Performance of an in-vivo, continuous blood-gas monitor with disposable probe
- PMID: 3113766
Performance of an in-vivo, continuous blood-gas monitor with disposable probe
Abstract
A fluorescence-based fiber optic measurement system has been developed for monitoring pH, pCO2, and pO2 through a 20-gauge radial artery catheter without compromising capabilities for monitoring arterial pressure or for blood withdrawal. The measuring probe consists of three optical fibers to which the sensing chemistries are attached, and a thermocouple that measures temperature. The probe is designed to meet in-vivo biocompatibility requirements for a one-time use of up to 72 h. The components that are in contact with the patient's blood are nontoxic, nonhemolytic, nonthrombogenic, and sterilizable. Blood compatibility is enhanced by including covalently bound heparin. The in-vitro accuracy of the system has been tested against commercial blood-gas measurement instruments; comparison with tonometry and blood gas values gave r greater than or equal to 0.98 for all three sensors. The standard error for all sensors was within the College of American Pathologists' accuracy guidelines for measuring blood gas. Instrumental drift was minimal, indicating that system performance characteristics should not be the limiting factor in obtaining clinically useful information for up to 72 h. The response times of the sensors in animal and in-vitro studies were less than 2 min, suitable for monitoring physiological changes in blood gas values.
Similar articles
-
Accuracy and utility of a continuous intra-arterial blood gas monitoring system in pediatric patients.Crit Care Med. 2001 Feb;29(2):420-6. doi: 10.1097/00003246-200102000-00037. Crit Care Med. 2001. PMID: 11246326
-
Clinical performance of a blood gas monitor: a prospective, multicenter trial.Crit Care Med. 1993 Apr;21(4):487-94. doi: 10.1097/00003246-199304000-00005. Crit Care Med. 1993. PMID: 8472565 Clinical Trial.
-
Multisite evaluation of a continuous intraarterial blood gas monitoring system.Anesthesiology. 1994 Sep;81(3):543-52. doi: 10.1097/00000542-199409000-00005. Anesthesiology. 1994. PMID: 8092498 Clinical Trial.
-
[Intravascular electrodes for blood gas monitoring].Iyodenshi To Seitai Kogaku. 1983 Dec;21(7):517-23. Iyodenshi To Seitai Kogaku. 1983. PMID: 6431155 Review. Japanese. No abstract available.
-
Emerging technology in critical care: continuous intra-arterial blood gas monitoring.Am J Crit Care. 1996 Jan;5(1):55-65. Am J Crit Care. 1996. PMID: 8680494 Review.
Cited by
-
Synthesis and photophysical and electrochemical properties of functionalized mono-, bis-, and trisanthracenyl bridged Ru(II) bis(2,2':6',2"-terpyridine) charge transfer complexes.ScientificWorldJournal. 2014;2014:570864. doi: 10.1155/2014/570864. Epub 2014 Apr 30. ScientificWorldJournal. 2014. PMID: 24883408 Free PMC article.
-
Continuous assessment of arterial blood gases.Crit Care. 1997;1(1):11-14. doi: 10.1186/cc2. Crit Care. 1997. PMID: 11094462 Free PMC article. No abstract available.
-
Continuous intra-arterial blood gas monitoring.Intensive Care Med. 1996 Aug;22(8):818-28. doi: 10.1007/BF01709527. Intensive Care Med. 1996. PMID: 8880253 Review.
-
The PB3300 intraarterial blood gas monitoring system.J Clin Monit. 1994 Jan;10(1):59-66. doi: 10.1007/BF01651467. J Clin Monit. 1994. PMID: 8126540
-
Intra-arterial blood gas monitoring system: more accurate values can be obtained.J Clin Monit. 1996 Mar;12(2):141-7. doi: 10.1007/BF02078134. J Clin Monit. 1996. PMID: 8823634
MeSH terms
Substances
LinkOut - more resources
Other Literature Sources