Enhancing primary implant stability by undersizing implant site preparation: A human cadaver study
- PMID: 31055093
- DOI: 10.1016/j.jormas.2019.04.016
Enhancing primary implant stability by undersizing implant site preparation: A human cadaver study
Abstract
Introduction: The aims of this human cadaver study were to analyze the primary stability of dental implants inserted by using different methods of preparation and to explore correlations among the bone density, peak insertion torque and resonance frequency analysis values.
Materials and methods: Forty-eight implants were placed into six human cadaver mandibles. The bone density of each implant recipient site was preoperatively measured by using computerized tomography (CT) in Hounsfield Units (HU). Three test groups were created according to the final drill size. The diameters of the final twist drills were 3 mm, 3.4 mm, and 3.8 mm in group T1, group T2, and group T3, respectively. The peak insertion torque and resonance frequency analysis values were recorded.
Results: The mean bone density, insertion torque, and RFA values were 285 ± 90 HU, 42.8 ± 5 Ncm, and 74 ± 6 ISQ for all 48 implants, respectively. Statistically significant correlations were noted between bone density and insertion torque values (P < 0.05), bone density and RFA values (P < 0.01), and insertion torque and RFA values (P < 0.001). When considering the three test groups, the mean insertion torque, and RFA values were 47.1 ± 3 Ncm, and 83.1 ± 6 ISQ in Group T1, and 43.4 ± 5 Ncm and 73.4 ± 6 ISQ in Group T2, and 37.9 ± 6 Ncm, and 65.3 ± 5 ISQ in Group T3.
Conclusion: The results of this study suggest that undersized implant site preparations may be helpful in enhancing primary implant stability and that there are significant correlations among bone density, maximum insertion torque, and RFA values.
Keywords: Bone density; Cadaver; Implant; Insertion torque; Primary stability; Resonance frequency analysis.
Copyright © 2019 Elsevier Masson SAS. All rights reserved.
Similar articles
-
Biomechanical aspects of primary implant stability: a human cadaver study.Clin Implant Dent Relat Res. 2009 Jun;11(2):113-9. doi: 10.1111/j.1708-8208.2008.00097.x. Epub 2008 Apr 16. Clin Implant Dent Relat Res. 2009. PMID: 18422713
-
Sensitivity of resonance frequency analysis method to assess implant stability.N Y State Dent J. 2011 Aug-Sep;77(5):44-9. N Y State Dent J. 2011. PMID: 22029115
-
Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: a human cadaver study.Clin Implant Dent Relat Res. 2009 Mar;11(1):52-8. doi: 10.1111/j.1708-8208.2008.00095.x. Epub 2008 Apr 1. Clin Implant Dent Relat Res. 2009. PMID: 18384400
-
Implant Insertion Torque: Its Role in Achieving Primary Stability of Restorable Dental Implants.Compend Contin Educ Dent. 2017 Feb;38(2):88-95; quiz 96. Compend Contin Educ Dent. 2017. PMID: 28156122 Review.
-
Implant insertion torque and marginal bone loss: A systematic review and meta-analysis.Int J Oral Implantol (Berl). 2020;13(4):345-353. Int J Oral Implantol (Berl). 2020. PMID: 33491366
Cited by
-
Primary Implant Stability Analysis of Different Dental Implant Connections and Designs-An In Vitro Comparative Study.Materials (Basel). 2022 Apr 23;15(9):3072. doi: 10.3390/ma15093072. Materials (Basel). 2022. PMID: 35591407 Free PMC article.
-
Influence of different drilling protocols and bone density on the insertion torque of dental implants.Med Oral Patol Oral Cir Bucal. 2023 Jul 1;28(4):e385-e394. doi: 10.4317/medoral.25804. Med Oral Patol Oral Cir Bucal. 2023. PMID: 37330951 Free PMC article.
-
Does the Modification of the Apical Geometry of a Dental Implant Affect Its Primary Stability? A Comparative Ex Vivo Study.Materials (Basel). 2021 Apr 1;14(7):1728. doi: 10.3390/ma14071728. Materials (Basel). 2021. PMID: 33915933 Free PMC article.
-
Reliability and Correlation of Different Devices for the Evaluation of Primary Implant Stability: An In Vitro Study.Materials (Basel). 2021 Sep 24;14(19):5537. doi: 10.3390/ma14195537. Materials (Basel). 2021. PMID: 34639933 Free PMC article.
-
Comparison of Primary Stability of Implants Installed by Two Different Methods in D3 and D4 Bone Types: An In Vitro Study.J Int Soc Prev Community Dent. 2020 Sep 28;10(5):620-626. doi: 10.4103/jispcd.JISPCD_160_20. eCollection 2020 Sep-Oct. J Int Soc Prev Community Dent. 2020. PMID: 33282772 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources