Dual-layer spectral computerized tomography for metal artifact reduction: small versus large orthopedic devices
- PMID: 31154493
- DOI: 10.1007/s00256-019-03248-3
Dual-layer spectral computerized tomography for metal artifact reduction: small versus large orthopedic devices
Abstract
Introduction: Metal artifacts limit the diagnostic utility of computerized tomography (CT) for implant-related complications. Dual-layer spectral detector CT imaging has been suggested for artifact reduction. Our objective was to evaluate the utility of spectral CT in artifact reduction in patients with small and large metal implants.
Methods: In this prospective study, patients with metallic orthopedic implants underwent CT imaging using a prototype spectral detector CT scanner. Conventional images were generated with iterative reconstruction at 120 kVp, and virtual monochromatic images were generated at 20-keV intervals between 40 to 200 keV. Conventional and monochromatic images were compared quantitatively using signal-to-noise ratio (SNR) and artifact improvement. Qualitative analysis was performed independently by two musculoskeletal radiologists and included six image quality indicators.
Results: A total of 12 patients were scanned. In monochromatic images, as the energy level increased, the artifact size decreased progressively (p < 0.01). When conventional and monochromatic images were compared, maximum reduction was seen at 200 keV. Using qualitative assessments, 160 and 180 keV levels had the best overall diagnostic image quality. With increased energy level, there was improvement in qualitative ratings of bone-metal interface conspicuity (p = 0.002), degree of streak artifact (p = 0.010) and trabecular bone definition at 1 cm from implant (p = 0.023), and a trend towards significance for bone definition at 5 cm, soft tissue detail and overall diagnostic quality. Subgroup analysis revealed superior artifact reduction in small implants compared to large hardware.
Discussion: Our results support the utility of dual-layer spectral CT in metal artifact reduction. Virtual monochromatic images were diagnostically superior, especially for smaller implants. Virtual monoenergetic images at 160-180 keV are ideal for reducing artifacts.
Keywords: Artifact; Dual-energy computed tomography; Dual-layer detector; Hip replacement arthroplasty; Metal implant; Spinal instrumentation; Virtual monoenergetic images.
Similar articles
-
Quantifying metal artefact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses.Eur J Radiol. 2017 Mar;88:61-70. doi: 10.1016/j.ejrad.2017.01.002. Epub 2017 Jan 4. Eur J Radiol. 2017. PMID: 28189210
-
Metal Artifact Reduction in Virtual Monoenergetic Spectral Dual-Energy CT of Patients With Metallic Orthopedic Implants in the Distal Radius.AJR Am J Roentgenol. 2018 Nov;211(5):1083-1091. doi: 10.2214/AJR.18.19514. Epub 2018 Sep 21. AJR Am J Roentgenol. 2018. PMID: 30240300
-
Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms.Skeletal Radiol. 2018 Feb;47(2):195-201. doi: 10.1007/s00256-017-2776-5. Epub 2017 Sep 21. Skeletal Radiol. 2018. PMID: 28932962
-
Reduction of Metal Artifact with Dual-Energy CT: Virtual Monospectral Imaging with Fast Kilovoltage Switching and Metal Artifact Reduction Software.Semin Musculoskelet Radiol. 2015 Dec;19(5):446-55. doi: 10.1055/s-0035-1569256. Epub 2015 Dec 22. Semin Musculoskelet Radiol. 2015. PMID: 26696083 Review.
-
Metal artifact reduction techniques in musculoskeletal CT-imaging.Eur J Radiol. 2018 Oct;107:60-69. doi: 10.1016/j.ejrad.2018.08.010. Epub 2018 Aug 12. Eur J Radiol. 2018. PMID: 30292274 Review.
Cited by
-
Dual-Energy Computed Tomography Applications to Reduce Metal Artifacts in Hip Prostheses: A Phantom Study.Diagnostics (Basel). 2022 Dec 23;13(1):50. doi: 10.3390/diagnostics13010050. Diagnostics (Basel). 2022. PMID: 36611342 Free PMC article.
-
Reduction of artifacts from contrast media in spectral-detector CT by combined use of virtual monoenergetic images and orthopedic metal-artifact-reduction algorithm.Medicine (Baltimore). 2025 May 16;104(20):e42370. doi: 10.1097/MD.0000000000042370. Medicine (Baltimore). 2025. PMID: 40388764 Free PMC article.
-
Updates on the Applications of Spectral Computed Tomography for Musculoskeletal Imaging.Diagnostics (Basel). 2024 Mar 29;14(7):732. doi: 10.3390/diagnostics14070732. Diagnostics (Basel). 2024. PMID: 38611645 Free PMC article. Review.
-
Reduction of Metal Artifacts Caused by Titanium Peduncular Screws in the Spine by Means of Monoenergetic Images and the Metal Artifact Reduction Software in Dual-Energy Computed Tomography.J Med Phys. 2022 Apr-Jun;47(2):152-158. doi: 10.4103/jmp.jmp_121_21. Epub 2022 Aug 5. J Med Phys. 2022. PMID: 36212200 Free PMC article.
-
Evaluation of peri-prosthetic radiolucent lines surrounding the cementless femoral stem using digital tomosynthesis with metal artifact reduction: a cadaveric study in comparison with radiography and computed tomography.Quant Imaging Med Surg. 2020 Sep;10(9):1786-1800. doi: 10.21037/qims-19-1018. Quant Imaging Med Surg. 2020. PMID: 32879857 Free PMC article.