. 2019 Oct;33(10):794-799.
doi: 10.1089/end.2019.0084. Epub 2019 May 27.

The Rise and Fall of High Temperatures During Ureteroscopic Holmium Laser Lithotripsy

Affiliations

The Rise and Fall of High Temperatures During Ureteroscopic Holmium Laser Lithotripsy

Brenton Winship et al. J Endourol. 2019 Oct.

Abstract

Introduction: Temperatures over 43°C-the threshold for cellular injury-may be achieved during ureteroscopic holmium laser lithotripsy. The time to reach and subsequently clear high temperatures at variable laser power settings and irrigation pressures has not been studied. Methods: A flexible or semirigid ureteroscope was placed within an 11/13 F ureteral access sheath inserted into a 250-mL saline bag simulating a normal-caliber ureter, renal pelvis reservoir, and antegrade irrigation flow. A thermocouple was placed adjacent to a 365 μm fiber fired for 45 seconds at 0.6 J/6 Hz, 0.8 J/8 Hz, 1 J/10 Hz, 1 J/20 Hz, and 0.2 J/80 Hz. Irrigation pressures of 200, 100, and 0 mm Hg were tested. Mean temperature changes were recorded with 6°C increase as a threshold for injury (as body temperature is 6°C below 43°C). Results: Semirigid scope: At 200 mm Hg no temperature changes >6°C were observed. At 100 mm Hg, changes >6°C occurred with 1 J/20 Hz within 1 second of activation and returned to ≤6°C within 1 second of cessation. At 0 mm Hg, changes >6°C occurred with all settings; within 1 second at power ≥10 W. Temperatures returned to ≤6°C within 5-10 seconds. Flexible scope: At 200 mm Hg, changes >6°C occurred at 1 J/10 Hz (15 seconds), 0.2 J/80 Hz (3 seconds), and 1 J/20 Hz (2 seconds). Temperatures returned within 6°C of baseline within 2 seconds. At 100 mm Hg, changes >6°C occurred in all but 0.6 J/6 Hz. Temperatures returned to ≤6°C in 5-10 seconds. At 0 mm Hg, all settings produced changes >6°C within 3 seconds, except 0.6 J/6 Hz (35 seconds). Temperatures returned to ≤6°C in under 10 seconds. Conclusions: High temperatures were achieved in our in vitro model in as little as 1 second at common irrigation pressures and laser settings, particularly with a flexible ureteroscope and power ≥10 W. However, with laser cessation, temperatures quickly returned to a safe level at each irrigation pressure.

Keywords: laser; thermal injury; ureteral stone; ureteroscopy instrumentation.