Review
. 2019 Mar;5(1):155-165.
doi: 10.21037/jss.2019.02.04.

What should my hospital buy next?-Guidelines for the acquisition and application of imaging, navigation, and robotics for spine surgery

Affiliations
Review

What should my hospital buy next?-Guidelines for the acquisition and application of imaging, navigation, and robotics for spine surgery

Gregory M Malham et al. J Spine Surg. 2019 Mar.

Abstract

The range of assistive technology options available for spinal fusion surgery has significantly increased. However, surgeons and hospital administrators may lack sufficient information to compare options and make purchasing decisions. We summarize currently available navigation, robotics, and imaging technologies for spinal surgery, highlighting key characteristics, utility, differences, price, and compatibility with other technologies and spinal implants. Guidelines for optimal use and combinations are provided based on surgical approach, operative site, patient anatomy, optimal image quality, and workflow efficiency. Key recommendations include the following. (I) Open-platform navigation and robotics systems that provide surgeons with access to all software and hardware features regardless of implant choice are preferred. (II) Imaging systems that have maximum compatibility with navigation and robotics platforms are optimal. (III) Navigation systems that offer a universal registration mechanism should be standard. (IV) 3D fluoroscopy provides the greatest benefit when speed, operative efficiency, and mobility are required. (V) Intraoperative CT is more useful for imaging long constructs, high BMI, or cervicothoracic anatomy. (VI) Radiation safety awareness that new 3D-fluoroscopy units can deliver radiation comparable to that of CT is needed. (VII) New robotic arm platforms require more clinical and health economic data to justify increased costs.

Keywords: 3D fluoroscopy; imaging; navigation; robotics; spinal fusion.

PubMed Disclaimer

Conflict of interest statement

Conflicts of Interest: GM Malham—Globus (consultancy), NuVasive (consultancy), Stryker (consultancy), Lifehealthcare (travel). T Wells-Quinn—LifeHealthcare Employee, Distributor in ANZ of: Mazor Renaissance, BodyTom, 7D Surgical.

Figures

Figure 1
Figure 1
3D fluoroscopy imaging systems.
Figure 2
Figure 2
Intraoperative mobile CT imaging systems.
Figure 3
Figure 3
Spinal navigation systems.
Figure 4
Figure 4
Spinal robotic platforms.

References

    1. Castro WH, Halm H, Jerosch J, et al. Accuracy of pedicle screw placement in lumbar vertebrae. Spine (Phila Pa 1976) 1996;21:1320-4. 10.1097/00007632-199606010-00008 - DOI - PubMed
    1. Malham GM, Goss B, Blecher C. Percutaneous Pedicle Screw Accuracy with Dynamic Electromyography: The Early Experience of a Traditionally Open Spine Surgeon. J Neurol Surg A Cent Eur Neurosurg 2015;76:303-8. 10.1055/s-0034-1373664 - DOI - PubMed
    1. Oh HS, Kim JS, Lee SH, et al. Comparison between the accuracy of percutaneous and open pedicle screw fixations in lumbosacral fusion. Spine J 2013;13:1751-7. 10.1016/j.spinee.2013.03.042 - DOI - PubMed
    1. Soriano-Sanchez JA, Ortega-Porcayo LA, Gutierrez-Partida CF, et al. Fluoroscopy-guided pedicle screw accuracy with a mini-open approach: a tomographic evaluation of 470 screws in 125 patients. Int J Spine Surg 2015;9:54. 10.14444/2054 - DOI - PMC - PubMed
    1. Wood MJ, McMillen J. The surgical learning curve and accuracy of minimally invasive lumbar pedicle screw placement using CT based computer-assisted navigation plus continuous electromyography monitoring - a retrospective review of 627 screws in 150 patients. Int J Spine Surg 2014;8. - PMC - PubMed