. 2019 Nov;74(11):1374-1380.
doi: 10.1111/anae.14697. Epub 2019 May 7.

Characterisation of aluminium release by the enFlow® fluid-warming system in crystalloids and blood products

Affiliations
Free article

Characterisation of aluminium release by the enFlow® fluid-warming system in crystalloids and blood products

M H Taylor et al. Anaesthesia. 2019 Nov.
Free article

Abstract

The use of uncoated aluminium-heated plates in an intravenous fluid-warming system has been shown to produce high levels of aluminium in Sterofundin 1/1E, a balanced crystalloid solution. However, the effect of this fluid-warming device on other balanced crystalloid solutions and blood products has not been studied. Using mass spectrometry we measured aluminium levels in Plasma-Lyte 148, compound sodium lactate solution, 4% human albumin solution, expired resuspended packed red cells and fresh frozen plasma that were pumped through an enFlow® fluid-warming system at 2 ml.min-1 . Samples were taken at baseline before heating and then at 10-min intervals up to 60 min with the system set to warm the fluids to 40 °C. High concentrations of aluminium were found for Plasma-Lyte 148 and compound sodium lactate solutions (mean (SD) 223 (0.6) μmol.l-1 and 163 (0.2) μmol.l-1 at 60 min, respectively); both concentrations were significantly greater than the United States Food and Drug Administration recommended maximum limit for aluminium in intravenous nutrition of 25 μg.l-1 (0.9 μmol.l-1 ). Lower aluminium levels were found in 4% human albumin solutions, expired resuspended red cells and fresh frozen plasma at 60 min (mean (SD) 5.7 (0.1) μmol.l-1 , 2.7 (0.0) μmol.l-1 and 2.3 (0.4) μmol.l-1 , respectively). The process allowing addition of aluminium to be added to Sterofundin 1/1E by the enFlow fluid warmer also occurs in Plasma-Lyte 148 and compound sodium lactate solutions and to a lesser degree in blood products. The exact mechanism facilitating this process and its clinical significance remain unclear.

Keywords: aluminium; blood product; crystalloid; fluid warming; hypothermia; prevention.

PubMed Disclaimer

Comment in

  • Fluid warmers, aluminium and what the regulator did.
    McGuire N, Kelly M, Mustafa E, Hannon J. McGuire N, et al. Anaesthesia. 2019 Nov;74(11):1354-1356. doi: 10.1111/anae.14743. Epub 2019 Jul 11. Anaesthesia. 2019. PMID: 31297796 Free PMC article. No abstract available.

References

    1. Perl T, Kunze-Szikszay N, Bräuer A, et al. Aluminium release by coated and uncoated fluid-warming devices. Anaesthesia 2019; 112: 166.
    1. NICE. Hypothermia: prevention and management in adults having surgery. Clinical guideline [CG65]. In: NICE Clinical Guideline. 2016. https://www.nice.org.uk/guidance/cg65: NICE Clinical Guideline. (accessed 19/03/2019).
    1. Food and Drug, Administration. Aluminum in large and small volume parenterals used in total parenteral nutrition. FDA Federal Register 1998, 63: 176-85.
    1. Bell JD, Kubal G, Radulovic S, Sadler PJ, Tucker A. Detection of aluminium(III) binding to citrate in human blood plasma by proton nuclear magnetic resonance spectroscopy. Analyst 1993; 118: 241-4.
    1. Froment DP, Molitoris BA, Buddington B, Miller N, Alfrey AC. Site and mechanism of enhanced gastrointestinal absorption of aluminum by citrate. Kidney International 1989; 36: 978-84.