First comparison of performances between the new whole-body cadmium-zinc-telluride SPECT-CT camera and a dedicated cardiac CZT camera for myocardial perfusion imaging: Analysis of phantom and patients
- PMID: 30963419
- DOI: 10.1007/s12350-019-01702-2
First comparison of performances between the new whole-body cadmium-zinc-telluride SPECT-CT camera and a dedicated cardiac CZT camera for myocardial perfusion imaging: Analysis of phantom and patients
Abstract
Background: Dedicated cardiac Cadmium-zinc-telluride (CZT) cameras show superior performances compared with Anger systems, particularly in terms of spatial resolution and count sensitivity. This study evaluated the performances of a new polyvalent whole body CZT camera (DNM 670CZT) compared with a cardiac dedicated CZT camera (DNM 530c) for myocardial perfusion SPECT.
Methods: The spatial resolution was evaluated with three linear sources filled with 99mTc. We used a cardiac phantom to evaluate count sensitivity, sharpness index, contrast-to-noise ratio, wall thickness, non-uniformity index, perfusion scores and ventricle volumes for both cameras. The impact of matrix size, and acquisition time was investigated. Concordance between the two cameras was evaluated in patients using QPS/QGS software for quantitative segmental perfusion, motion and thickness scores.
Results: The spatial resolution was identical with the two cameras. Count sensitivity of the DNM 670CZT was twofold lower compared with the DNM 530c, leading to lower sharpness index and contrast-to-noise ratio. The wall thickness and the myocardial volumes were similar. Visual and quantitative assessments of the perfusion patterns have shown a good concordance of the two cameras on phantoms and in patients.
Conclusion: This study demonstrated the feasibility of myocardial perfusion SPECT imaging using the new whole-body DNM 670CZT camera.
Keywords: CZT-cameras; SPECT; myocardial perfusion imaging.
Comment in
-
Cardiac myocardial perfusion imaging with new SPECT cameras: Comparing apples and oranges.J Nucl Cardiol. 2020 Aug;27(4):1270-1273. doi: 10.1007/s12350-019-01729-5. Epub 2019 May 1. J Nucl Cardiol. 2020. PMID: 31044404 No abstract available.
Similar articles
-
A performance comparison of novel cadmium-zinc-telluride camera and conventional SPECT/CT using anthropomorphic torso phantom and water bags to simulate soft tissue and breast attenuation.Ann Nucl Med. 2015 May;29(4):342-50. doi: 10.1007/s12149-015-0952-z. Epub 2015 Jan 28. Ann Nucl Med. 2015. PMID: 25628019
-
Comparison of myocardial perfusion imaging using thallium-201 between a new cadmium-zinc-telluride cardiac camera and a conventional SPECT camera.Clin Nucl Med. 2011 Sep;36(9):776-80. doi: 10.1097/RLU.0b013e31821a294e. Clin Nucl Med. 2011. PMID: 21825848 Clinical Trial.
-
Comparison between stress myocardial perfusion SPECT recorded with cadmium-zinc-telluride and Anger cameras in various study protocols.Eur J Nucl Med Mol Imaging. 2013 Feb;40(3):331-40. doi: 10.1007/s00259-012-2292-8. Epub 2012 Nov 27. Eur J Nucl Med Mol Imaging. 2013. PMID: 23184308 Clinical Trial.
-
Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM).Eur J Nucl Med Mol Imaging. 2016 Dec;43(13):2423-2432. doi: 10.1007/s00259-016-3467-5. Epub 2016 Aug 19. Eur J Nucl Med Mol Imaging. 2016. PMID: 27542010 Review.
-
Novel Cadmium Zinc Telluride Devices for Myocardial Perfusion Imaging-Technological Aspects and Clinical Applications.Semin Nucl Med. 2016 Jul;46(4):273-85. doi: 10.1053/j.semnuclmed.2016.01.002. Semin Nucl Med. 2016. PMID: 27237438 Review.
Cited by
-
Comparative Analysis of Cardiac SPECT Myocardial Perfusion Imaging: Full-Ring Solid-State Detectors Versus Dedicated Cardiac Fixed-Angle Gamma Camera.Medicina (Kaunas). 2025 Apr 4;61(4):665. doi: 10.3390/medicina61040665. Medicina (Kaunas). 2025. PMID: 40282956 Free PMC article.
-
High-quality brain perfusion SPECT images may be achieved with a high-speed recording using 360° CZT camera.EJNMMI Phys. 2020 Nov 4;7(1):65. doi: 10.1186/s40658-020-00334-7. EJNMMI Phys. 2020. PMID: 33146804 Free PMC article.
-
Reduced acquisition time for thallium myocardial perfusion imaging with large field cadmium-zinc-telluride SPECT/CT cameras: An equivalence study.J Nucl Cardiol. 2022 Aug;29(4):1933-1941. doi: 10.1007/s12350-021-02611-z. Epub 2021 Apr 22. J Nucl Cardiol. 2022. PMID: 33890184
-
A review of state-of-the-art resolution improvement techniques in SPECT imaging.EJNMMI Phys. 2025 Jan 30;12(1):9. doi: 10.1186/s40658-025-00724-9. EJNMMI Phys. 2025. PMID: 39883257 Free PMC article. Review.
-
Same same but different: dopamine transporter SPECT on scanners with CZT vs. NaI detectors.EJNMMI Res. 2023 Mar 22;13(1):24. doi: 10.1186/s13550-023-00973-8. EJNMMI Res. 2023. PMID: 36949290 Free PMC article.
References
-
- Blaire T, Bailliez A, Bouallegue FB, Bellevre D, Agostini D, Manrique A. Left ventricular function assessment using 123I/99mTc dual-isotope acquisition with two semi-conductor cadmium–zinc–telluride (CZT) cameras: A gated cardiac phantom study. EJNMMI Phys 2016;3:1-11. - DOI
-
- Bailliez A, Blaire T, Mouquet F, Legghe R, Etienne B, Legallois D, et al. Segmental and global left ventricular function assessment using gated SPECT with a semiconductor Cadmium Zinc Telluride (CZT) camera: Phantom study and clinical validation vs cardiac magnetic resonance. J Nucl Cardiol 2014;21:712-22. - DOI
-
- Gambhir SS, Berman DS, Ziffer J, Nagler M, Sandler M, Patton J, et al. A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera. J Nucl Med 2009;50:635-43. - DOI
-
- Erlandsson K, Kacperski K, van Gramberg D, Hutton BF. Performance evaluation of D-SPECT: A novel SPECT system for nuclear cardiology. Phys Med Biol 2009;54:2635-49. - DOI
-
- Bellevre D, Manrique A, Legallois D, Bross S, Baavour R, Roth N, et al. First determination of the heart-to-mediastinum ratio using cardiac dual isotope (123I-MIBG/99mTc-tetrofosmin) CZT imaging in patients with heart failure: The ADRECARD study. Eur J Nucl Med Mol Imaging 2015;42:1912-9. - DOI
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources