Premature battery depletion due to compromised low-voltage capacitor in a family of defibrillators

Affiliations

Premature battery depletion due to compromised low-voltage capacitor in a family of defibrillators

Jay A Montgomery et al. Pacing Clin Electrophysiol. 2019 Jul.

Abstract

Background: Boston Scientific (Marlborough, MA, USA) implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured between 2008 and 2014 are potentially subject to premature battery depletion through a low-voltage capacitor malfunction occurring as a result of hydrogen buildup within the device. Although some of these devices are currently under advisory, other devices manufactured during this timeframe carry a lower risk of the same malfunction. These same devices are known to have superior longevity in general, and the overall mean lifespan of the devices remains long.

Methods: All patients implanted or followed at our two centers who experienced premature battery depletion and had a Boston Scientific ICD or CRT-D potentially at risk for low-voltage capacitor malfunction were studied retrospectively.

Results: Nineteen out of 838 patients (2.3%) with devices potentially at risk have had premature battery depletion: 5.7% of those under advisory and 1.1% of those not under advisory. None of our patients had compromised therapy, and all had >27 days of projected battery longevity remaining.

Conclusions: Undetected premature battery depletion in this population of ICDs has the potential to expose a patient to an interval of time where the device is unable to provide therapy. However, with enrollment in remote monitoring, regular follow-up, and attention to audible alerts, the risk of therapy loss due to low-voltage state can be effectively mitigated. For these reasons, prophylactic generator replacement is not recommended.

Keywords: battery failure; generator change; high-voltage capacitor; implantable cardioverter-defibrillator; low-voltage capacitor; premature battery depletion.