High-attenuation artifact reduction in breast tomosynthesis using a novel reconstruction algorithm
- PMID: 31153567
- DOI: 10.1016/j.ejrad.2019.04.014
High-attenuation artifact reduction in breast tomosynthesis using a novel reconstruction algorithm
Abstract
Purpose: To assess the effect on reducing the out-of-plane artifacts from metal objects in breast tomosynthesis (BT) using a novel artifact-reducing reconstruction algorithm in specimen radiography.
Methods and materials: The study was approved by the Regional Ethical Review Board. BT images of 18 partial- and whole mastectomy specimens from women with breast cancer were acquired before and after a needle was inserted close to the lesion. The images were reconstructed using both a standard reconstruction algorithm, and a novel algorithm; the latter uses pre-segmentation to remove highly attenuating artifact-inducing objects from projection images before reconstruction. Images were separately reconstructed with and without segmentation, and combined into an artifact-reduced reconstruction. Standard and artifact-reduced BT-algorithms were compared visually and quantitatively using clinical images of mastectomy specimens and a physical anthropomorphic phantom. Six readers independently assessed the visibility of the lesion with and without artifact-reduction in a side-by-side comparison. A quantitative analysis was performed, comparing the signal-difference to background ratio (SDBR) and artifact spread function (ASF) between the two reconstruction methods.
Results: The magnitude of out-of-plane artifacts was clearly reduced with the novel reconstruction compared to BT-images without artifact reduction. Lesion masking by artifacts was largely averted; tumour visibility was comparable to standard BT images without a needle. In 76 ± 8% (standard deviation) of cases overall, readers could confidently state needle location. The same figure was 94 ± 6% for whole mastectomy cases, compared to 62 ± 17% for partial mastectomies. With metal artifact reduction, SDBR increased by 97% in the phantom, and by 69% in the mastectomies. The artifact spread function was substantially narrower.
Conclusion: Artifact reduction in BT using a novel reconstruction method enables qualitatively and quantitatively improved clinical use of BT when metal artifacts can be a limiting factor such as in tomosynthesis-guided biopsy.
Keywords: Artifact reduction; Breast imaging; Breast tomosynthesis; Image reconstruction; Mammography.
Copyright © 2019. Published by Elsevier B.V.
Similar articles
-
Characterization of a constrained paired-view technique in iterative reconstruction for breast tomosynthesis.Med Phys. 2013 Oct;40(10):101901. doi: 10.1118/1.4819943. Med Phys. 2013. PMID: 24089903
-
A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis.Med Phys. 2006 Oct;33(10):3781-95. doi: 10.1118/1.2237543. Med Phys. 2006. PMID: 17089843 Free PMC article.
-
Improving image quality for digital breast tomosynthesis: an automated detection and diffusion-based method for metal artifact reduction.Phys Med Biol. 2017 Sep 15;62(19):7765-7783. doi: 10.1088/1361-6560/aa8803. Phys Med Biol. 2017. PMID: 28832336 Free PMC article.
-
Evaluation of back projection methods for breast tomosynthesis image reconstruction.J Digit Imaging. 2015 Jun;28(3):338-45. doi: 10.1007/s10278-014-9736-6. J Digit Imaging. 2015. PMID: 25384538 Free PMC article. Review.
-
A review of breast tomosynthesis. Part II. Image reconstruction, processing and analysis, and advanced applications.Med Phys. 2013 Jan;40(1):014302. doi: 10.1118/1.4770281. Med Phys. 2013. PMID: 23298127 Free PMC article. Review.
Cited by
-
Digital breast tomosynthesis system concept addressing the needs in breast cancer screening and diagnosis.J Med Imaging (Bellingham). 2025 Jan;12(Suppl 1):S13010. doi: 10.1117/1.JMI.12.S1.S13010. Epub 2024 Dec 17. J Med Imaging (Bellingham). 2025. PMID: 39703347
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical