Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2018 Oct;18(19):7899-7906.
doi: 10.1109/JSEN.2018.2863644. Epub 2018 Aug 6.

CMOS Monolithic Electrochemical Gas Sensor Microsystem Using Room Temperature Ionic Liquid

Affiliations

CMOS Monolithic Electrochemical Gas Sensor Microsystem Using Room Temperature Ionic Liquid

Heyu Yin et al. IEEE Sens J. 2018 Oct.

Abstract

The growing demand for personal healthcare monitoring requires a challenging combination of performance, size, power, and cost that is difficult to achieve with existing gas sensor technologies. This paper presents a new CMOS monolithic gas sensor microsystem that meets these requirements through a unique combination of electrochemical readout circuits, post-CMOS planar electrodes, and room temperature ionic liquid (RTIL) sensing materials. The architecture and design of the CMOS-RTIL-based monolithic gas sensor are described. The monolithic device occupies less than 0.5mm2 per sensing channel and incorporates electrochemical biasing and readout functions with only 1.4mW of power consumption. Oxygen was tested as an example gas, and results show that the microsystem demonstrates a highly linear response (R2 = 0.995) over a 0 - 21% oxygen concentration range, with a limit of detection of 0.06% and a 1 second response time. Monolithic integration reduces manufacturing cost and is demonstrated to improve limits of detection by a factor of five compared to a hybrid implementation. The combined characteristics of this device offer an ideal platform for portable/wearable gas sensing in applications such as air pollutant monitoring.

Keywords: electrochemical; gas sensor; monolithic microsystem; room temperature ionic liquid.

PubMed Disclaimer

Figures

Fig. 1.
Fig. 1.
Block diagram of an amperometric electrochemical sensor system.
Fig. 2.
Fig. 2.
Conceptual illustration of the CMOS monolithic RTIL-based electrochemical gas sensor microsystem.
Fig. 3.
Fig. 3.
Schematic of the monolithic microsystem, consisting of the amperometric instrumentation circuit and the RTIL-based sensor. Green rectangle block illustrates the clock waveform.
Fig. 4.
Fig. 4.
Die photographs of fabricated CMOS electrochemical instrumentation chip; (a) chip as received from foundry with functional blocks labeled, (b) chip after post-CMOS fabrication of (top) on-CMOS electrodes and (bottom) on-CMOS RTIL sensor/electrolyte layer.
Fig. 5.
Fig. 5.
Post-CMOS pocess flow for on-chip RTIL-based sensor fabrication: Photoresist is spin-coated (a) and developed (b). Titanium and gold is deposited by PVD (c). Photoresist is then rinsed off to leave the electrode (d). CMOS chip with on-chip electrode is wire bonded to package (e). A droplet of RTIL is casted on the electrodes (f).
Fig. 6.
Fig. 6.
The electrochemical test setup.
Fig. 7.
Fig. 7.
The R2rop output voltage range and linearity error as a function of input voltage. The absolute error of the output voltage is less than 0.05% over the range from 0.03 V to 4.90 V.
Fig. 8.
Fig. 8.
Amperometric readout output range and linearity error as a function of input current when fs = 100 kHz, Vref2 = 1.3 V. In the range of −1 μA to 1 μA, the absolute linearity error is less than 4 nA.
Fig. 9.
Fig. 9.
Output of amperometric readout circuit for on-CMOS RTIL-based sensor response to oxygen. Oxygen concentration varies from 0 – 21% with a step of 2.1%. Oxygen exposures were 10 s in duration, preceded and followed by exposure to pure nitrogen for 30 s. The amperometric circuit was set to Vref1 = 2.1 V and Vref2 = 1.3 V (the sensor was applied at −0.8V respectively), and fs=100 kHz.
Fig. 10.
Fig. 10.
Oxygen calibration curve for monolithic RTIL-based gas sensor using data extracted from Fig. 9. A high level of linearity (R2 = 0.995) was achieved.
Fig. 11.
Fig. 11.
Repeatability test. Constant-potential current response measured over five cycles of alternate exposure to 4.2% oxygen and pure nitrogen flow at an applied bias of −0.8 V. The standard deviation of the five test cycles is 89 μV and the corresponding repeatable LOD of oxygen is 0.06%.
Fig. 12.
Fig. 12.
Off-CMOS and on-CMOS sensor response time characterized by the amperometric readout circuit. The response time saturates around 1 sec.

References

    1. Brook RD, “Cardiovascular effects of air pollution,” Clin. Sci, vol. 115, no. 6, pp. 175–187, 2008. - PubMed
    1. Lave LB and Seskin EP, Air pollution and human health. 2013. - PubMed
    1. Hagleitner C, Hierlemann A, Lange D, Kummer A, Kerness N, Brand O and Batltes H, “Smart single-chip gas sensor microsystem.,” Nature, vol. 414, no. 6861, pp. 293–296, 2001. - PubMed
    1. Philippe J, Arndt G, Colinet E, Savoye M, Ernst T, Ollier E and Arcamone J, “Fully monolithic and ultra-compact NEMS-CMOS selfoscillator based-on single-crystal silicon resonators and low-cost CMOS circuitry,” Proc. IEEE Int. Conf. Micro Electro Mech. Syst, pp. 1071–1074, 2014.
    1. Afridi MY, Suehle JS, Zaghloul ME, Berning DW, Cavicchi RE, Semanick S, Montgomery CB and Taylor CJ, “A monolithic CMOS microhotplate-based gas sensor system,” IEEE Sens. J, vol. 2, no. 6, pp. 644–655, 2002.

LinkOut - more resources