Imaging evaluation of polyethylene liner dissociation in total hip arthroplasty
- PMID: 31104144
- DOI: 10.1007/s00256-019-03232-x
Imaging evaluation of polyethylene liner dissociation in total hip arthroplasty
Abstract
Objective: To describe the imaging findings of polyethylene liner dissociation in total hip arthroplasty.
Materials and methods: Retrospective search of our institution's radiology database identified 12 patients with polyethylene liner dissociation of a total hip arthroplasty. Clinical and operative notes were reviewed. All radiological studies were reviewed independently by two radiologists.
Results: Among 12 patients (seven females/five males; mean age: 67 years; median interval after surgery at diagnosis: 8.5 months) with polyethylene liner dissociation, 11 had radiographs, six had CT, seven had MRI, and two had arthrography. "Bubble sign" and "crescent sign" on radiography were insensitive, seen only in three patients, but all showed abrupt eccentric positioning of the femoral head. CT identified the dislocated liner in five of six patients but failed to identify one liner, which was partially associated with the acetabular cup. MRI identified the dislocated liner in all seven patients, with the MAVRIC (multiacquisition variable resonance image combination) sequences either increasing the diagnostic confidence relative to the routine pulse sequences or being the sole sequences in which the liner can be identified if the liner remained partially associated with the acetabular cup. Arthrography identified the dislocated liner in one of two patients.
Conclusions: Previously described radiographic signs of polyethylene liner dissociation are insensitive, but abrupt eccentric positioning of the femoral head in the correct clinical context is highly suggestive of the diagnosis. MRI and CT can accurately localize the displaced liner. MAVRIC is particularly helpful if the dislocated liner remains partially associated with the acetabular cup.
Keywords: Dual-mobility arthroplasty; Intraprosthetic dislocation; MAVRIC; Polyethylene liner; Total hip arthroplasty.
Similar articles
-
The tram track sign: a characteristic sonographic feature of polyethylene liner dissociation after total-hip arthroplasty.J Ultrasound Med. 2014 Nov;33(11):1931-7. doi: 10.7863/ultra.33.11.1931. J Ultrasound Med. 2014. PMID: 25336480
-
The crescent sign: dissociation of the polyethylene liner from a modular acetabular component in total hip arthroplasty.Skeletal Radiol. 2005 Oct;34(10):620-4. doi: 10.1007/s00256-005-0927-6. Epub 2005 Jun 25. Skeletal Radiol. 2005. PMID: 15981009
-
Correlation of Cup Inclination Angle with Liner Wear for Metal-on-polyethylene in Hip Primary Arthroplasty.Orthop Surg. 2017 May;9(2):186-190. doi: 10.1111/os.12337. Epub 2017 May 30. Orthop Surg. 2017. PMID: 28557297 Free PMC article.
-
Dissociation of the S-ROM metal-backed polyethylene acetabular liner--a case report and literature review.Acta Orthop Belg. 2003;69(1):86-8. Acta Orthop Belg. 2003. PMID: 12666298 Review.
-
Dissociation and Intrapelvic Entrapment of a Dual-mobility Polyethylene Component.Clin Orthop Relat Res. 2016 Apr;474(4):1072-6. doi: 10.1007/s11999-015-4381-z. Epub 2015 Jun 19. Clin Orthop Relat Res. 2016. PMID: 26088765 Free PMC article. Review.
Cited by
-
Polyethylene Liner Dissociation in the Early Post-Operative Period Following Total Hip Arthroplasty: A Case Report and Literature Review.Cureus. 2022 May 18;14(5):e25119. doi: 10.7759/cureus.25119. eCollection 2022 May. Cureus. 2022. PMID: 35733505 Free PMC article.
-
Intraprosthetic Dislocation Following Dual Mobility Total Hip Arthroplasty: A Retrieval Analysis Study.Arthroplast Today. 2024 Dec 20;31:101596. doi: 10.1016/j.artd.2024.101596. eCollection 2025 Feb. Arthroplast Today. 2024. PMID: 39811773 Free PMC article.
-
Recurrent and radiographically unrecognized iatrogenic intra-prosthesis hip dislocations.J Am Coll Emerg Physicians Open. 2020 May 10;1(4):419-422. doi: 10.1002/emp2.12050. eCollection 2020 Aug. J Am Coll Emerg Physicians Open. 2020. PMID: 33000065 Free PMC article.
-
Intraprosthetic dislocation of dual-mobility total hip arthroplasty implant.Radiol Case Rep. 2023 Aug 30;18(11):3964-3967. doi: 10.1016/j.radcr.2023.08.070. eCollection 2023 Nov. Radiol Case Rep. 2023. PMID: 37680661 Free PMC article.
-
Intraprosthetic Dislocation of a Modular Dual-Mobility Total Hip Arthroplasty With Posterior Migration of the Polyethylene Liner.Arthroplast Today. 2024 Jan 16;25:101290. doi: 10.1016/j.artd.2023.101290. eCollection 2024 Feb. Arthroplast Today. 2024. PMID: 38298810 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical