Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2019 Oct;82(4):1385-1397.
doi: 10.1002/mrm.27819. Epub 2019 Jun 12.

Optimized quantification of spin relaxation times in the hybrid state

Affiliations

Optimized quantification of spin relaxation times in the hybrid state

Jakob Assländer et al. Magn Reson Med. 2019 Oct.

Abstract

Purpose: The optimization and analysis of spin ensemble trajectories in the hybrid state-a state in which the direction of the magnetization adiabatically follows the steady state while the magnitude remains in a transient state.

Methods: Numerical optimizations were performed to find spin ensemble trajectories that minimize the Cramér-Rao bound for T1 -encoding, T2 -encoding, and their weighted sum, respectively, followed by a comparison between the Cramér-Rao bounds obtained with our optimized spin-trajectories, Look-Locker sequences, and multi-spin-echo methods. Finally, we experimentally tested our optimized spin trajectories with in vivo scans of the human brain.

Results: After a nonrecurring inversion segment on the southern half of the Bloch sphere, all optimized spin trajectories pursue repetitive loops on the northern hemisphere in which the beginning of the first and the end of the last loop deviate from the others. The numerical results obtained in this work align well with intuitive insights gleaned directly from the governing equation. Our results suggest that hybrid-state sequences outperform traditional methods. Moreover, hybrid-state sequences that balance T1 - and T2 -encoding still result in near optimal signal-to-noise efficiency for each relaxation time. Thus, the second parameter can be encoded at virtually no extra cost.

Conclusions: We provided new insights into the optimal encoding processes of spin relaxation times in order to guide the design of robust and efficient pulse sequences. We found that joint acquisitions of T1 and T2 in the hybrid state are substantially more efficient than sequential encoding techniques.

Keywords: HSFP; MRF; SSFP; optimal control; parameter mapping; quantitative MRI.

PubMed Disclaimer

Figures

FIGURE 1
FIGURE 1
Equation (1) describes the spin dynamics and is visualized in (A). The white area indicates the steady-state ellipse which separates the area in which the magnetization grows (red) and shrinks (blue). This particular sub-figure is valid for the ratio T1∕T2 = 781 ms∕65 ms ≈ 12, which are values reported for brain white matter. The derivatives of Equation (1) with respect to T1 and T2 are depicted in (B) and (C), respectively. These plots are normalized by the respective relaxation times and are, therefore, valid for any combination of T1 and T2
FIGURE 2
FIGURE 2
The spin dynamics in inversion recovery balanced hybrid-state free precession (IR-bHSFP) sequences are depicted on Bloch-spheres (A, D, G, J, M, and P). The polar angle patterns are shown in (B, E, H, K, N, and Q), with the color scale providing a reference for the trajectories on the Bloch-sphere. The absolute value of the magnetization (with a negative sign indicating the southern hemisphere) and its normalized derivatives with respect to the relaxation times are the foundation of the relative Cramér-Rao bound and are shown in (C, F, I, L, O, and R). The optimizations depicted in the left-hand column are limited to 0 ≤ ϑπ/2, while the right-hand column shows the same optimizations with the limit 0 ≤ ϑπ/4
FIGURE 3
FIGURE 3
The depicted relative Cramér-Rao bounds are defined by Equations (4) and (5) and can be understood as a lower bound of the squared inverse SNR efficiency per unit time. They result from numerical optimization for rCRB(T1), rCRB(T2), and rCRB(T1) + rCRB(T2), while limiting the polar angle to 0 ≤ ϑπ/2 (lightly colored markers) and 0 ≤ ϑπ/4 (dark colored markers), respectively
FIGURE 4
FIGURE 4
The performance of IR-bHSFP sequences optimized with 0 ≤ ϑπ/4 (Figure 2J–R) is illustrated through plots of the relative Cramér-Rao bounds, which provide a lower bound for the noise in the estimated relaxation times. All patterns were optimized for T1 = 781 ms and T2 = 65 ms, as indicated by the red square, and were tested for the entire parameter space in a sample MRF dictionary. Note the logarithmic scale in all three dimensions
FIGURE 5
FIGURE 5
The performance of IR-bHSFP optimized with 0 ≤ ϑπ/4 (Figure 2J–R) is illustrated through plots of the correlation of the fingerprint using the optimization parameters T1 = 781 ms and T2 = 65 ms (red square) with the rest of the parameter space
FIGURE 6
FIGURE 6
The performance of IR-bHSFP optimized with 0 ≤ ϑπ/4 (Figure 2J–R) is depicted through plots of the correlation coefficient. The correlation matrix was calculated for the entire parameter space. Thereafter, a maximum intensity projection was performed along the T2 (A-D) and the T1-dimension (E-H), respectively. The result can be understood as the worst case correlations for a single parameter under consideration of all values of the other parameter. The red lines indicates the set of parameters used for optimizing the patterns
FIGURE 7
FIGURE 7
The in vivo data were acquired with the excitation patterns depicted in Figure 2K, N, and Q with the limit 0 ≤ ϑπ/4. The parameter maps have an in-plane resolution of 1 mm and were acquired in 3.8 s. Note the logarithmic scale of the T1 and T2 color maps
FIGURE 8
FIGURE 8
The parameter maps are the average of 49 measurements like the one shown in Figure 7, each acquired in 3.8 s with a 10 s pause in between. Note the logarithmic scale of the T1 and T2 color maps. The red rectangles indicate the regions of interest used for calculating the values in Table 1
FIGURE 9
FIGURE 9
In order to verify the theoretically derived noise properties, we depict the mean relaxation times divided by the standard deviation over the 49 consecutive experiments. This metric provides an analog to the signal-to-noise ratio. The upper limit of this metric is directly proportional to the square root of the inverse rCRB(T1,2)

Similar articles

Cited by

References

    1. Jiang Y, Ma D, Seiberlich N, Gulani V, Griswold MA. MR fingerprinting using fast imaging with steady state precession (FISP) with spiral readout. Magn Reson Med. 2015;74:1621–1631. - PMC - PubMed
    1. Crawley AP, Henkelman RM. A comparison of one-shot and recovery methods in T1 imaging. Magn Reson Med. 1988; 7: 23–34. - PubMed
    1. Cheng HLM, Wright GA. Rapid high-resolution T1 mapping by variable flip angles: accurate and precise measurements in the presence of radiofrequency field inhomogeneity. Magn Reson Med. 2006;55:566–574. - PubMed
    1. Deoni SCL, Rutt BK, Peters TM. Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state. Magn Reson Med. 2003;49:515–526. - PubMed
    1. Deoni SCL, Peters TM, Rutt BK. Determination of optimal angles for variable nutation proton magnetic spin-lattice, T1, and Spin-Spin, T2, relaxation times measurement. Magn Reson Med. 2004;51:194–199. - PubMed

Publication types

MeSH terms