com]

References

    1. Lazzerini M, Sonego M, Pellegrin MC. Hypoxaemia as a mortality risk factor in acute lower respiratory infections in children in low and middle‐income countries: systematic review and meta‐analysis. PLOS One. 2015;10(9):e0136166. - PMC - PubMed
    1. World Health Organization . Health A. Handbook IMCI: Integrated management of childhood illness. Geneva: WHO; 2005.
    1. Alwadhi V, Dewan P, Malhotra RK, Shah D, Gupta P. Tachypnea and other danger signs vs pulse oximetry for prediction of hypoxia in severe pneumonia/very severe disease. Indian Pediatr. 2017;54(9):729‐734. - PubMed
    1. Basnet S, Adhikari RK, Gurung CK. Hypoxemia in children with pneumonia and its clinical predictors. Indian J Pediatr. 2006;73(9):777‐781. - PubMed
    1. Duke T, Subhi R, Peel D, Frey B. Pulse oximetry: technology to reduce child mortality in developing countries. Ann Trop Paediatr. 2009;29(3):165‐175. - PubMed

Publication types