The accuracy and precision of an open-circuit system to measure oxygen consumption and carbon dioxide production in neonates
- PMID: 3099255
- DOI: 10.1203/00006450-198701000-00014
The accuracy and precision of an open-circuit system to measure oxygen consumption and carbon dioxide production in neonates
Abstract
We measured the oxygen consumption, carbon dioxide production, and respiratory quotient during the combustion of a known mass of anhydrous ethanol and methanol to assess the accuracy of an open-circuit flow-through system. Continuous measurements were made of the mass of alcohol burned, the velocity of gas flow through the apparatus, and simultaneous measurements of the fractional concentration of oxygen, carbon dioxide and nitrogen of the inlet and outlet gas using paramagnetic oxygen analyzer, infrared carbon dioxide meter, and mass spectrometer. Standard respiratory and stoichiometric equations were used to calculate the oxygen consumption, carbon dioxide production and RQ for the mass of absolute alcohol combustion per unit time. In a series of 12 consecutive laboratory experiments (on 7 days), the measured values of gas exchange (similar to the rate of respiratory gas exchange by an infant of 1-4 kg) were in agreement within 5% of the true values for ethanol and methanol combustion, confirming the validity of the open-circuit method. The paramagnetic oxygen analyzer and the mass spectrometer gave similar oxygen consumption results and differed very little when the rate of absolute alcohol combustion was used to quantify the accuracy of the complete measurement system. A positive measurement error was observed for the carbon dioxide production results from both the IR meter and mass spectrometer, with the result that the respiratory quotient measurements were 3.4-4.7% higher than the true value. The mass spectrometer gave more precise oxygen consumption results, whereas smaller variance of carbon dioxide production measurements was observed using the infrared CO2 meter.(ABSTRACT TRUNCATED AT 250 WORDS)
Similar articles
-
Day-to-day energy expenditure variability in low birth weight neonates.Pediatr Res. 1987 Jan;21(1):66-71. doi: 10.1203/00006450-198701000-00015. Pediatr Res. 1987. PMID: 3099256
-
[Noninvasive, continuous monitoring of artificial respiration in premature and newborn infants by the constant measurement of respiratory minute volume, oxygen consumption and carbon dioxide release].Klin Padiatr. 1986 Jul-Aug;198(4):321-5. doi: 10.1055/s-2008-1033880. Klin Padiatr. 1986. PMID: 3093754 German.
-
Some mathematical and technical issues in the measurement and interpretation of open-circuit indirect calorimetry in small animals.Int J Obes (Lond). 2006 Sep;30(9):1322-31. doi: 10.1038/sj.ijo.0803280. Epub 2006 Jun 27. Int J Obes (Lond). 2006. PMID: 16801931 Review.
-
Comparison of metabolic monitors in critically ill, ventilated patients.Nutrition. 2006 Nov-Dec;22(11-12):1077-86. doi: 10.1016/j.nut.2006.06.007. Epub 2006 Sep 14. Nutrition. 2006. PMID: 16973331
-
Exhaled gas analysis. Technical and clinical aspects of capnography and oxygen consumption.Crit Care Nurs Clin North Am. 1989 Dec;1(4):669-79. Crit Care Nurs Clin North Am. 1989. PMID: 2516723 Review.
Cited by
-
A new paediatric metabolic monitor.Intensive Care Med. 1994;20(1):51-7. doi: 10.1007/BF02425058. Intensive Care Med. 1994. PMID: 8163761
-
Multiple propane gas burn rates procedure to determine accuracy and linearity of indirect calorimetry systems: an experimental assessment of a method.PeerJ. 2022 Aug 29;10:e13882. doi: 10.7717/peerj.13882. eCollection 2022. PeerJ. 2022. PMID: 36061755 Free PMC article.
-
Validation of a portable indirect calorimetry system for measurement of energy expenditure in sick preterm infants.Arch Dis Child. 1992 Oct;67(10 Spec No):1207-11. doi: 10.1136/adc.67.10_spec_no.1207. Arch Dis Child. 1992. PMID: 1444562 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources