. 2019 Feb 22;35(2):276-278.
doi: 10.1002/joa3.12170. eCollection 2019 Apr.

Undersensing of ventricular fibrillation by a biventricular implantable cardioverter-defibrillator: What is the cause and the troubleshooting?

Affiliations

Undersensing of ventricular fibrillation by a biventricular implantable cardioverter-defibrillator: What is the cause and the troubleshooting?

Kien Vo Le et al. J Arrhythm. .
No abstract available

Keywords: CRT‐D; dissociation; undersensing; ventricular tachycardia.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflict of interests for this article.

Figures

Figure 1
Figure 1
Event recording by cardioverter‐defibrillator (CRT‐D) Model H197 (Boston Scientific, Marlborough, MA) during DFT testing. (This recording is not a continuous one.)
Figure 2
Figure 2
A, Left ventricular protection period. B, Troubleshooting of this case. LV lead was plugged to the RV IS‐1 port and vice versa (RV IS‐1 to the LV IS‐1 port)

Similar articles

Cited by

References

    1. Bristow MR, Saxon LA, Boehmer J, et al. Cardiac‐resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med. 2004;350:2140–50. - PubMed
    1. Liqun Wu, Jin Qi, Zhang Ning, et al. The effects of acute amiodarone on short‐ and long‐duration ventricular defibrillation threshold in canines. J Cardiovasc Pharmacol. 2011;58:432–8. - PubMed
    1. Dosdall DJ, Osorio J, Robichaux RP, Huang J, Li L, Ideker RE. Purkinje activation preceds myocardial activation following defibrillation after long‐duration ventricular fibrillation. Heart Rhythm. 2010;7:405–12. - PMC - PubMed
    1. Windecker S, Ideker RE, Plumb VJ, Kay GN, Walcott GP, Epstein AE. The influence of ventricular fibrillation duration on defibrillation efficacy using biphasic waveforms in humans. J Am Coll Cardiol. 1999;33:33–38. - PubMed
    1. Poole JE, Johnson GW, Hellkamp AS, et al. Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med. 2008;359:1009–17. - PMC - PubMed