Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2019 Jun;477(6):1315-1321.
doi: 10.1097/CORR.0000000000000710.

Are Hooded, Crosslinked Polyethylene Liners Associated with a Reduced Risk of Revision After THA?

Affiliations

Are Hooded, Crosslinked Polyethylene Liners Associated with a Reduced Risk of Revision After THA?

Adrian Bauze et al. Clin Orthop Relat Res. 2019 Jun.

Abstract

Background: Hooded acetabular liners and head sizes ≥ 32 mm have both shown to have a beneficial effect on the revision rate for dislocation in THA. Experience with noncrosslinked polyethylene (nonXLPE) raised concerns regarding the risk of impingement damage, loosening, and osteolysis with hooded liners; however, the evidence for this in crosslinked polyethylene (XLPE) is inconclusive. The interaction between different femoral head sizes and hooded liners is not well understood, and it is unclear whether hooded XLPE liners have a beneficial effect on overall long-term survivorship.

Questions/purposes: We analyzed a large national joint registry to ask: (1) Is the use of hooded XLPE liners associated with a reduced revision rate for dislocation compared with nonhooded liners? (2) Is there a difference in the revision rate for aseptic loosening/osteolysis? (3) Is head size associated with any difference in the revision rate between hooded and nonhooded liners?

Methods: The Australian Orthopaedic Association National Joint Replacement Registry longitudinally maintains data on all primary and revision joint arthroplasties with nearly 100% capture. We analyzed all conventional primary THAs performed from registry inception in September 1999 until December 31, 2016 in patients with a diagnosis of osteoarthritis who had nonhooded or hooded XLPE bearings in a cementless acetabular shell. The study group included 192,659 THA procedures with XLPE liners, of which 67,904 were nonhooded and 124,755 were hooded. The mean age of patients receiving nonhooded liners was 70 years (range, 11-100 years); 44% were males. This was similar to the patients with hooded liners, who had a mean age of 70 years (range, 16-100 years); 45% were males. The main outcome measure was the cumulative percent revision at 15 years of the THA using Kaplan-Meier estimates of survivorship. We examined reasons for revision and and performed multivariable analysis to control for the confounding factors of three head size groups (< 32mm, 32mm, and > 32mm) and for the method of femoral fixation.

Results: There was a higher revision rate for dislocation for patients with nonhooded liners at all times to 15 years (HR, 1.31; 95% CI, 1.17-1.47; p < 0.001). There was a higher revision rate for the diagnosis of aseptic loosening/osteolysis with patients with nonhooded liners compared with hooded liners (HR, 1.19; 95% CI, 1.05-1.34; p = 0.006). Head sizes of 32 mm or larger were independently associated with a lower comparative revision rate between hooded and nonhooded liners, but this was not apparent for head sizes smaller than 32 mm. It appeared that the main driver of the finding in larger heads was a reduced dislocation risk with hooded liners for 32 mm heads (HR, 1.50; 95% CI, 1.23-1.80; p < 0.001) and for heads larger than 32 mm (HR, 1.50; 95% CI, 1.20-1.89; p < 0.001).

Conclusions: Prior research has suggested that hooded acetabular liners may be associated with impingement, loosening, and osteolysis; however, in this large, registry-based report we found that XLPE hooded liners are not associated with an increased revision rate for aseptic loosening/osteolysis. Although there are many potential confounding variables in this registry analysis, if anything, surgeons using larger femoral heads and hooded liners likely did so in patients with a higher perceived dislocation risk. Patients with larger heads and XLPE hooded liners were, however, less likely to experience revision for dislocation. These liners therefore appear reasonable to use in primary THA at the surgeon's discretion.

Level of evidence: Level III, therapeutic study.

PubMed Disclaimer

Conflict of interest statement

Each author certifies that he or she has no commercial associations (eg consultancies, stock ownership, equity interest, patent/licensing arrangements etc) that might pose a conflict of interest in connection with the submitted article.

All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research® editors and board members are on file with the publication and can be viewed on request.

Figures

Fig. 1
Fig. 1
The cumulative percent revision for dislocation of primary conventional THA by acetabular hood type for patients with a primary diagnosis of osteoarthritis is shown.
Fig. 2
Fig. 2
The cumulative percent revision for loosening/osteolysis of primary conventional THA by acetabular hood type for patients with a primary diagnosis of osteoarthritis is shown.
Fig. 3
Fig. 3
The cumulative percent revision for dislocation of primary conventional THA with head size 32 mm by acetabular hood type for patients with a primary diagnosis of osteoarthritis is shown.
Fig. 4
Fig. 4
The cumulative percent revision for dislocation of primary conventional THA with head size greater than 32 mm by acetabular hood type for patients with a primary diagnosis of osteoarthritis is shown.

Comment in

Similar articles

Cited by

References

    1. Amstutz HC, Lodwig RM, Schurman DJ, Hodgson AG. Range of motion studies for total hip replacements. A comparative study with a new experimental apparatus. Clin Orthop Relat Res. 1975;111:124-30 - PubMed
    1. Australian Orthopaedic Association National Joint Replacement Registry. Annual Report. Adelaide: AOA; 2018:15, 135 Available at: https://aoanjrr.sahmri.com/annual-reports-2018 Accessed January 30, 2019.
    1. Biau DJ, Latouche A, Porcher R. Competing events influence estimated survival probability: when is Kaplan-Meier analysis appropriate? Clin Orthop Relat Res. 2007;462:229-233. - PubMed
    1. Brooks PJ. Dislocation following total hip replacement: causes and cures. Bone Joint J. 2013;95-B(11 Suppl A):67-69. - PubMed
    1. Charnley J. Low friction Arthroplasty of the Hip: Theory and Practice. Spinger; New York, 1979:315-316.