Monitoring Hemostasis During Extracorporeal Life Support
- PMID: 30913107
- DOI: 10.1097/MAT.0000000000000993
Monitoring Hemostasis During Extracorporeal Life Support
Abstract
To balance the risk of bleeding versus circuit thrombosis during extracorporeal life support (ECLS), it is important to monitor anticoagulants and hemostasis. We evaluated the prothrombin time (PT), partial thromboplastin time (PTT), activated clotting time (ACT), and antifactor Xa heparin activity (aXa) correlation with changes in coagulation factor and heparin levels using in vitro and in vivo samples. aXa correlated with heparin (r = 0.97) and antithrombin (r = 0.98) but was unaffected by other parameters. PT correlated with coagulation factors (r = 0.88) but was minimally affected by heparin or other parameters. When single parameters were changed, ACT was insensitive to <0.5 U/ml heparin, correlated with coagulation factors (r = 0.99), and was affected by factor XII and platelets. When multiple parameters changed in vitro and in vivo, ACT was not correlated with heparin or coagulation factors. PTT correlated with heparin and coagulation factors individually but had low correlation when multiple parameters changed in vitro and in vivo. In conclusion, aXa is the most specific for heparin levels, and PT is most specific for coagulation factor levels making these assays well suited to monitor anticoagulation and hemostasis for patients on ECLS. PTT is highly variable when multiple parameters are changing in vitro and in vivo, but may be useful when aXa cannot be used because of interference. ACT is too insensitive to heparin, sensitive to too many other factors, and too imprecise to be useful for monitoring hemostasis during ECLS.
References
-
- Urlesberger B, Zobel G, Zenz W, et al. Activation of the clotting system during extracorporeal membrane oxygenation in term newborn infants. J Pediatr 1996.129: 264–268.
-
- Bembea MM, Schwartz JM, Shah N, et al. Anticoagulation monitoring during pediatric extracorporeal membrane oxygenation. ASAIO J 2013.59: 63–68.
-
- Bembea MM, Annich G, Rycus P, Oldenburg G, Berkowitz I, Pronovost P. Variability in anticoagulation management of patients on extracorporeal membrane oxygenation: an international survey. Pediatr Crit Care Med 2013.14: e77–e84.
-
- Guervil DJ, Rosenberg AF, Winterstein AG, Harris NS, Johns TE, Zumberg MS. Activated partial thromboplastin time versus antifactor Xa heparin assay in monitoring unfractionated heparin by continuous intravenous infusion. Ann Pharmacother 2011.45: 861–868.
-
- Vandiver JW, Vondracek TG. A comparative trial of anti-factor Xa levels versus the activated partial thromboplastin time for heparin monitoring. Hosp Pract (1995) 2013.41: 16–24.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources