False-positive troponin I elevation in a newborn with neonatal encephalopathy
Abstract
Serum biomarkers of myocardial damage are commonly used in babies after perinatal asphyxia. We present a case report of a persistently troponin I elevation without evidence of clinical or instrumental signs of myocardial ischaemia in a baby with perinatal asphyxia. When the blood was mixed with polyethylene glycol we found that the troponin I levels were falsely elevated due to interfering antibodies. This case shows that analytical errors may still occur despite modern immunoassay systems and underlines the need for further investigations to identify false-positive values in case of disagreement between clinical conditions and laboratory values.
Keywords: ischaemic heart disease; neonatal intensive care; neurological injury.
© BMJ Publishing Group Limited 2019. No commercial re-use. See rights and permissions. Published by BMJ.
Conflict of interest statement
Competing interests: None declared.
Similar articles
-
Use of high-sensitivity cardiac troponin I levels for early diagnosis of myocardial injury after neonatal asphyxia.J Int Med Res. 2019 Jul;47(7):3234-3242. doi: 10.1177/0300060519831187. Epub 2019 Jun 13. J Int Med Res. 2019. PMID: 31189368 Free PMC article.
-
A patient with an increased troponin level without evidence of ischaemic cardiac injury.Hong Kong Med J. 2004 Aug;10(4):277-9. Hong Kong Med J. 2004. PMID: 15299174
-
Cardiac biomarkers in neonatal hypoxic ischaemia.Acta Paediatr. 2012 Apr;101(4):338-43. doi: 10.1111/j.1651-2227.2011.02539.x. Epub 2011 Dec 23. Acta Paediatr. 2012. PMID: 22118561 Review.
-
Is serum troponin T a useful marker of myocardial damage in newborn infants with perinatal asphyxia?Acta Paediatr. 2007 Feb;96(2):181-4. doi: 10.1111/j.1651-2227.2007.00104.x. Acta Paediatr. 2007. PMID: 17429901
-
Troponin.Neonatal Netw. 2013 Sep-Oct;32(5):365-8. doi: 10.1891/0730-0832.32.5.365. Neonatal Netw. 2013. PMID: 23985475 Review.
Cited by
-
Unexpected high troponin T and I values in a child with hypertrophic cardiomyopathy and acute chest pain: a case report.Eur Heart J Case Rep. 2023 Aug 2;7(8):ytad375. doi: 10.1093/ehjcr/ytad375. eCollection 2023 Aug. Eur Heart J Case Rep. 2023. PMID: 37575535 Free PMC article.
-
Case report: False-positive B-type natriuretic peptide results with the cyclic amplified fluorescence immunoassay.J Int Med Res. 2025 Mar;53(3):3000605251325942. doi: 10.1177/03000605251325942. Epub 2025 Mar 22. J Int Med Res. 2025. PMID: 40119688 Free PMC article.
-
Predictive value of cord blood myocardial enzyme and troponin levels for myocardial injury after neonatal asphyxia.Am J Transl Res. 2023 Jan 15;15(1):241-248. eCollection 2023. Am J Transl Res. 2023. PMID: 36777838 Free PMC article.
-
Characterization of canine anti-mouse antibodies highlights that multiple strategies are needed to combat immunoassay interference.Sci Rep. 2019 Oct 10;9(1):14521. doi: 10.1038/s41598-019-51228-3. Sci Rep. 2019. PMID: 31601945 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources