Sensitivity and Specificity of Smartphone-Based Retinal Imaging for Diabetic Retinopathy: A Comparative Study
- PMID: 31014763
- DOI: 10.1016/j.oret.2018.09.016
Sensitivity and Specificity of Smartphone-Based Retinal Imaging for Diabetic Retinopathy: A Comparative Study
Abstract
Purpose: To determine the sensitivity and specificity of a smartphone-based fundus camera, the Remidio Fundus on Phone (FOP; Remidio Innovative Solutions Pvt. Ltd., Bengaluru, India) in detecting diabetic retinopathy (DR) compared with a conventional tabletop fundus camera and clinical examination.
Design: Cross-sectional, single-site, instrument validation study.
Participants: Consecutive patients with diabetes who had no DR (n = 55 eyes), mild to moderate nonproliferative diabetic retinopathy (NPDR; n = 70 eyes), severe NPDR (n = 46 eyes), proliferative diabetic retinopathy (PDR; n = 62 eyes), and diabetic macular edema (DME; n = 44 eyes).
Methods: All participants underwent a dilated examination to determine the grade of DR. Then all participants had mydriatic 45° fundus photographs obtained from three fields of view with the Remidio FOP and a Topcon tabletop fundus camera (Topcon Medical Systems, Inc., Oakland, NJ). Two masked retina specialists graded images for DR and DME, using National Health Service guidelines as well as for image quality using predefined criteria.
Main outcome measure: Sensitivity and specificity of the Remidio FOP for the detection of any DR compared to clinical examination.
Results: One hundred thirty-five subjects (233 eyes) were recruited for the study. Compared with the reference clinical examination, using images from the Remidio FOP, graders 1 and 2 reported a sensitivity of 93.1% (95% confidence interval [CI] = 88.3-96.4) and 94.3% (95% CI = 89.7-97.2) and a specificity of 89.1% (95% CI = 68.2-92.2) and 94.5% (95% CI = 84.9-98.9), respectively, in identifying any DR (κ = 0.55; 95% CI = 0.50-0.57). With images from the Topcon camera, graders reported similar sensitivities and specificities with marginally better agreement (κ = 0.68; 95% CI = 0.67-0.73). The sensitivity of detecting DR gradually increased from R1 to R3 disease using both cameras. Both graders classified significantly fewer images as "ungradable" (2.6%-4.3% for Topcon vs. 1.7%-2.1% for Remidio FOP) and more images as excellent from the Remidio FOP (59%-74%) than the Topcon device (52%-61%).
Conclusions: The Remidio FOP device was found to have high sensitivity and specificity for the detection of any grade of DR with good agreement between graders. The rate of ungradable images was acceptably low and image quality was marginally better with the Remidio FOP.
Copyright © 2018 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Validation of Smartphone Based Retinal Photography for Diabetic Retinopathy Screening.PLoS One. 2015 Sep 24;10(9):e0138285. doi: 10.1371/journal.pone.0138285. eCollection 2015. PLoS One. 2015. PMID: 26401839 Free PMC article.
-
Automated diabetic retinopathy detection in smartphone-based fundus photography using artificial intelligence.Eye (Lond). 2018 Jun;32(6):1138-1144. doi: 10.1038/s41433-018-0064-9. Epub 2018 Mar 9. Eye (Lond). 2018. PMID: 29520050 Free PMC article.
-
Nonmydriatic ultrawide field retinal imaging compared with dilated standard 7-field 35-mm photography and retinal specialist examination for evaluation of diabetic retinopathy.Am J Ophthalmol. 2012 Sep;154(3):549-559.e2. doi: 10.1016/j.ajo.2012.03.019. Epub 2012 May 23. Am J Ophthalmol. 2012. PMID: 22626617
-
Use of Smartphones to Detect Diabetic Retinopathy: Scoping Review and Meta-Analysis of Diagnostic Test Accuracy Studies.J Med Internet Res. 2020 May 15;22(5):e16658. doi: 10.2196/16658. J Med Internet Res. 2020. PMID: 32347810 Free PMC article.
-
Clinical utility of handheld fundus and smartphone-based camera for monitoring diabetic retinal diseases: a review study.Int Ophthalmol. 2024 Feb 9;44(1):41. doi: 10.1007/s10792-024-02975-4. Int Ophthalmol. 2024. PMID: 38334896 Review.
Cited by
-
Detection and Diagnosis of Retinoblastoma: Can Mobile Devices Be the Next Step Toward Early Intervention?Cureus. 2022 Oct 8;14(10):e30074. doi: 10.7759/cureus.30074. eCollection 2022 Oct. Cureus. 2022. PMID: 36381807 Free PMC article. Review.
-
Validity of smartphone-based retinal photography (PEEK-retina) compared to the standard ophthalmic fundus camera in diagnosing diabetic retinopathy in Uganda: A cross-sectional study.PLoS One. 2022 Sep 6;17(9):e0273633. doi: 10.1371/journal.pone.0273633. eCollection 2022. PLoS One. 2022. PMID: 36067194 Free PMC article.
-
The Toronto Tele-Retinal Screening Program for the Elderly in Long-Term Care: A Pilot Project.Clin Ophthalmol. 2024 Dec 20;18:3881-3892. doi: 10.2147/OPTH.S491154. eCollection 2024. Clin Ophthalmol. 2024. PMID: 39720185 Free PMC article.
-
Application of smartphone ophthalmoscope in ophthalmic clinical practice and teaching.Int J Ophthalmol. 2025 May 18;18(5):951-957. doi: 10.18240/ijo.2025.05.22. eCollection 2025. Int J Ophthalmol. 2025. PMID: 40385125 Free PMC article. Review.
-
Accuracy of Integrated Artificial Intelligence Grading Using Handheld Retinal Imaging in a Community Diabetic Eye Screening Program.Ophthalmol Sci. 2023 Dec 15;4(3):100457. doi: 10.1016/j.xops.2023.100457. eCollection 2024 May-Jun. Ophthalmol Sci. 2023. PMID: 38317871 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical