Trends in radiation dose and image quality for pediatric patients with a multidetector CT and a third-generation dual-source dual-energy CT
- PMID: 31004322
- DOI: 10.1007/s11547-019-01037-5
Trends in radiation dose and image quality for pediatric patients with a multidetector CT and a third-generation dual-source dual-energy CT
Abstract
Aim: To provide an overview on dose reduction and image quality after the installation of a third-generation dual-source CT (dsCT) in a Pediatric Radiology Department.
Materials and methods: We included pediatric patients (< 20 years old) undergoing CT for oncological staging (neck, chest and abdomen) or low-dose chest CT for lung diseases. Each of these two groups were further divided in two age groups (≤ or > 10 years old) including patients scanned in the same period of two consecutive years, in 2017 with a 16-row LightSpeed CT (GE Healthcare) or in 2018 with a Somatom Force dsCT (Siemens Healthineers). Technical parameters such as kVp, mAs, slice thickness, exposure times and dose indicators were retrieved and compared. Image quality was evaluated in consensus by two radiologists on a five-point semiquantitative scale. Nonparametric tests were used.
Results: In oncological patients, significantly lower kVp and tube current with better image quality were achieved with the dsCT. Radiation dose (total DLP) was 5-6 times lower with dsCT, thanks also to virtual non-contrast images. In low-dose chest CT, the frequent use of tin filter required higher tube current; a total DLP 3 times lower was achieved with dsCT in patients ≤ 10 years old. The image quality was better with the dsCT in low-dose chest CT protocols.
Conclusion: The third-generation dsCT provides high-quality images with reduced motion artifacts at lower dose.
Keywords: Dose; Dose reduction; Dual-energy CT; Dual-source CT; Pediatric radiology.
Similar articles
-
Ultra-high pitch chest computed tomography at 70 kVp tube voltage in an anthropomorphic pediatric phantom and non-sedated pediatric patients: Initial experience with 3rd generation dual-source CT.Z Med Phys. 2016 Dec;26(4):349-361. doi: 10.1016/j.zemedi.2015.11.002. Epub 2015 Dec 15. Z Med Phys. 2016. PMID: 26702762
-
Diagnosis of Pulmonary Artery Embolism: Comparison of Single-Source CT and 3rd Generation Dual-Source CT using a Dual-Energy Protocol Regarding Image Quality and Radiation Dose.Rofo. 2017 Jun;189(6):527-536. doi: 10.1055/s-0043-103089. Epub 2017 Apr 26. Rofo. 2017. PMID: 28445908 English.
-
Radiation exposure related to cardiovascular CT examination: comparison between conventional 64-MDCT and third-generation dual-source MDCT.Radiol Med. 2019 Aug;124(8):753-761. doi: 10.1007/s11547-019-01036-6. Epub 2019 Apr 22. Radiol Med. 2019. PMID: 31011995
-
Third-generation dual-source dual-energy CT in pediatric congenital heart disease patients: state-of-the-art.Radiol Med. 2019 Dec;124(12):1238-1252. doi: 10.1007/s11547-019-01097-7. Epub 2019 Oct 19. Radiol Med. 2019. PMID: 31630332 Review.
-
Dual-energy CT revisited with multidetector CT: review of principles and clinical applications.Diagn Interv Radiol. 2011 Sep;17(3):181-94. doi: 10.4261/1305-3825.DIR.3860-10.0. Epub 2010 Nov 14. Diagn Interv Radiol. 2011. PMID: 20945292 Review.
Cited by
-
Clinical utility of Dual Energy Computed Tomography in gout: current concepts and applications.Acta Biomed. 2020 Jul 13;91(8-S):116-124. doi: 10.23750/abm.v91i8-S.9942. Acta Biomed. 2020. PMID: 32945286 Free PMC article. Review.
-
Editorial from guest editors current Euratom legislation (DE 59/2013): new patient management in radiation protection.Radiol Med. 2019 Aug;124(8):711-713. doi: 10.1007/s11547-019-01070-4. Epub 2019 Aug 29. Radiol Med. 2019. PMID: 31468300 No abstract available.
-
Spectral CT quantification stability and accuracy for pediatric patients: A phantom study.J Appl Clin Med Phys. 2021 Mar;22(3):16-26. doi: 10.1002/acm2.13161. Epub 2021 Jan 10. J Appl Clin Med Phys. 2021. PMID: 33426801 Free PMC article.
-
Anterior chest wall non-traumatic diseases: a road map for the radiologist.Acta Biomed. 2020 Jul 13;91(8-S):43-50. doi: 10.23750/abm.v91i8-S.9972. Acta Biomed. 2020. PMID: 32945278 Free PMC article. Review.
-
Masses in right side of the heart: spectrum of imaging findings.Acta Biomed. 2020 Jul 13;91(8-S):60-70. doi: 10.23750/abm.v91i8-S.9940. Acta Biomed. 2020. PMID: 32945280 Free PMC article. Review.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical