Nasolaryngeal Distances in the Adult Population and an Evaluation of Commercially Available Nasotracheal Tubes
- PMID: 31162158
- DOI: 10.1213/ANE.0000000000004241
Nasolaryngeal Distances in the Adult Population and an Evaluation of Commercially Available Nasotracheal Tubes
Abstract
Background: Preformed nasal endotracheal tubes (NETs) come with a predefined insertion depth due to their curved design. While size indication refers to internal diameter, there is a considerable variability in the corresponding lengths and proportions of same-sized tubes of different manufacturers which is probably based on the lack of data of nasolaryngeal distances (NLDs) in the adult population. Choosing the best-fitting NET is therefore difficult and carries the risk of endobronchial intubation or, on the contrary, cuff inflation at the vocal cord level. The aim of this study was to develop a prediction model for NLD and a selection guide to choose the appropriate NET based on a radiographic description of NLD in comparison to the measurements of available NETs of several manufacturers.
Methods: After institutional ethics board review, 388 computed tomography (CT) scan images of head, neck, and upper thorax in a heterogeneous adult cohort were included. Mean distances from the nares to the lower border of the thyroid cartilage were measured. NETs from different manufacturers were measured and compared to the NLD derived from the radiographic analysis. The patients' sex, body height, and weight were considered as possible covariates in quantile regression models for predicting the NLD.
Results: Data from 200 patients were analyzed. NLD was associated with sex, body height, and weight. A simple quantile regression model using the body height as the only covariate sufficed to achieve accurate predictions of NLD. Validation on independent test data showed that 92.8% of the NLD predictions were closer than ±20 mm to the observed NLD values. Measurements of equal-sized NETs varied considerably in outer diameter, proportion, the nasopharyngeal part, and guide marks. Length differences of the bend-to-cuff distance, containing the anatomically NLD, ranged between 218 and 270 mm at same sizes.
Conclusions: A reliable prediction of NLD can be obtained simply by body height, using the formula (Equation is included in full-text article.). As manufacturers' tube lengths vary substantially, additional information about the bend-to-cuff distance as corresponding tube section would allow for more accurate tube selection.
Similar articles
-
Comparison of cuffed and uncuffed preformed oral pediatric tracheal tubes.Paediatr Anaesth. 2006 Jul;16(7):734-42. doi: 10.1111/j.1460-9592.2006.01844.x. Paediatr Anaesth. 2006. PMID: 16879515
-
Nares-to-carina distance in children: does a 'modified Morgan formula' give useful guidance during nasal intubation?Paediatr Anaesth. 2015 Sep;25(9):936-42. doi: 10.1111/pan.12693. Epub 2015 May 29. Paediatr Anaesth. 2015. PMID: 26033678
-
The relationship between a person's height and appropriate endotracheal tube length.Anaesth Intensive Care. 1992 May;20(2):156-60. doi: 10.1177/0310057X9202000206. Anaesth Intensive Care. 1992. PMID: 1595848
-
Selection and placement of oral ventilation tubes based on tracheal morphometry.Laryngoscope. 2011 Jun;121(6):1225-30. doi: 10.1002/lary.21752. Epub 2011 May 6. Laryngoscope. 2011. PMID: 21557233
-
Anatomical In Vitro Investigations of the Pediatric Larynx: A Call for Manufacturer Redesign of Tracheal Tube Cuff Location and Perhaps a Call to Reconsider the Use of Uncuffed Tracheal Tubes.Anesth Analg. 2021 Oct 1;133(4):894-902. doi: 10.1213/ANE.0000000000005565. Anesth Analg. 2021. PMID: 33901057
Cited by
-
Nasotracheal intubation for airway management during anesthesia.Anesth Pain Med (Seoul). 2021 Jul;16(3):232-247. doi: 10.17085/apm.21040. Epub 2021 Jul 30. Anesth Pain Med (Seoul). 2021. PMID: 34352965 Free PMC article. Review.
-
Does intubation while observing the glottis with a fiberoptic scope reduce postoperative sore throat?BMC Anesthesiol. 2024 Nov 18;24(1):419. doi: 10.1186/s12871-024-02807-z. BMC Anesthesiol. 2024. PMID: 39558251 Free PMC article.
-
Comparison of the selection of nasotracheal tube diameter based on the patient's sex or size of the nasal airway: A prospective observational study.PLoS One. 2021 Mar 8;16(3):e0248296. doi: 10.1371/journal.pone.0248296. eCollection 2021. PLoS One. 2021. PMID: 33684167 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials